Advertisement





Click here for more guidelines.
CME Topic Collections Past Issues Search Current Issue Home
     

J Am Coll Cardiol, 1999; 33:883-891
© 1999 by the American College of Cardiology Foundation
This Article
Right arrow Figures Only
Right arrow Full Text
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Lehmann, K. G.
Right arrow Articles by Platt, M. S.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Lehmann, K. G.
Right arrow Articles by Platt, M. S.

EXPERIMENTAL STUDIES

Improved accuracy and precision of thermodilution cardiac output measurement using a dual thermistor catheter system

Kenneth G. Lehmann, MD, FACCa and Miriam S. Platt*

a Sections of Cardiology, University of Washington School of Medicine, Seattle, Washington, USA
* Veterans Affairs Puget Sound Health Care System, Seattle, Washington, USA

Manuscript received April 17, 1998; revised manuscript received September 18, 1998, accepted November 2, 1998.

Reprint requests and correspondence: Dr. Kenneth G. Lehmann, Section of Cardiology (111C), VA Puget Sound Health Care System, 1660 South Columbian Way, Seattle, Washington 98108

OBJECTIVES

To assess whether thermodilution cardiac output determination based on measurement of injectate temperature in vivo leads to more accurate and precise estimates and to study the influence of chilled injectate on test performance.

BACKGROUND

Cardiac output measurement via right heart catheterization is used extensively for hemodynamic evaluation in a variety of diagnostic, perioperative and critical care settings. Maximizing accuracy is essential for optimal patient care.

METHODS

This prospective study of 960 thermodilution cardiac output measurements was conducted using conventional and dual thermistor techniques. Specialized dual thermistor right heart catheters were constructed using a second thermistor positioned to measure injectate temperature in vivo just prior to entry into the right atrium. To eliminate interinjection variability, a custom set-up was developed that permitted output measurement using both techniques simultaneously. Both ambient temperature injections and cooled injections were investigated.

RESULTS

The dual thermistor technique demonstrated significantly less measurement variability than the conventional technique for both ambient temperature (precision = 0.41 vs. 0.55 L/min, p < 0.001) and cooled (precision = 0.35 vs. 0.43 L/min, p = 0.01) injections. Similarly, the average range of cardiac output values obtained during five sequential injections in each patient was less using the dual thermistor approach (1.05 vs. 1.55 L/min, p < 0.001). The use of cooled injectate reduced the mean error of the dual thermistor technique but actually increased the mean error of the conventional technique. Even with ambient temperature injections, injectate warming during catheter transit varied considerably and unpredictably from injection to injection (2 SD range = –0.22 to 5.74°C). Conventional ambient temperature and cooled measurements significantly overestimated Fick cardiac output measurements by 0.32 and 0.50 L/min, respectively (p < 0.001). In contrast, dual thermistor measurements were statistically similar (–0.08 and –0.08 L/min, p = 0.34) to Fick measurements.

CONCLUSIONS

This new dual thermistor approach results in a significant improvement in both precision and accuracy of thermodilution cardiac output measurement.

Abbreviations and Acronyms
  cardiac cath = cardiac catheterization
  CO = cardiac output
  temp = temperature




This article has been cited by other articles:


Home page
Am. J. Neuroradiol.Home page
A. Sorteberg, W. Sorteberg, A. Rappe, and C. M. Strother
Effect of Guglielmi Detachable Coils on Intraaneurysmal Flow: Experimental Study in Canines
AJNR Am. J. Neuroradiol., February 1, 2002; 23(2): 288 - 294.
[Abstract] [Full Text] [PDF]


Home page
Br J AnaesthHome page
Editorial I: The utility of pulmonary artery catheterization
Br. J. Anaesth., October 1, 2000; 85(4): 501 - 503.
[Full Text] [PDF]



 
  CME Topic Collections Past Issues Search Current Issue Home

Advertisement