Immediate sealing of arterial puncture sites after cardiac catheterization and coronary angioplasty using a biodegradable collagen plug: results of an international registry
SM Ernst,
RM Tjonjoegin,
R Schrader,
M Kaltenbach,
U Sigwart,
TA Sanborn,
and
HW Plokker
Department of Cardiology of St. Antonius Hospital, Nieuwegein, The Netherlands.
OBJECTIVES. The aim of this study was to evaluate the safety and efficacy of a biodegradable collagen plug that has been developed to reduce the arterial compression time required to achieve hemostasis at the arterial puncture site after diagnostic and interventional coronary procedures. BACKGROUND. After diagnostic and interventional coronary catheterization procedures, local arterial compression is required to achieve hemostasis and complications may ensue, especially in patients on full anticoagulation. METHODS. Between March 1991 and July 1991, 252 patients admitted for routine coronary angiography or angioplasty to four large hospitals received such a hemostatic device immediately after the procedure. Hemostasis was achieved with collagen in 87% of patients after a mean compression time of 4.8 min. Time to hemostasis was independent of the heparin load. A total of 54 hematomas (21%) was reported; all but 2 resolved without additional treatment. Two patients had a severe hematoma, requiring blood transfusion, and two patients required surgery to repair a pseudoaneurysm. During a follow-up period of 4 weeks no severe late complications were reported. CONCLUSIONS. We conclude that the collagen plug appears to be a safe device to achieve hemostasis at the arterial puncture site, independent of anticoagulation.
This article has been cited by other articles:

|
 |

|
 |
 
K. Hamraoui, S. M. P. G. Ernst, P. F. H. M. van Dessel, J. C. Kelder, J. M. ten Berg, M. J. Suttorp, W. Jaarsma, and T. H. W. Plokker
Efficacy and safety of percutaneous treatment of iatrogenic femoral artery pseudoaneurysm by biodegradable collagen injection
J. Am. Coll. Cardiol.,
April 17, 2002;
39(8):
1297 - 1304.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
G. Dangas, R. Mehran, S. Kokolis, D. Feldman, L. F. Satler, A. D. Pichard, K. M. Kent, A. J. Lansky, G. W. Stone, and M. B. Leon
Vascular complications after percutaneous coronary interventions following hemostasis with manual compression versus arteriotomy closure devices
J. Am. Coll. Cardiol.,
September 1, 2001;
38(3):
638 - 641.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
M. Rabah, D. Mason, D. W. M. Muller, R. Hundley, A. D. Kugelmass, B. Weiner, L. Cannon, W. W. O'Neill, and R. D. Safian
Heparin after percutaneous intervention (HAPI): a prospective multicenter randomized trial of three heparin regimens after successful coronary intervention
J. Am. Coll. Cardiol.,
August 1, 1999;
34(2):
461 - 467.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
S. H. Duda, J. Wiskirchen, M. Erb, U. Schott, K. Khaligi, P. L. Pereira, J. Albes, and C. D. Claussen
Suture-mediated Percutaneous Closure of Antegrade Femoral Arterial Access Sites in Patients Who Have Full Anticoagulation Therapy
Radiology,
January 1, 1999;
210(1):
47 - 52.
[Abstract]
[Full Text]
|
 |
|

|
 |

|
 |
 
M. Gwechenberger, R. Katzenschlager, G. Heinz, M. Gottsauner-Wolf, P. Probst, and M. Gwechenberger
Use of a Collagen Plug Versus Manual Compression for Sealing Arterial Puncture Site After Cardiac Catheterization
Angiology,
February 1, 1997;
48(2):
121 - 126.
[Abstract]
[PDF]
|
 |
|

|
 |

|
 |
 
COLLAGEN PLUGS STANCH BLEEDING AFTER CARDIAC CATHETERIZATION
Journal Watch (General),
April 13, 1993;
1993(413):
4 - 4.
[Full Text]
|
 |
|
|