OBJECTIVES: The purpose of this study was to determine the impact of changes in flow on aortic valve area (AVA) as measured by the Gorlin formula and transesophageal echocardiographic (TEE) planimetry. BACKGROUND: The meaning of flow-related changes in AVA calculations using the Gorlin formula in patients with aortic stenosis remains controversial. It has been suggested that flow dependence of the calculated area could be due to a true widening of the orifice as flow increases or to a disproportionate flow dependence of the formula itself. Alternatively, anatomic AVA can be measured by direct planimetry of the valve orifice with TEE. METHODS: Simultaneous measurement of the planimetered and Gorlin valve area was performed intraoperatively under different hemodynamic conditions in 11 patients. Left ventricular and ascending aortic pressures were measured simultaneously after transventricular and aortic punctures. Changes in flow were induced by dobutamine infusion. Using multiplane TEE, AVA was planimetered at the level of the leaflet tips in the short-axis view. RESULTS: Overall, cardiac output, stroke volume and transvalvular volume flow rate ranged from 2.5 to 7.3 liters/min, from 43 to 86 ml and from 102 to 306 ml/min, respectively. During dobutamine infusion, cardiac-output increased by 42% and mean aortic valve gradient by 54%. When minimal flow was compared with maximal flow, the Gorlin area varied from (mean +/- SD) 0.44 +/- 0.12 to 0.60 +/- 0.14 cm2 (p < 0.005). The mean change in Gorlin area under different flow rates was 36 +/- 32%. Despite these changes, there was no significant change in the planimetered area when minimal flow was compared with maximal flow. The mean difference in planimetered area under different flow rates was 0.002 +/- 0.01 cm2 (p = 0.86). CONCLUSIONS: By simultaneous determination of Gorlin formula and TEE planimetry valve areas, we showed that acute changes in transvalvular volume flow substantially altered valve area calculated by the Gorlin formula but did not result in significant alterations of the anatomic valve area in aortic stenosis. These results suggest that the flow-related variation in the Gorlin AVA is due to a disproportionate flow dependence of the formula itself and not a true change in valve area.
This article has been cited by other articles:
![]() |
T. Mihaljevic, M. R. Sayeed, S. C. Stamou, and S. Paul Pathophysiology of Aortic Valve Disease Card. Surg. Adult, January 1, 2008; 3(2008): 825 - 840. [Full Text] |
||||
![]() |
A.-C. Pouleur, J.-B. le Polain de Waroux, A. Pasquet, J.-L. J. Vanoverschelde, and B. L. Gerber Aortic Valve Area Assessment: Multidetector CT Compared with Cine MR Imaging and Transthoracic and Transesophageal Echocardiography Radiology, September 1, 2007; 244(3): 745 - 754. [Abstract] [Full Text] [PDF] |
||||
![]() |
J Bermejo and R Yotti Low-gradient aortic valve stenosis: value and limitations of dobutamine stress testing Heart, March 1, 2007; 93(3): 298 - 302. [Abstract] [Full Text] [PDF] |
||||
![]() |
H. Alkadhi, S. Wildermuth, A. Plass, D. Bettex, B. Baumert, S. Leschka, L. M. Desbiolles, B. Marincek, and T. Boehm Aortic Stenosis: Comparative Evaluation of 16-Detector Row CT and Echocardiography Radiology, July 1, 2006; 240(1): 47 - 55. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. Kadem, R. Rieu, J. G. Dumesnil, L.-G. Durand, and P. Pibarot Flow-Dependent Changes in Doppler-Derived Aortic Valve Effective Orifice Area Are Real and Not Due to Artifact J. Am. Coll. Cardiol., January 3, 2006; 47(1): 131 - 137. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. Kadem, R. Rieu, J. G. Dumesnil, L.-G. Durand, and P. Pibarot Flow-Dependent Changes in Doppler-Derived Aortic Valve Effective Orifice Area Are Real and Not Due to Artifact J. Am. Coll. Cardiol., December 14, 2005; (2005) j.jacc.2005.05.100v1. [Abstract] [Full Text] [PDF] |
||||
![]() |
K. A. Busing, C. Schulte-Sasse, S. Fluchter, T. Suselbeck, K. K. Haase, W. Neff, J. G. Hirsch, M. Borggrefe, and C. Duber Cerebral Infarction: Incidence and Risk Factors after Diagnostic and Interventional Cardiac Catheterization--Prospective Evaluation at Diffusion-weighted MR Imaging Radiology, April 1, 2005; 235(1): 177 - 183. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. S. John, T. Dill, R. R. Brandt, M. Rau, W. Ricken, G. Bachmann, and C. W. Hamm Magnetic resonance to assess the aortic valve area in aortic stenosis: How does it compare to current diagnostic standards? J. Am. Coll. Cardiol., August 6, 2003; 42(3): 519 - 526. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. Das, H. Rimington, N. Smeeton, and J. Chambers Determinants of symptoms and exercise capacity in aortic stenosis: a comparison of resting haemodynamics and valve compliance during dobutamine stress Eur. Heart J., July 1, 2003; 24(13): 1254 - 1263. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. E. Gray and A. C. Perrino Jr. Hemodynamic-Induced Changes in Aortic Valve Area: Implications for Doppler Cardiac Output Determinations Anesth. Analg., March 1, 2001; 92(3): 584 - 589. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. Shandas, J. Kwon, and L. Valdes-Cruz A Method for Determining the Reference Effective Flow Areas for Mechanical Heart Valve Prostheses : In Vitro Validation Studies Circulation, April 25, 2000; 101(16): 1953 - 1959. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Arsenault, N. Masani, G. Magni, J. Yao, L. Deras, and N. Pandian Variation of anatomic valve area during ejection in patients with valvular aortic stenosis evaluated by two-dimensional echocardiographic planimetry: comparison with traditional Doppler data J. Am. Coll. Cardiol., December 1, 1998; 32(7): 1931 - 1937. [Abstract] [Full Text] [PDF] |
||||
| HOME | SUBSCRIPTIONS | CURRENT ISSUE | PAST ISSUES | CARDIOSOURCE | SEARCH | HELP | FEEDBACK |