Estimation of regurgitant flow volume based on centerline velocity/distance profiles using digital color M-Q Doppler: application to orifices of different shapes
T Shiota,
D Teien,
YB Deng,
S Ge,
R Shandas,
S Holcomb,
and
DJ Sahn
Clinical Care Center for Congenital Heart Disease, Oregon Health Sciences University, Portland 97201.
OBJECTIVES. In this study we investigated the centerline velocity profile method for flow computation as applied to noncircular, as well as circular, orifices using digital color flow data. BACKGROUND. Recently it has been suggested that flow volume through an orifice can be estimated more accurately by computing the axial "centerline" flow velocity/distance profile proximal to the orifice. METHODS. A total of seven different orifices were mounted in a constant-flow model: four circular orifices, two rectangular orifices with a major/minor axis ratio of 4:1 and 8:1 and an ovoid orifice having a major/minor axis ratio of 2:1. Three different flow rates were examined (1.68, 3.48 and 6.48 liters/min). Digital measurements of flow velocity at discrete positions along the centerline progressing toward the orifice were analyzed to yield complete flow velocity profiles for each orifice at each flow rate. RESULTS. A clear separation of the flow profiles for the three different flow rates was observed independent of orifice size for all of the circular orifices. The velocity/distance acceleration curves showed highly significant correlations using multiplicative regression fits (y = ax-b, r = 0.94 to 0.99, all p < 0.0001). An equation for quantitatively correlating the a and b coefficients from the multiplicative regression fits with flow rates was derived from stepwise regression analysis: Flow rate = 23a + 3.3b - 1.5 (r = 0.97, p < 0.0001, SEE 0.46 liter/min). CONCLUSIONS. In view of the various sizes and shapes encountered clinically for regurgitant orifices, the simplicity of this method for the estimation of the severity of regurgitant lesions might be of importance for clinical applications of this method.
This article has been cited by other articles:

|
 |

|
 |
 
M. S. Firstenberg, P. M. Vandervoort, N. L. Greenberg, N. G. Smedira, P. M. McCarthy, M. J. Garcia, and J. D. Thomas
Noninvasive estimation of transmitral pressure drop across the normal mitral valve in humans: importance of convective and inertial forces during left ventricular filling
J. Am. Coll. Cardiol.,
November 15, 2000;
36(6):
1942 - 1949.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
T. Shiota, M. Jones, A. Delabays, X. Li, I. Yamada, M. Ishii, P. Acar, S. Holcomb, N. G. Pandian, and D. J. Sahn
Direct Measurement of Three-dimensionally Reconstructed Flow Convergence Surface Area and Regurgitant Flow in Aortic Regurgitation : In Vitro and Chronic Animal Model Studies
Circulation,
November 18, 1997;
96(10):
3687 - 3695.
[Abstract]
[Full Text]
|
 |
|

|
 |

|
 |
 
T. Shiota, M. Jones, I. Yamada, R. S. Heinrich, M. Ishii, B. Sinclair, S. Holcomb, A. P. Yoganathan, and D. J. Sahn
Effective Regurgitant Orifice Area by the Color Doppler Flow Convergence Method for Evaluating the Severity of Chronic Aortic Regurgitation : An Animal Study
Circulation,
February 1, 1996;
93(3):
594 - 602.
[Abstract]
[Full Text]
|
 |
|
|