J Am Coll Cardiol, 1996; 27:161-172 © 1996 by the American College of Cardiology Foundation |
Cardiac Ultrasound Laboratory, Massachusetts General Hospital, Boston 02114, USA.
OBJECTIVES. The purpose of this study was to develop a rational and objective method for selecting a region in the proximal flow field where the hemispheric formula for calculating regurgitant flow rates by the flow convergence technique is most accurate. BACKGROUND. A major obstacle to clinical implementation of the proximal flow convergence method is that it assumes hemispheric isovelocity contours throughout the Doppler color flow map, whereas contour shape depends critically on location in the flow field. METHODS. Twenty mitral regurgitant flow rate stages were produced in six dogs by implanting grommet orifices into the anterior mitral leaflet and varying driving pressures so that actual peak flow rate could be determined from the known effective regurgitant orifice times the orifice velocity. Because plotting flow rate calculated by using a hemispheric formula versus alias velocities produces underestimation near the orifice and overestimation far from it, this plot was fitted to a polynomial function to allow identification of an inflection point within a relatively flat intermediate zone, where factors causing overestimation and underestimation are expected to be unimportant or balanced. The accuracy of flow rate calculation by the inflection point was compared with unselective and selective averaging techniques. Clinical relevance, initial feasibility and correlation with an independent measure were tested in 13 consecutive patients with mitral regurgitation who underwent cardiac catheterization. RESULTS. 1) The accuracy of single-point calculations was improved by selecting points in the flat portion of the curve (y = 1.15x - 3.34, r = 0.87, SEE = 22.1 ml/s vs. y = 1.34x - 1.99, r = 0.71, SEE = 45.6 ml/s, p < 0.01). 2) Selective averaging of points in the flat portion of the curve further improved accuracy and decreased scatter compared with unselective averaging (y = 1.08x + 4.8, r = 0.96, SEE = 11.6 ml/s vs. y = 1.30x + 0.6, r = 0.90, SEE = 20.9 ml/s, p < 0.01). 3) The proposed algorithm for mathematically identifying the inflection point provided the best results (y = 0.96x + 4.5, r = 0.96, SEE = 9.9 ml/s), with a mean error of 1.6 +/- 9.7 ml/s vs. 11.4 +/- 11.7 ml/s for selective averaging (p < 0.01). In patients, the proposed algorithm identified an inflection point at which calculated regurgitant volume agreed best with invasive measurements (y = 1.1x - 0.61, r = 0.93, SEE = 17 ml). CONCLUSIONS. The accuracy of the proximal flow convergence method can be significantly improved by analyzing the flow field mathematically to identify the optimal isovelocity zone before using the hemispheric formula to calculate regurgitant flow rates. Because the proposed algorithm is objective, operator independent and, thus, suitable for automatization, it could provide the clinician with a powerful quantitative tool to assess valvular regurgitation.
This article has been cited by other articles:
![]() |
T. Buck, B. Plicht, P. Hunold, R. A. Mucci, R. Erbel, and R. A. Levine Broad-beam spectral Doppler sonification of the vena contracta using matrix-array technology: A new solution for semi-automated quantification of mitral regurgitant flow volume and orifice area J. Am. Coll. Cardiol., March 1, 2005; 45(5): 770 - 779. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. A. Levine and J. Hung Ischemic mitral regurgitation, the dynamic lesion: clues to the cure J. Am. Coll. Cardiol., December 3, 2003; 42(11): 1929 - 1932. [Full Text] [PDF] |
||||
![]() |
W. A. Zoghbi, M. Enriquez-Sarano, E. Foster, P. A. Grayburn, C. D. Kraft, R. A. Levine, P. Nihoyannopoulos, C. M. Otto, M. A. Quinones, H. Rakowski, et al. American Society of Echocardiography: recommendations for evaluation of the severity of native valvular regurgitation with two-dimensional and Doppler echocardiography : A report from the American Society of Echocardiography's Nomenclature and Standards Committee and The Task Force on Valvular Regurgitation, developed in conjunction with the American College of Cardiology Echocardiography Committee, The Cardiac Imaging Committee, Council on Clinical Cardiology, The American Heart Association, and the European Society of Cardiology Working Group on Echocardiography, represented by: Eur J Echocardiogr, December 1, 2003; 4(4): 237 - 261. [Full Text] [PDF] |
||||
![]() |
J. D Thomas DOPPLER ECHOCARDIOGRAPHIC ASSESSMENT OF VALVAR REGURGITATION Heart, December 1, 2002; 88(6): 651 - 657. [Full Text] [PDF] |
||||
![]() |
T. Buck, R. A. Mucci, J. L. Guerrero, G. Holmvang, M. D. Handschumacher, and R. A. Levine The Power-Velocity Integral at the Vena Contracta : A New Method for Direct Quantification of Regurgitant Volume Flow Circulation, August 29, 2000; 102(9): 1053 - 1061. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Hung, Y. Otsuji, M. D. Handschumacher, E. Schwammenthal, and R. A. Levine Mechanism of dynamic regurgitant orifice area variation in functional mitral regurgitation: Physiologic insights from the proximal flow convergence technique J. Am. Coll. Cardiol., February 1, 1999; 33(2): 538 - 545. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. Shiota, M. Jones, A. Delabays, X. Li, I. Yamada, M. Ishii, P. Acar, S. Holcomb, N. G. Pandian, and D. J. Sahn Direct Measurement of Three-dimensionally Reconstructed Flow Convergence Surface Area and Regurgitant Flow in Aortic Regurgitation : In Vitro and Chronic Animal Model Studies Circulation, November 18, 1997; 96(10): 3687 - 3695. [Abstract] [Full Text] |
||||
| HOME | SUBSCRIPTIONS | CURRENT ISSUE | PAST ISSUES | CARDIOSOURCE | SEARCH | HELP | FEEDBACK |