cardiology careers collections past issues search home
     

J Am Coll Cardiol, 2004; 44:1898-1902, doi:10.1016/j.jacc.2004.07.042
© 2004 by the American College of Cardiology Foundation
This Article
Right arrow Figures Only
Right arrow Full Text
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Gunderson, B. D.
Right arrow Articles by Ellenbogen, K. A.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Gunderson, B. D.
Right arrow Articles by Ellenbogen, K. A.

HEART RHYTHM DISTURBANCES

An algorithm to predict implantable cardioverter-defibrillator lead failure

Bruce D. Gunderson, MS*, Amisha S. Patel, MS*, Chad A. Bounds, BSEE*, Richard K. Shepard, MD{dagger}, Mark A. Wood, MD, FACC{dagger} and Kenneth A. Ellenbogen, MD, FACC{dagger},*

* *Medtronic, Inc., Minneapolis, Minnesota, USA
{dagger} Division of Cardiology, Medical College of Virginia/Virginia Commonwealth University, Richmond, Virginia, USA

Manuscript received December 4, 2003; accepted July 29, 2004.

* Reprint requests and correspondence: Dr. Kenneth A. Ellenbogen, Medical College of Virginia, PO Box 980053, Richmond, Virginia 23398-0053 (Email: kellenbogen{at}pol.net).

OBJECTIVES: The goal of this analysis was to test an algorithm that identifies implantable cardioverter-defibrillator (ICD) lead problems before clinical failure and/or inappropriate therapy.

BACKGROUND: The ICD lead failures typically present as inappropriate shock therapy. Identifying lead failures before their clinical presentation may prevent patient discomfort, improve device longevity, and avoid device-induced proarrhythmia.

METHODS: We tested an algorithm that uses two measures of oversensing and one measure of abnormal impedance to detect a lead failure. The oversensing measures consisted of a counter for RR intervals <140 ms and nonsustained ventricular tachycardia episodes with mean RR interval <200 ms. The impedance measure tracked lead impedances every day and each week. Abnormal impedance was defined as a decrease in impedances or an outlier value compared with baseline. Lead failures were identified when both oversensing measures were met or abnormal impedance and one oversensing measure occurred. The stored data from 696 patients with an ICD were analyzed to determine the sensitivity and specificity of the algorithm to detect lead failures.

RESULTS: Twenty-nine patients demonstrated clinical lead failures with an average of 6 ± 9 inappropriate shocks per patient. The two oversensing measures used in the algorithm predicted 72% (21 of 29) of the lead failures. Fulfilling at least two of the three impedance and oversensing measures, the sensitivity of our algorithm was 83% (24 of 29) with a 100% (667 of 667) specificity.

CONCLUSION: Oversensing combined with abnormal impedance trends may be used to identify ICD lead failures with high sensitivity and very high specificity.

Abbreviations and Acronyms
  HV = high voltage
  ICD = implantable cardioverter-defibrillator
  NST = nonsustained tachycardia
  RR = consecutive QRS complexes
  RV = right ventricle/ventricular
  SIC = sensing integrity counter




This article has been cited by other articles:


Home page
CirculationHome page
C. D. Swerdlow, B. D. Gunderson, K. T. Ousdigian, A. Abeyratne, R. W. Stadler, J. M. Gillberg, A. S. Patel, and K. A. Ellenbogen
Downloadable Algorithm to Reduce Inappropriate Shocks Caused by Fractures of Implantable Cardioverter-Defibrillator Leads
Circulation, November 18, 2008; 118(21): 2122 - 2129.
[Abstract] [Full Text] [PDF]


Home page
EuropaceHome page
A. Goette, F. Cantu, L. van Erven, P. Geelen, F. Halimi, J. L. Merino, J. M. Morgan, and on behalf of the Scientific Initiative Committee o
Performance and survival of transvenous defibrillation leads: need for a European data registry
Europace, November 11, 2008; (2008) eun301v1.
[Abstract] [Full Text] [PDF]


Home page
EuropaceHome page
D. Vollmann, L. Luthje, and M. Zabel
Far-field oversensing of atrial signals: an unusual cause for very short V-V intervals and inappropriate implantable cardioverter defibrillator therapy
Europace, August 1, 2008; 10(8): 1009 - 1011.
[Full Text] [PDF]


Home page
EuropaceHome page
J. C. Nielsen, H. Kottkamp, M. Zabel, E. Aliot, U. Kreutzer, A. Bauer, A. Schuchert, H. Neuser, B. Schumacher, H. Schmidinger, et al.
Automatic home monitoring of implantable cardioverter defibrillators
Europace, June 1, 2008; 10(6): 729 - 735.
[Abstract] [Full Text] [PDF]


Home page
J Am Coll CardiolHome page
A. E. Epstein, J. P. DiMarco, K. A. Ellenbogen, N.A. M. Estes III, R. A. Freedman, L. S. Gettes, A. M. Gillinov, G. Gregoratos, S. C. Hammill, D. L. Hayes, et al.
ACC/AHA/HRS 2008 Guidelines for Device-Based Therapy of Cardiac Rhythm Abnormalities: A Report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines (Writing Committee to Revise the ACC/AHA/NASPE 2002 Guideline Update for Implantation of Cardiac Pacemakers and Antiarrhythmia Devices) Developed in Collaboration With the American Association for Thoracic Surgery and Society of Thoracic Surgeons
J. Am. Coll. Cardiol., May 27, 2008; 51(21): e1 - e62.
[Full Text] [PDF]


Home page
CirculationHome page
A. E. Epstein, J. P. DiMarco, K. A. Ellenbogen, N.A. M. Estes III, R. A. Freedman, L. S. Gettes, A. M. Gillinov, G. Gregoratos, S. C. Hammill, D. L. Hayes, et al.
ACC/AHA/HRS 2008 Guidelines for Device-Based Therapy of Cardiac Rhythm Abnormalities: A Report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines (Writing Committee to Revise the ACC/AHA/NASPE 2002 Guideline Update for Implantation of Cardiac Pacemakers and Antiarrhythmia Devices): Developed in Collaboration With the American Association for Thoracic Surgery and Society of Thoracic Surgeons
Circulation, May 27, 2008; 117(21): e350 - e408.
[Full Text] [PDF]


Home page
EuropaceHome page
D. Vollmann, L. Luthje, and M. Zabel
Unusual cause for an increase of the sensing integrity counter in a patient with inappropriate implantable cardioverter-defibrillator therapy
Europace, May 1, 2007; 9(5): 275 - 277.
[Abstract] [Full Text] [PDF]


Home page
EuropaceHome page
D. Vollmann, A. Erdogan, E. Himmrich, J. Neuzner, D. Becker, C. Unterberg-Buchwald, J. Sperzel, and for the SAFE Study Investigators
Patient AlertTM to detect ICD lead failure: efficacy, limitations, and implications for future algorithms.
Europace, January 1, 2006; 8(5): 371 - 376.
[Abstract] [Full Text] [PDF]



 
  cardiology careers collections past issues search home