OBJECTIVES: This study evaluated the long-term reliability of an implantable cardioverter defibrillator (ICD) lead to determine the incidence, clinical presentation, and management of lead failure.
BACKGROUND: Despite recent advances in ICD technology, the long-term reliability of ICD leads remains a significant problem.
METHODS: Concern about long-term reliability of coaxial polyurethane ICD leads caused us to systematically study all patients implanted with Medtronic (Minneapolis, Minnesota) 6936 lead at our institution. We performed follow-up of 74 patients with 76 ICD leads that were implanted from February 28, 1995 to September 8, 1997. Thirty-seven patients underwent routine clinical ICD follow-up testing and ventricular fibrillation induction to determine the status of their ICD lead after a mean follow-up of 68.6 ± 8.2 months.
RESULTS: The lead survival analysis shows a cumulative failure probability of 37% (confidence interval, 24% to 54%) at 68.6 months. Six patients demonstrated a previously undescribed mode of ICD lead failure: prolonged oversensing immediately after shock therapy. The use of short interval counters to monitor nonphysiologic R-R intervals and the measurement of ring-to-coil impedance detected early lead failures in five patients.
CONCLUSIONS: This analysis shows: 1) problems with ICD leads may not become apparent until late during follow-up and may become a significant late problem, 2) a "signature" mode of lead failure for the 6936 consisting of oversensing of electrical noise following shocks, 3) early detection of lead failure with a short interval counter algorithm or measurement of ring-to-coil impedance may be clinically useful.
| ||||||||||
This article has been cited by other articles:
![]() |
M. Chinushi, Y. Hosaka, N. Ikarashi, K. Iijima, H. Furushima, and Y. Aizawa Automatic R-wave and impedance testing with the modern patient alert system to reduce inappropriate implantable cardioverter defibrillator shocks due to lead fracture Europace, June 1, 2008; 10(6): 738 - 740. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. C. Nielsen, H. Kottkamp, M. Zabel, E. Aliot, U. Kreutzer, A. Bauer, A. Schuchert, H. Neuser, B. Schumacher, H. Schmidinger, et al. Automatic home monitoring of implantable cardioverter defibrillators Europace, June 1, 2008; 10(6): 729 - 735. [Abstract] [Full Text] [PDF] |
||||
![]() |
W. H. Maisel and D. B. Kramer Implantable Cardioverter-Defibrillator Lead Performance Circulation, May 27, 2008; 117(21): 2721 - 2723. [Full Text] [PDF] |
||||
![]() |
J. Eckstein, M. T. Koller, M. Zabel, D. Kalusche, B. A. Schaer, S. Osswald, and C. Sticherling Necessity for Surgical Revision of Defibrillator Leads Implanted Long-Term: Causes and Management Circulation, May 27, 2008; 117(21): 2727 - 2733. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Rozner Management of Implanted Cardiac Defibrillators During Eye Surgery Anesth. Analg., February 1, 2008; 106(2): 671 - 672. [Full Text] [PDF] |
||||
![]() |
T. Rauwolf, M. Guenther, N. Hass, A. Schnabel, M. Bock, M.U. Braun, and R.H. Strasser Ventricular oversensing in 518 patients with implanted cardiac defibrillators: incidence, complications, and solutions Europace, November 1, 2007; 9(11): 1041 - 1047. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. Kleemann, T. Becker, K. Doenges, M. Vater, J. Senges, S. Schneider, W. Saggau, U. Weisse, and K. Seidl Response to Letter Regarding Article, "Annual Rate of Transvenous Defibrillation Lead Defects in Cardioverter-Defibrillators Over a Period of >10 Years" Circulation, October 30, 2007; 116(18): e513 - e513. [Full Text] [PDF] |
||||
![]() |
J. A. Ezekowitz, B. H. Rowe, D. M. Dryden, N. Hooton, B. Vandermeer, C. Spooner, and F. A. McAlister Systematic Review: Implantable Cardioverter Defibrillators for Adults with Left Ventricular Systolic Dysfunction Ann Intern Med, August 21, 2007; 147(4): 251 - 262. [Abstract] [Full Text] [PDF] |
||||
![]() |
W. H. Maisel Transvenous Implantable Cardioverter-Defibrillator Leads: The Weakest Link Circulation, May 15, 2007; 115(19): 2461 - 2463. [Full Text] [PDF] |
||||
![]() |
