Three-dimensional echocardiographic measurement of left ventricular volume in vitro: comparison with two-dimensional echocardiography and cineventriculography
PM Sapin,
KD Schroeder,
MD Smith,
AN DeMaria,
and
DL King
Division of Cardiology, University of Kentucky Medical Center, Lexington 40536.
OBJECTIVES. This study was designed to compare three-dimensional echocardiography, two-dimensional echocardiography and cineventriculography for the purpose of measuring left ventricular volume in vitro. BACKGROUND. Three-dimensional echocardiographic systems have been shown to be highly accurate in measuring the volumes of balloon phantoms. However, three-dimensional techniques have not been compared with standard two-dimensional echocardiography in vitro or with cineventriculography, the clinical standard for left ventricular volume measurement. METHODS. Excised porcine hearts were prepared with an internal latex sheath that could be filled and maintained with a known ("true") volume of liquid. Each heart was then imaged by cineventriculography, standard two-dimensional echocardiography and three-dimensional echocardiography. Left ventricular volumes were calculated from 15 hearts at 25 volumes ranging from 50 to 280 ml by the following methods: 1) biplane cineventriculography using the area-length method; 2) two-dimensional echocardiography by the apical biplane method using a summation of discs algorithm in 15 cases and the single-plane, four-chamber method using a summation of discs algorithm in 10 cases; and 3) three-dimensional echocardiography using a polyhedral surface reconstruction volume computation algorithm based on multiple nonparallel, nonevenly spaced short-axis cross sections. RESULTS. Results were compared with true volume, and a nonparametric analysis of variance was performed. Both measurement bias (systematic error) and imprecision (random error) were assessed. All methods tended to underestimate the true volume (two-dimensional echocardiography -6.1 +/- 17.6%, three-dimensional echocardiography -4.7 +/- 5.0% and biplane cineventriculography -3.9 +/- 8.2%), although differences were not significant. Although there was a significant correlation between the magnitude of measurement bias and the size of the volume being measured for two-dimensional echocardiography and cineventriculography, the bias of three-dimensional echocardiography was fairly constant over the range of volumes. When bias was accounted for, two-dimensional echocardiography was significantly less precise than cineventriculography and three-dimensional echocardiography in terms of percent error (15.3 +/- 11.9%, 5.6 +/- 5.7% and 3.9 +/- 3.4%, respectively). CONCLUSIONS. Three-dimensional echocardiography using a polyhedral surface reconstruction algorithm for volume computation provides accuracy comparable to that of biplane cineventriculography in this in vitro model. Standard two-dimensional echocardiographic volume computation is significantly less accurate than the other two methods.
This article has been cited by other articles:

|
 |

|
 |
 
A. Lyshchik, S. B. Hobbs, A. C. Fleischer, D. Khabele, D.-S. Son, J. C. Gore, and R. R. Price
Ovarian Volume Measurements in Mice With High-Resolution Ultrasonography
J. Ultrasound Med.,
October 1, 2007;
26(10):
1419 - 1425.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
R. C. Houck, J. E. Cooke, and E. A. Gill
Live 3D Echocardiography: A Replacement for Traditional 2D Echocardiography?
Am. J. Roentgenol.,
October 1, 2006;
187(4):
1092 - 1106.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
J Kawai, K Tanabe, C.-L Wang, T Tani, T Yagi, H Shiotani, and S Morioka
Comparison of left atrial size by freehand scanning three-dimensional echocardiography and two-dimensional echocardiography
Eur Heart J Cardiovasc Imaging,
January 1, 2004;
5(1):
18 - 24.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
L Sugeng, L Weinert, and R M Lang
Left ventricular assessment using real time three dimensional echocardiography
Heart,
November 1, 2003;
89(90003):
iii29 - 36.
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
D. Gilon, E. G. Cape, M. D. Handschumacher, J.-K. Song, J. Solheim, M. VanAuker, M. E. E. King, and R. A. Levine
Effect of three-dimensional valve shape on the hemodynamics of aortic stenosis: Three-dimensional echocardiographic stereolithography and patient studies
J. Am. Coll. Cardiol.,
October 16, 2002;
40(8):
1479 - 1486.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
M. Ahmad, T. Xie, M. McCulloch, G. Abreo, and M. Runge
Real-time three-dimensional dobutamine stress echocardiography in assessment of ischemia: comparison with two-dimensional dobutamine stress echocardiography
J. Am. Coll. Cardiol.,
April 1, 2001;
37(5):
1303 - 1309.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
N Sutaria, D Northridge, N Masani, and N Pandian
Three dimensional echocardiography for the assessment of mitral valve disease
Heart,
November 1, 2000;
84(90002):
7i - 10.
[Full Text]
|
 |
|

