J Am Coll Cardiol, 1998; 32:2065-2071
© 1998 by the American College of Cardiology Foundation
This Article
Right arrow Figures Only
Right arrow Full Text
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Citing Articles
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Pagan-Carlo, L. A.
Right arrow Articles by Kerber, R. E.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Pagan-Carlo, L. A.
Right arrow Articles by Kerber, R. E.

EXPERIMENTAL STUDIES

Encircling overlapping multipulse shock waveforms for transthoracic defibrillation

Luis A. Pagan-Carlo, MDa, John J. Allan, MDa, Kirk T. Spencer, MDa, Clay L. Birkett, BSEEa, Rick Myers, BSEEa and Richard E. Kerber, MD, FACCa

a Department of Internal Medicine, University of Iowa Hospital, Iowa City, Iowa, USA

Manuscript received January 5, 1998; revised manuscript received July 29, 1998, accepted August 20, 1998.

Address for correspondence: Dr. Kerber, Department of Internal Medicine, University of Iowa Hospital, 200 Hawkins Drive, Iowa City, Iowa 52242
dick-kerber{at}uiowa.edu

Objectives. This study was performed to determine the efficacy of new encircling overlapping multipulse, multipathway waveforms for transthoracic defibrillation.

Background. Alternative waveforms for transthoracic defibrillation may improve shock success.

Methods. First, we determined the shock success achieved by three different waveforms at varying energies (18–150 J) in 21 mongrel dogs after short-duration ventricular fibrillation. The waveforms tested included the traditional damped sinusoidal waveform, a single pathway biphasic waveform, and a new encircling overlapping multipulse waveform delivered from six electrode pads oriented circumferentially. Second, in 11 swine we compared the efficacy of encircling overlapping multipulse shocks given from six electrode pads and three capacitors versus encircling overlapping shocks given from a device utilizing three electrodes and one capacitor.

Results. In the first experiment, the encircling overlapping waveform performed significantly better than biphasic and damped sinusoidal waveforms at lower energies. The shock success rate of the overlapping waveform (six pads) ranged from 67 ± 4% (at 18–49 J energy) to 99 ± 3% at ≥150 J; at comparable energies biphasic waveform shock success ranged from 26 ± 5% (p < 0.01 vs. encircling overlapping waveforms) to 99 ± 5% (p = NS). Damped sinusoidal waveform shock success ranged from 4 ± 1% (p < 0.01 vs. encircling overlapping waveform) to 73 ± 9% (p = NS). In the second experiment the three electrode pads, one capacitor encircling waveform achieved shock success rates comparable with the six-pad, three-capacitor waveform; at 18–49 J, success rates were 45 ± 15% versus 57 ± 12%, respectively (p = NS). At 100 J, success rates for both were 100%.

Conclusions. We conclude that encircling overlapping multipulse multipathway waveforms facilitate transthoracic defibrillation at low energies. These waveforms can be generated from a device that requires only three electrodes and one capacitor.

Abbreviations and Acronyms
  DS = damped sinusoidal waveform
  EO = encircling overlapping multipulse waveform
  VF = ventricular fibrillation






 