Feasibility of atrial fibrillation detection and use of a preceding synchronization interval as a criterion for shock delivery in humans with atrial fibrillation
JS Sra,
C Maglio,
A Dhala,
Z Blanck,
M Biehl,
S Deshpande,
ET Keelan,
MR Jazayeri,
and
M Akhtar
Electrophysiology Laboratory, St. Luke's Medical Center, University of Wisconsin-Milwaukee Clinical Campus, Milwaukee, USA.
OBJECTIVES: This study assessed the feasibility of detecting atrial fibrillation (AF) and delivery of appropriately timed R wave shocks using an implantable atrial defibrillator. BACKGROUND: For atrial defibrillation therapy to be feasible in an implantable form, AF must be detected in a specific fashion, and the risk of ventricular proarrhythmia should be minimized. METHODS: Eleven patients with AF underwent testing with an implantable atrial defibrillator (METRIX 3000 Automatic Atrial Defibrillator, InControl, Inc.). Wideband electrograms (EGMs) were recorded from the right ventricular (RV) bipolar catheter and from the multipolar catheters located in the right atrium (RA) and coronary sinus (CS). Atrial fibrillation detection was performed using two serial algorithms-quiet interval analysis and baseline crossing analysis-that detect atrial activity on the RA-CS channel. Ventricular sensing using a minimal preceding synchronization interval of 500 ms as a criterion for synchronous shock delivery was performed from filtered RV and RV-CS EGMs. RESULTS: The AF detection algorithms were applied to 53 AF data segments and 18 normal sinus rhythm data segments. Atrial fibrillation was detected appropriately in 49 instances, and the specificity for detecting AF and normal sinus rhythm was 100%. Synchronization criterion efficacy was assessed by delivering shock markers and shocks. Of the 2,025 R waves processed, 557 (27.5%) were marked as suitable for shock delivery. In addition, 69 therapeutic and 11 test shocks were delivered during AF. All shock markers and shocks were delivered synchronously with the R wave, and the synchronization criterion was never violated. CONCLUSIONS: Atrial fibrillation can be detected in a specific fashion using the RA-CS lead configuration and serial detection algorithms for atrial sensing. The delivery of properly timed shocks is feasible and should minimize the risk of ventricular proarrhythmia.
This article has been cited by other articles:

|
 |

|
 |
 
C. M. Tracy, M. Akhtar, J. P. DiMarco, D. L. Packer, H. H. Weitz, M. A. Creager, D. R. Holmes Jr, G. Merli, G. P. Rodgers, C. M. Tracy, et al.
American College of Cardiology/American Heart Association 2006 Update of the Clinical Competence Statement on Invasive Electrophysiology Studies, Catheter Ablation, and Cardioversion: A Report of the American College of Cardiology/American Heart Association/American College of Physicians Task Force on Clinical Competence and Training Developed in Collaboration With the Heart Rhythm Society
J. Am. Coll. Cardiol.,
October 3, 2006;
48(7):
1503 - 1517.
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
C. M. Tracy, M. Akhtar, J. P. DiMarco, D. L. Packer, H. H. Weitz, W. L. Winters, J. L. Achord, A. W. Boone, J. W. Hirshfeld Jr, B. H. Lorell, et al.
American College of Cardiology/American Heart Association Clinical Competence Statement on invasive electrophysiology studies, catheter ablation, and cardioversion: A report of the american college of cardiology/american heart association/american college of physicians-american society of internal medicine task force on clinical competence
J. Am. Coll. Cardiol.,
November 1, 2000;
36(5):
1725 - 1736.
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
C. M. Tracy, M. Akhtar, J. P. DiMarco, D. L. Packer, H. H. Weitz, W. L. Winters, J. L. Achord, A. W. Boone, J. W. Hirshfeld Jr, B. H. Lorell, et al.
American College of Cardiology/American Heart Association Clinical Competence Statement on Invasive Electrophysiology Studies, Catheter Ablation, and Cardioversion : A Report of the American College of Cardiology/American Heart Association/American College of Physicians-American Society of Internal Medicine Task Force on Clinical Competence
Circulation,
October 31, 2000;
102(18):
2309 - 2320.
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
H.-F. Tse, C.-P. Lau, B. M. Yomtov, and G. M. Ayers
Implantable atrial defibrillator with a single-pass dual-electrode lead
J. Am. Coll. Cardiol.,
June 1, 1999;
33(7):
1974 - 1980.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
H.-F. Tse, C.-P. Lau, J. S. Sra, H. J. G. M. Crijns, N. Edvardsson, S. Kacet, and D. G. Wyse
Atrial Fibrillation Detection and R-Wave Synchronization by Metrix Implantable Atrial Defibrillator : Implications for Long-Term Efficacy and Safety
Circulation,
March 23, 1999;
99(11):
1446 - 1451.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
H. J. J. Wellens, C.-P. Lau, B. Luderitz, M. Akhtar, A. L. Waldo, A. J. Camm, C. Timmermans, H.-F. Tse, W. Jung, L. Jordaens, et al.
Atrioverter: An Implantable Device for the Treatment of Atrial Fibrillation
Circulation,
October 20, 1998;
98(16):
1651 - 1656.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
A. C. Skanes, R. A. Gray, C. L. Zuur, and J. Jalife
Effects of Postshock Atrial Pacing on Atrial Defibrillation Outcome in the Isolated Sheep Heart
Circulation,
July 7, 1998;
98(1):
64 - 72.
[Abstract]
[Full Text]
[PDF]
|
 |
|
|