|
 |

|
 |
 
R. De Simone, G. Glombitza, C. F. Vahl, J.o. Albers, H. P. Meinzer, and S. Hagl
Three-dimensional color doppler: a clinical study in patients with mitral regurgitation
J. Am. Coll. Cardiol.,
May 1, 1999;
33(6):
1646 - 1654.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
O. H. Gilja, T. Hausken, A. Berstad, and S. Odegaard
Measurements of organ volume by ultrasonography
Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine,
March 1, 1999;
213(3):
247 - 259.
[Abstract]
[PDF]
|
 |
|

|
 |

|
 |
 
T. Buck, P. Hunold, K. U. Wentz, W. Tkalec, H. J. Nesser, and R. Erbel
Tomographic Three-dimensional Echocardiographic Determination of Chamber Size and Systolic Function in Patients With Left Ventricular Aneurysm : Comparison to Magnetic Resonance Imaging, Cineventriculography, and Two-dimensional Echocardiography
Circulation,
December 16, 1997;
96(12):
4286 - 4297.
[Abstract]
[Full Text]
|
 |
|

|
 |

|
 |
 
L. Gepstein, G. Hayam, S. Shpun, and S. A. Ben-Haim
Hemodynamic Evaluation of the Heart With a Nonfluoroscopic Electromechanical Mapping Technique
Circulation,
November 18, 1997;
96(10):
3672 - 3680.
[Abstract]
[Full Text]
|
 |
|

|
 |

|
 |
 
D. Gilon, E. G. Cape, M. D. Handschumacher, L. Jiang, C. Sears, J. Solheim, E. Morris, J. T. Strobel, S. M. Miller-Jones, A. E. Weyman, et al.
Insights From Three-dimensional Echocardiographic Laser Stereolithography: Effect of Leaflet Funnel Geometry on the Coefficient of Orifice Contraction, Pressure Loss, and the Gorlin Formula in Mitral Stenosis
Circulation,
August 1, 1996;
94(3):
452 - 459.
[Abstract]
[Full Text]
|
 |
|

|
 |

|
 |
 
A. S. Gopal, Z. Shen, P. M. Sapin, A. M. Keller, M. J. Schnellbaecher, D. W. Leibowitz, O. O. Akinboboye, R. A. Rodney, D. K. Blood, and D. L. King
Assessment of Cardiac Function by Three-dimensional Echocardiography Compared With Conventional Noninvasive Methods
Circulation,
August 15, 1995;
92(4):
842 - 853.
[Abstract]
[Full Text]
|
 |
|

|
 |

|
 |
 
L. Jiang, J. A. Vazquez de Prada, M. D. Handschumacher, C. Vuille, J. L. Guererro, M. H. Picard, J. T. Joziatis, J. T. Fallon, A. E. Weyman, and R. A. Levine
Quantitative Three-Dimensional Reconstruction of Aneurysmal Left Ventricles : In Vitro and In Vivo Validation
Circulation,
January 1, 1995;
91(1):
222 - 230.
[Abstract]
[Full Text]
|
 |
|
|