Reasons for prolonged or failed attempts at radiofrequency catheter ablation of accessory pathways
F Morady,
A Strickberger,
KC Man,
E Daoud,
M Niebauer,
R Goyal,
M Harvey,
and
F Bogun
Division of Cardiology, Department of Internal Medicine, University of Michigan Medical Center, Ann Arbor, Michigan 48109-0022, USA.
OBJECTIVES: The purpose of this study was to categorize the reasons for a prolonged or failed procedure in a series of patients undergoing catheter ablation of an accessory pathway. BACKGROUND: Radiofrequency ablation of accessory pathways at times requires a lengthy procedure or a second ablation session, or both, and not prior studies have systematically investigated the reasons for this. METHODS: In a consecutive series of 619 patients undergoing catheter ablation of an accessory pathway, the mean ablation time +/- SD was 68 +/- 64 min. The subjects of this study were 14 patients who had an ablation time >2 SD greater than the mean (>196 min) and 51 patients who required a second ablation session for a successful outcome. The accessory pathway in the 65 patients in this study was located in the right free wall in 19 patients (29%), septum in 14 (22%) and left free wall in 32 (49%). RESULTS: The primary reasons for a lengthy or failed ablation attempt were 1) inability to position the ablation catheter at the effective target site (16 patients, 25%); 2) instability of the ablation catheter or inadequate tissue contact at the target site, or both (15 patients, 23%); 3) mapping error due to an oblique course of the accessory pathway (7 patients, 11%); 4) failure to recognize a posteroseptal accessory pathway as being left-sided instead of right-sided (4 patients, 6%); 5) other errors in accessory pathway localization (6 patients, 9%); 6) epicardial location of the accessory pathway (5 patients, 8%); 7) recurrent atrial fibrillation (2 patients, 3%); 8) occurrence of a complication (2 patients, 3%); 9) unusual right-sided accessory pathway that inserted in the anterior right ventricle, 2 cm away from the lateral tricuspid annulus (1 patient, 1.5%); and 10) unexplained factors (7 patients, 11%). The most common effective strategies employed to achieve a successful outcome in these patients were 1) substitution of a more experienced operator; 2) use of ablation catheters of varying configurations; 3) switching from a retrograde aortic to a trans-septal approach; 4) switching from an inferior to a superior vena caval approach; 5) use of a 60-cm guiding sheath; 6) detailed mapping of the atrial or ventricular insertion of the accessory pathway; and 7) searching within the coronary sinus for a presumed accessory pathway potential. CONCLUSIONS: A lengthy or failed attempt at catheter ablation of an accessory pathway may be due to a variety of reasons, the most common of which are problems related to some aspect of catheter manipulation and errors in accessory pathway localization. Knowledge of the most common reasons for a lengthy or ineffective procedure may facilitate successful outcome of accessory pathway ablation.
This article has been cited by other articles:

|
 |

|
 |
 
F. Sacher, M. Wright, U. B. Tedrow, M. D. O'Neill, P. Jais, M. Hocini, R. Macdonald, D. W. Davies, P. Kanagaratnam, N. Derval, et al.
Wolff-Parkinson-White ablation after a prior failure: a 7-year multicentre experience
Europace,
June 1, 2010;
12(6):
835 - 841.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
E. Wissner, F. Ouyang, and K. H. Kuck
Examining the causes of ablation failure in the Wolff-Parkinson-White syndrome
Europace,
June 1, 2010;
12(6):
772 - 773.
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
A. Chiribiri, S. Kelle, U. Kohler, L. F. Tops, B. Schnackenburg, R. Bonamini, J. J. Bax, E. Fleck, and E. Nagel
Magnetic resonance cardiac vein imaging: relation to mitral valve annulus and left circumflex coronary artery.
J. Am. Coll. Cardiol. Img.,
November 1, 2008;
1(6):
729 - 738.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
H. Nakagawa and W. M. Jackman
Catheter Ablation of Paroxysmal Supraventricular Tachycardia
Circulation,
November 20, 2007;
116(21):
2465 - 2478.
[Full Text]
[PDF]
|
 |
|
|