T. Kleemann, T. Becker, K. Doenges, M. Vater, J. Senges, S. Schneider, W. Saggau, U. Weisse, and K. Seidl Annual Rate of Transvenous Defibrillation Lead Defects in Implantable Cardioverter-Defibrillators Over a Period of >10 Years Circulation, May 15, 2007; 115(19): 2474 - 2480. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. Vollmann, L. Luthje, and M. Zabel Unusual cause for an increase of the sensing integrity counter in a patient with inappropriate implantable cardioverter-defibrillator therapy Europace, May 1, 2007; 9(5): 275 - 277. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. H. Schoenfeld Contemporary Pacemaker and Defibrillator Device Therapy: Challenges Confronting the General Cardiologist Circulation, February 6, 2007; 115(5): 638 - 653. [Full Text] [PDF] |
||||
![]() |
E. Occhetta, M. Bortnik, A. Magnani, G. Francalacci, and P. Marino Inappropriate implantable cardioverter-defibrillator discharges unrelated to supraventricular tachyarrhythmias Europace, October 1, 2006; 8(10): 863 - 869. [Abstract] [Full Text] [PDF] |
||||
![]() |
Developed in Collaboration With the European Heart, D. P. Zipes, A. J. Camm, M. Borggrefe, A. E. Buxton, B. Chaitman, M. Fromer, G. Gregoratos, G. Klein, A. J. Moss, et al. ACC/AHA/ESC 2006 Guidelines for Management of Patients With Ventricular Arrhythmias and the Prevention of Sudden Cardiac Death: A Report of the American College of Cardiology/American Heart Association Task Force and the European Society of Cardiology Committee for Practice Guidelines (Writing Committee to Develop Guidelines for Management of Patients With Ventricular Arrhythmias and the Prevention of Sudden Cardiac Death) J. Am. Coll. Cardiol., September 5, 2006; 48(5): e247 - e346. [Full Text] [PDF] |
||||
![]() |
Writing Committee Members, D. P. Zipes, A. J. Camm, M. Borggrefe, A. E. Buxton, B. Chaitman, M. Fromer, G. Gregoratos, G. Klein, A. J. Moss, et al. ACC/AHA/ESC 2006 guidelines for management of patients with ventricular arrhythmias and the prevention of sudden cardiac death: A report of the American College of Cardiology/American Heart Association Task Force and the European Society of Cardiology Committee for Practice Guidelines (Writing Committee to Develop Guidelines for Management of Patients With Ventricular Arrhythmias and the Prevention of Sudden Cardiac Death) Developed in collaboration with the European Heart Rhythm Association and the Heart Rhythm Society Europace, September 1, 2006; 8(9): 746 - 837. [Full Text] [PDF] |
||||
![]() |
D. Vollmann, A. Erdogan, E. Himmrich, J. Neuzner, D. Becker, C. Unterberg-Buchwald, J. Sperzel, and for the SAFE Study Investigators Patient AlertTM to detect ICD lead failure: efficacy, limitations, and implications for future algorithms. Europace, January 1, 2006; 8(5): 371 - 376. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. C. Senges-Becker, M. Klostermann, R. Becker, A. Bauer, K. E. Siegler, H. A. Katus, and W. Schoels What is the "Optimal" follow-up schedule for ICD patients? Europace, January 1, 2005; 7(4): 319 - 326. [Abstract] [Full Text] [PDF] |
||||
![]() |
B. D. Gunderson, A. S. Patel, C. A. Bounds, R. K. Shepard, M. A. Wood, and K. A. Ellenbogen An algorithm to predict implantable cardioverter-defibrillator lead failure J. Am. Coll. Cardiol., November 2, 2004; 44(9): 1898 - 1902. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. Becker, J. Ruf-Richter, J. C. Senges-Becker, A. Bauer, S. Weretka, F. Voss, H. A. Katus, and W. Schoels Patient alert in implantable cardioverter defibrillators: toy or tool? J. Am. Coll. Cardiol., July 7, 2004; 44(1): 95 - 98. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. Roguin, C. S. Bomma, K. Nasir, H. Tandri, C. Tichnell, C. James, J. Rutberg, J. Crosson, P. J. Spevak, R. D. Berger, et al. Implantable Cardioverter-Defibrillators in patients with arrhythmogenic right ventricular Dysplasia/Cardiomyopathy J. Am. Coll. Cardiol., May 19, 2004; 43(10): 1843 - 1852. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. Wichter, M. Paul, C. Wollmann, T. Acil, P. Gerdes, O. Ashraf, T. D.T. Tjan, R. Soeparwata, M. Block, M. Borggrefe, et al. Implantable Cardioverter/Defibrillator Therapy in Arrhythmogenic Right Ventricular Cardiomyopathy: Single-Center Experience of Long-Term Follow-Up and Complications in 60 Patients Circulation, March 30, 2004; 109(12): 1503 - 1508. [Abstract] [Full Text] [PDF] |
||||
![]() |
Lead Failure Remains the Achilles' Heel of the ICD Journal Watch Cardiology, February 21, 2003; 2003(221): 2 - 2. [Full Text] |
||||
| HOME | SUBSCRIPTIONS | CURRENT ISSUE | PAST ISSUES | CARDIOSOURCE | SEARCH | HELP | FEEDBACK |