J Am Coll Cardiol, 1993; 22:1367-1372
© 1993 by the American College of Cardiology Foundation
This Article
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Avitall, B
Right arrow Articles by Hare, J
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Avitall, B
Right arrow Articles by Hare, J

Repeated use of ablation catheters: a prospective study

B Avitall, M Khan, D Krum, M Jazayeri, and J Hare

Electrophysiology Laboratory, University of Wisconsin-Milwaukee Clinical Campus, Sinai Samaritan Medical Center 53201.

OBJECTIVES. The objective of this study was to provide insight into the time course of electrical, physical and mechanical changes in ablation catheters after each use that may affect the safety and efficacy of the ablation procedure. BACKGROUND. An increasing number of institutions are reusing deflectable ablation catheters. At present, there are no data concerning the safety of reusing ablation catheters. METHODS. Over a period of 1 year, 69 Webster/Mansfield deflectable catheters used in 336 ablation procedures were prospectively studied. An additional 18 new catheters were tested after multiple sterilizations only. The catheters were evaluated for electrical and physical integrity and mechanical capabilities. These include deflection at room temperature and 37 degrees C, shaft compression and buckling during deflection, tip craters, torquing ability, glue separation and tip attachment using a stereoscope at x30 magnification and electrical resistance for each electrode. After each use, the catheters were gas-sterilized with ethylene oxide. RESULTS. The most common reasons for catheter rejection were tip electrode glue separation after 4.3 +/- 4.3 uses and loss of deflection after 5.0 +/- 3.3 uses. Electrical discontinuity between the catheter handle and electrodes was observed after 10.0 +/- 3.7 uses. There was no significant decrease in catheter torquing ability with repeated use. In this study the total estimated savings was $128,133, which includes the cost of catheter reprocessing. The reuse of Webster/Mansfield ablation catheters has not resulted in any major catheter failure or any major adverse clinical complications. CONCLUSIONS. On the basis of these observations, we believe that the Webster/Mansfield catheter can be reused an average of five times. It is strongly recommended that after each use catheters be carefully examined under appropriate magnification (x30) and that special attention be given to the ablation tip electrode. The catheters should also be tested for deflection and electrical integrity.


This article has been cited by other articles:


Home page
J Am Coll CardiolHome page
M. N. Bathina, S. Mickelsen, C. Brooks, J. Jaramillo, T. Hepton, and F. M. Kusumoto
Safety and efficacy of hydrogen peroxide plasma sterilization for repeated use of electrophysiology catheters
J. Am. Coll. Cardiol., November 1, 1998; 32(5): 1384 - 1388.
[Abstract] [Full Text] [PDF]




HOME SUBSCRIPTIONS CURRENT ISSUE PAST ISSUES CARDIOSOURCE SEARCH HELP FEEDBACK
Copyright © 1993 by the American College of Cardiology Foundation.