JACC
HOME SUBSCRIPTIONS CURRENT ISSUE PAST ISSUES CARDIOSOURCE SEARCH HELP FEEDBACK
 QUICK SEARCH:   [advanced]


     


J Am Coll Cardiol, 2003; 41:73-80
© 2003 by the American College of Cardiology Foundation
This Article
Right arrow Figures Only
Right arrow Full Text
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Ellenbogen, K. A.
Right arrow Articles by Verness, D.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Ellenbogen, K. A.
Right arrow Articles by Verness, D.

CLINICAL STUDY: CARDIAC ARRHYTHMIAS

Detection and management of animplantable cardioverter defibrillator lead failure

Incidence and clinical implications

Kenneth A. Ellenbogen, MD, FACC*,*, Mark A. Wood, MD, FACC*, Richard K. Shepard, MD*, Henry F. Clemo, MD, PhD, FACC*, Tim Vaughn, BS{dagger}, Keith Holloman, BA{dagger}, Marlene Dow, RN{dagger}, Jeanne Leffler, RN*, Athula Abeyratne, PhD{dagger} and Dave Verness, BS, PE, CQE{dagger}

* Division of Cardiology, Medical College of Virginia/Virginia Commonwealth University, Richmond, Virginia, USA
{dagger} Medtronic, Inc., Minneapolis, Minnesota, USA

Manuscript received May 6, 2002; revised manuscript received August 6, 2002, accepted August 19, 2002.

* Reprint requests and correspondence: Dr. Kenneth A. Ellenbogen, Medical College of Virginia, P.O. Box 980053, Richmond, Virginia 23398-0053, USA.
kellenbogen{at}pol.net

OBJECTIVES: This study evaluated the long-term reliability of an implantable cardioverter defibrillator (ICD) lead to determine the incidence, clinical presentation, and management of lead failure.

BACKGROUND: Despite recent advances in ICD technology, the long-term reliability of ICD leads remains a significant problem.

METHODS: Concern about long-term reliability of coaxial polyurethane ICD leads caused us to systematically study all patients implanted with Medtronic (Minneapolis, Minnesota) 6936 lead at our institution. We performed follow-up of 74 patients with 76 ICD leads that were implanted from February 28, 1995 to September 8, 1997. Thirty-seven patients underwent routine clinical ICD follow-up testing and ventricular fibrillation induction to determine the status of their ICD lead after a mean follow-up of 68.6 ± 8.2 months.

RESULTS: The lead survival analysis shows a cumulative failure probability of 37% (confidence interval, 24% to 54%) at 68.6 months. Six patients demonstrated a previously undescribed mode of ICD lead failure: prolonged oversensing immediately after shock therapy. The use of short interval counters to monitor nonphysiologic R-R intervals and the measurement of ring-to-coil impedance detected early lead failures in five patients.

CONCLUSIONS: This analysis shows: 1) problems with ICD leads may not become apparent until late during follow-up and may become a significant late problem, 2) a "signature" mode of lead failure for the 6936 consisting of oversensing of electrical noise following shocks, 3) early detection of lead failure with a short interval counter algorithm or measurement of ring-to-coil impedance may be clinically useful.

Abbreviations and Acronyms
  CI
  confidence interval
  DFT
  defibrillation threshold
  ICD
  implantable cardioverter defibrillator
  MIO
  metal ion oxidation




This article has been cited by other articles:


Home page
EuropaceHome page
M. Chinushi, Y. Hosaka, N. Ikarashi, K. Iijima, H. Furushima, and Y. Aizawa
Automatic R-wave and impedance testing with the modern patient alert system to reduce inappropriate implantable cardioverter defibrillator shocks due to lead fracture
Europace, June 1, 2008; 10(6): 738 - 740.
[Abstract] [Full Text] [PDF]


Home page
EuropaceHome page
J. C. Nielsen, H. Kottkamp, M. Zabel, E. Aliot, U. Kreutzer, A. Bauer, A. Schuchert, H. Neuser, B. Schumacher, H. Schmidinger, et al.
Automatic home monitoring of implantable cardioverter defibrillators
Europace, June 1, 2008; 10(6): 729 - 735.
[Abstract] [Full Text] [PDF]


