OBJECTIVES: This study was designed to test the feasibility of high-resolution phased-array intracardiac imaging.
BACKGROUND: Intracardiac echocardiographic imaging of the heart during interventional electrophysiologic (EP) procedures has been limited by inadequate ultrasound penetration and absence of Doppler hemodynamic and flow information produced by rotating mechanical ultrasound elements.
METHODS: A 10F (3.2 mm) phased-array, variable 5.5 to 10 MHz frequency imaging catheter with a four-way deflectable tip was applied in 24 patients undergoing EP studies. Sixteen prespecified cardiac targets were imaged from a right heart venue.
RESULTS: Fifteen patients had no underlying organic heart disease; nine had ischemic, cardiomyopathic, valvular or congenital heart disorders. Longitudinal and short-axis imaging readily disclosed each cardiac valve, support structures and chamber, as well as the pericardium, right and left atrial appendages, the junction of the right atrium and superior vena cava, crista terminalis, tricuspid valve isthmus, coronary sinus orifice, membranous fossa ovalis and pulmonary veins. The average target depth was 8.8 ± 1.5 cm (range 0.5 to 15 cm), with adequate penetration at a 7.5 MHz imaging frequency. Color flow and Doppler utilities clearly characterized transaortic and pulmonic valve and pulmonary vein blood flow, including during low output states.
CONCLUSIONS: These first human studies with this technology demonstrate the methods, feasibility and utility of intracardiac phased-array vector and Doppler imaging for long-axis, apex-to-base global cardiac imaging. High resolution of endocardial structures and catheters suggests additional utility for visualizing interventional procedures from the right heart.
| |||||||||||||||||
This article has been cited by other articles:
![]() |
D. L. Packer, S. B. Johnson, M. W. Kolasa, T. J. Bunch, B. D. Henz, and Y. Okumura New generation of electro-anatomic mapping: full intracardiac ultrasound image integration Europace, November 1, 2008; 10(suppl_3): iii35 - iii41. [Abstract] [Full Text] [PDF] |
||||
![]() |
Y. Okumura, B. D. Henz, S. B. Johnson, T. J. Bunch, C. J. O'Brien, D. O. Hodge, A. Altman, A. Govari, and D. L. Packer Three-Dimensional Ultrasound for Image-Guided Mapping and Intervention: Methods, Quantitative Validation, and Clinical Feasibility of a Novel Multimodality Image Mapping System Circ Arrhythmia Electrophysiol, June 1, 2008; 1(2): 110 - 119. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. Ota, D. Schwartzman, D. Francischelli, D. A. Hettrick, and M. A. Zenati Impact of beating heart left atrial ablation on left-sided heart mechanics. J. Thorac. Cardiovasc. Surg., October 1, 2007; 134(4): 982 - 988. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. Ota, T. W. Gilbert, S. F. Badylak, D. Schwartzman, and M. A. Zenati Electromechanical characterization of a tissue-engineered myocardial patch derived from extracellular matrix J. Thorac. Cardiovasc. Surg., April 1, 2007; 133(4): 979 - 985. [Abstract] [Full Text] [PDF] |
||||
![]() |
B. D. Hoit Echocardiographic Characterization of the Cardiovascular Phenotype in Rodent Models Toxicol Pathol, January 1, 2006; 34(1): 105 - 110. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. L. Packer, P. Keelan, T. M. Munger, J. F. Breen, S. Asirvatham, L. A. Peterson, K. H. Monahan, M. F. Hauser, K. Chandrasekaran, L. J. Sinak, et al. Clinical Presentation, Investigation, and Management of Pulmonary Vein Stenosis Complicating Ablation for Atrial Fibrillation Circulation, February 8, 2005; 111(5): 546 - 554. [Abstract] [Full Text] [PDF] |
||||
![]() |
E. B. Saad, A. Rossillo, C. P. Saad, D. O. Martin, M. Bhargava, D. Erciyes, D. Bash, M. Williams-Andrews, S. Beheiry, N. F. Marrouche, et al. Pulmonary Vein Stenosis After Radiofrequency Ablation of Atrial Fibrillation: Functional Characterization, Evolution, and Influence of the Ablation Strategy Circulation, December 23, 2003; 108(25): 3102 - 3107. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. Bonanomi, D. Schwartzman, D. Francischelli, K. Hebsgaard, and M. A. Zenati A new device for beating heart bipolar radiofrequency atrial ablation J. Thorac. Cardiovasc. Surg., December 1, 2003; 126(6): 1859 - 1866. [Abstract] [Full Text] [PDF] |
||||
![]() |
A Doi, M Takagi, I Toda, M Teragaki, M Yoshiyama, K Takeuchi, and J Yoshikawa Real time quantification of low temperature radiofrequency ablation lesion size using phased array intracardiac echocardiography in the canine model: comparison of two dimensional images with pathological lesion characteristics Heart, August 1, 2003; 89(8): 923 - 927. [Abstract] [Full Text] [PDF] |
||||
![]() |
N. F. Marrouche, D. O. Martin, O. Wazni, A. M. Gillinov, A. Klein, M. Bhargava, E. Saad, D. Bash, H. Yamada, W. Jaber, et al. Phased-Array Intracardiac Echocardiography Monitoring During Pulmonary Vein Isolation in Patients With Atrial Fibrillation: Impact on Outcome and Complications Circulation, June 3, 2003; 107(21): 2710 - 2716. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. Tabata, J. D. Thomas, and A. L. Klein Pulmonary venous flow by doppler echocardiography: revisited 12 years later J. Am. Coll. Cardiol., April 16, 2003; 41(8): 1243 - 1250. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. Schwartzman, J. Lacomis, and W. G. Wigginton Characterization of left atrium and distal pulmonary vein morphology using multidimensional computed tomography J. Am. Coll. Cardiol., April 16, 2003; 41(8): 1349 - 1357. [Abstract] [Full Text] [PDF] |
||||
| HOME | SUBSCRIPTIONS | CURRENT ISSUE | PAST ISSUES | CARDIOSOURCE | SEARCH | HELP | FEEDBACK |