cardiology careers collections past issues search home
     

J Am Coll Cardiol, 1993; 21:885-894
© 1993 by the American College of Cardiology Foundation
This Article
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Wietholt, D
Right arrow Articles by Breithardt, G
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Wietholt, D
Right arrow Articles by Breithardt, G

Clinical experience with antitachycardia pacing and improved detection algorithms in a new implantable cardioverter-defibrillator

D Wietholt, M Block, F Isbruch, D Bocker, M Borggrefe, M Shenasa, and G Breithardt

Hospital of the Westfalische-Wilhelms University of Munster, Department of Cardiology and Angiology, Germany.

OBJECTIVES. This study was conducted to assess the effectiveness of antitachycardia pacing modes and detection algorithms in patients with a new third-generation implantable cardioverter-defibrillator. METHODS. Twenty-three of 42 consecutive patients had coronary artery disease, 14 had dilated cardiomyopathy, 2 had prior valve replacement and 3 had arrhythmogenic right ventricular dysplasia. The mean ejection fraction was 41 +/- 14%; there were 31 men (74%) and 11 women, with a mean age of 53 years. On the basis of preoperative and postoperative electrophysiologic studies, in 28 patients antitachycardia pacing was postoperatively programmed randomly as "burst" (66%) or autodecremental "ramp" (34%) stimulation with a first coupling interval of 81% of tachycardia cycle length and up to 8 sequences with 3 to 10 stimuli. RESULTS. During a follow-up interval of 6.3 +/- 2.2 months, 15 patients were treated by antitachycardia pacing for a median of 6 (range 1 to 59) hemodynamically stable ventricular tachycardias (175 +/- 12 beats/min). In 5 patients, 22 ventricular tachycardias (9%) were not terminated by antitachycardia pacing but by cardioversion. Seven (3%) of these episodes accelerated (> 50 ms) during antitachycardia pacing. Syncope did not occur during these episodes. In seven patients initial antitachycardia pacing in cases of supraventricular tachycardias delayed charging and redetection prevented inappropriate discharges. Additional detection algorithms were programmed only after inappropriate therapy. The sudden "onset" and "sustained rate duration" criteria were programmed in three patients and the cycle length "stability" criteria in six patients, respectively. After activation of these detection algorithms only two of the seven patients had further inappropriate device discharges. CONCLUSIONS. Thus, antitachycardia pacing by this implantable cardioverter-defibrillator effectively and appropriately terminated 91% of hemodynamically stable ventricular tachycardias. Inappropriate device discharges were prevented in some patients by antitachycardia pacing and additional detection algorithms.


This article has been cited by other articles:


Home page
EuropaceHome page
G. Boriani, E. Occhetta, S. Cesario, S. Grossi, M. Marconi, G. Speca, P. Silvestri, M. Biffi, M. Bortnik, C. Martignani, et al.
Contribution of morphology discrimination algorithm for improving rhythm discrimination in slow and fast ventricular tachycardia zones in dual-chamber implantable cardioverter-defibrillators
Europace, August 1, 2008; 10(8): 918 - 925.
[Abstract] [Full Text] [PDF]


Home page
JAMAHome page
A. K. Gehi, D. Mehta, and J. A. Gomes
Evaluation and Management of Patients After Implantable Cardioverter-Defibrillator Shock
JAMA, December 20, 2006; 296(23): 2839 - 2847.
[Abstract] [Full Text] [PDF]


Home page
HeartHome page
J M Morgan
The MADIT II and COMPANION studies: will they affect uptake of device treatment?
Heart, March 1, 2004; 90(3): 243 - 245.
[Full Text] [PDF]


Home page
J Am Coll CardiolHome page
L. B. Mitchell, E. A. Pineda, J. L. Titus, P. M. Bartosch, and D. G. Benditt
Sudden death in patients with implantable cardioverter defibrillators: The importance of post-shock electromechanical dissociation
J. Am. Coll. Cardiol., April 17, 2002; 39(8): 1323 - 1328.
[Abstract] [Full Text] [PDF]


Home page
CirculationHome page
M. S. Wathen, M. O. Sweeney, P. J. DeGroot, A. J. Stark, J. L. Koehler, M. B. Chisner, C. Machado, and W. O. Adkisson
Shock Reduction Using Antitachycardia Pacing for Spontaneous Rapid Ventricular Tachycardia in Patients With Coronary Artery Disease
Circulation, August 14, 2001; 104(7): 796 - 801.
[Abstract] [Full Text] [PDF]


Home page
Eur Heart JHome page
C.M Pratt, A.J Camm, J.T Bigger Jr, G Breithardt, R.W.F Campbell, A.E Epstein, L.J Kappenberger, K.-H Kuck, S Pocock, S Saksena, et al.
Evaluation of antiarrhythmic drug efficacy in patients with an ICD. Unlimited potential or replete with complexity and problems?
Eur. Heart J., November 1, 1999; 20(21): 1538 - 1552.
[Abstract] [PDF]


Home page
CirculationHome page
A. Arenal, J. Almendral, J. Villacastin, R. Morris, E. Castellanos, and J. L. Delcan
First Postpacing Interval Variability During Right Ventricular Stimulation : A Single Algorithm for the Differential Diagnosis of Regular Tachycardias
Circulation, August 18, 1998; 98(7): 671 - 677.
[Abstract] [Full Text] [PDF]


Home page
CirculationHome page
A. Schaumann, F. v. z. Muhlen, B. Herse, B.-D. Gonska, and H. Kreuzer
Empirical Versus Tested Antitachycardia Pacing in Implantable Cardioverter Defibrillators : A Prospective Study Including 200 Patients
Circulation, January 13, 1998; 97(1): 66 - 74.
[Abstract] [Full Text] [PDF]


Home page
CirculationHome page
G. J. Fahy, E. B. Sgarbossa, P. J. Tchou, and S. L. Pinski
Hospital Readmission in Patients Treated With Tiered-Therapy Implantable Defibrillators
Circulation, September 15, 1996; 94(6): 1350 - 1356.
[Abstract] [Full Text]


Home page
CirculationHome page
D. P. Zipes and D. Roberts
Results of the International Study of the Implantable Pacemaker Cardioverter-Defibrillator : A Comparison of Epicardial and Endocardial Lead Systems
Circulation, July 1, 1995; 92(1): 59 - 65.
[Abstract] [Full Text]



 
  cardiology careers collections past issues search home