Home page
CirculationHome page
W. H. Maisel and D. B. Kramer
Implantable Cardioverter-Defibrillator Lead Performance
Circulation, May 27, 2008; 117(21): 2721 - 2723.
[Full Text] [PDF]


Home page
CirculationHome page
J. Eckstein, M. T. Koller, M. Zabel, D. Kalusche, B. A. Schaer, S. Osswald, and C. Sticherling
Necessity for Surgical Revision of Defibrillator Leads Implanted Long-Term: Causes and Management
Circulation, May 27, 2008; 117(21): 2727 - 2733.
[Abstract] [Full Text] [PDF]


Home page
Anesth. Analg.Home page
M. Rozner
Management of Implanted Cardiac Defibrillators During Eye Surgery
Anesth. Analg., February 1, 2008; 106(2): 671 - 672.
[Full Text] [PDF]


Home page
EuropaceHome page
T. Rauwolf, M. Guenther, N. Hass, A. Schnabel, M. Bock, M.U. Braun, and R.H. Strasser
Ventricular oversensing in 518 patients with implanted cardiac defibrillators: incidence, complications, and solutions
Europace, November 1, 2007; 9(11): 1041 - 1047.
[Abstract] [Full Text] [PDF]


Home page
CirculationHome page
T. Kleemann, T. Becker, K. Doenges, M. Vater, J. Senges, S. Schneider, W. Saggau, U. Weisse, and K. Seidl
Response to Letter Regarding Article, "Annual Rate of Transvenous Defibrillation Lead Defects in Cardioverter-Defibrillators Over a Period of >10 Years"
Circulation, October 30, 2007; 116(18): e513 - e513.
[Full Text] [PDF]


Home page
ANN INTERN MEDHome page
J. A. Ezekowitz, B. H. Rowe, D. M. Dryden, N. Hooton, B. Vandermeer, C. Spooner, and F. A. McAlister
Systematic Review: Implantable Cardioverter Defibrillators for Adults with Left Ventricular Systolic Dysfunction
Ann Intern Med, August 21, 2007; 147(4): 251 - 262.
[Abstract] [Full Text] [PDF]


Home page
CirculationHome page
W. H. Maisel
Transvenous Implantable Cardioverter-Defibrillator Leads: The Weakest Link
Circulation, May 15, 2007; 115(19): 2461 - 2463.
[Full Text] [PDF]


Home page
CirculationHome page
T. Kleemann, T. Becker, K. Doenges, M. Vater, J. Senges, S. Schneider, W. Saggau, U. Weisse, and K. Seidl
Annual Rate of Transvenous Defibrillation Lead Defects in Implantable Cardioverter-Defibrillators Over a Period of >10 Years
Circulation, May 15, 2007; 115(19): 2474 - 2480.
[Abstract] [Full Text] [PDF]


Home page
EuropaceHome page
D. Vollmann, L. Luthje, and M. Zabel
Unusual cause for an increase of the sensing integrity counter in a patient with inappropriate implantable cardioverter-defibrillator therapy
Europace, May 1, 2007; 9(5): 275 - 277.
[Abstract] [Full Text] [PDF]


Home page
CirculationHome page
M. H. Schoenfeld
Contemporary Pacemaker and Defibrillator Device Therapy: Challenges Confronting the General Cardiologist
Circulation, February 6, 2007; 115(5): 638 - 653.
[Full Text] [PDF]


Home page
EuropaceHome page
E. Occhetta, M. Bortnik, A. Magnani, G. Francalacci, and P. Marino
Inappropriate implantable cardioverter-defibrillator discharges unrelated to supraventricular tachyarrhythmias
Europace, October 1, 2006; 8(10): 863 - 869.
[Abstract] [Full Text] [PDF]


Home page
J Am Coll CardiolHome page
Developed in Collaboration With the European Heart, D. P. Zipes, A. J. Camm, M. Borggrefe, A. E. Buxton, B. Chaitman, M. Fromer, G. Gregoratos, G. Klein, A. J. Moss, et al.
ACC/AHA/ESC 2006 Guidelines for Management of Patients With Ventricular Arrhythmias and the Prevention of Sudden Cardiac Death: A Report of the American College of Cardiology/American Heart Association Task Force and the European Society of Cardiology Committee for Practice Guidelines (Writing Committee to Develop Guidelines for Management of Patients With Ventricular Arrhythmias and the Prevention of Sudden Cardiac Death)
J. Am. Coll. Cardiol., September 5, 2006; 48(5): e247 - e346.
[Full Text] [PDF]


Home page
EuropaceHome page
Writing Committee Members, D. P. Zipes, A. J. Camm, M. Borggrefe, A. E. Buxton, B. Chaitman, M. Fromer, G. Gregoratos, G. Klein, A. J. Moss, et al.
ACC/AHA/ESC 2006 guidelines for management of patients with ventricular arrhythmias and the prevention of sudden cardiac death: A report of the American College of Cardiology/American Heart Association Task Force and the European Society of Cardiology Committee for Practice Guidelines (Writing Committee to Develop Guidelines for Management of Patients With Ventricular Arrhythmias and the Prevention of Sudden Cardiac Death) Developed in collaboration with the European Heart Rhythm Association and the Heart Rhythm Society
Europace, September 1, 2006; 8(9): 746 - 837.
[Full Text] [PDF]


Home page
EuropaceHome page
D. Vollmann, A. Erdogan, E. Himmrich, J. Neuzner, D. Becker, C. Unterberg-Buchwald, J. Sperzel, and for the SAFE Study Investigators
Patient AlertTM to detect ICD lead failure: efficacy, limitations, and implications for future algorithms.
Europace, January 1, 2006; 8(5): 371 - 376.
[Abstract] [Full Text] [PDF]


Home page
EuropaceHome page
J. C. Senges-Becker, M. Klostermann, R. Becker, A. Bauer, K. E. Siegler, H. A. Katus, and W. Schoels
What is the "Optimal" follow-up schedule for ICD patients?
Europace, January 1, 2005; 7(4): 319 - 326.
[Abstract] [Full Text] [PDF]


Home page
J Am Coll CardiolHome page
B. D. Gunderson, A. S. Patel, C. A. Bounds, R. K. Shepard, M. A. Wood, and K. A. Ellenbogen
An algorithm to predict implantable cardioverter-defibrillator lead failure
J. Am. Coll. Cardiol., November 2, 2004; 44(9): 1898 - 1902.
[Abstract] [Full Text] [PDF]


Home page
J Am Coll CardiolHome page
R. Becker, J. Ruf-Richter, J. C. Senges-Becker, A. Bauer, S. Weretka, F. Voss, H. A. Katus, and W. Schoels
Patient alert in implantable cardioverter defibrillators: toy or tool?
J. Am. Coll. Cardiol., July 7, 2004; 44(1): 95 - 98.
[Abstract] [Full Text] [PDF]


Home page
J Am Coll CardiolHome page
A. Roguin, C. S. Bomma, K. Nasir, H. Tandri, C. Tichnell, C. James, J. Rutberg, J. Crosson, P. J. Spevak, R. D. Berger, et al.
Implantable Cardioverter-Defibrillators in patients with arrhythmogenic right ventricular Dysplasia/Cardiomyopathy
J. Am. Coll. Cardiol., May 19, 2004; 43(10): 1843 - 1852.
[Abstract] [Full Text] [PDF]


Home page
CirculationHome page
T. Wichter, M. Paul, C. Wollmann, T. Acil, P. Gerdes, O. Ashraf, T. D.T. Tjan, R. Soeparwata, M. Block, M. Borggrefe, et al.
Implantable Cardioverter/Defibrillator Therapy in Arrhythmogenic Right Ventricular Cardiomyopathy: Single-Center Experience of Long-Term Follow-Up and Complications in 60 Patients
Circulation, March 30, 2004; 109(12): 1503 - 1508.
[Abstract] [Full Text] [PDF]


Home page
Journal Watch CardiologyHome page
Lead Failure Remains the Achilles' Heel of the ICD
Journal Watch Cardiology, February 21, 2003; 2003(221): 2 - 2.
[Full Text]




HOME SUBSCRIPTIONS CURRENT ISSUE PAST ISSUES CARDIOSOURCE SEARCH HELP FEEDBACK
Copyright © 2003 by the American College of Cardiology Foundation.