Advertisement

Click here for more guidelines.

 
 




CME Topic Collections Past Issues Search Current Issue Home
     

J Am Coll Cardiol, 1992; 19:324-332
© 1992 by the American College of Cardiology Foundation
This Article
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Baumgartner, H
Right arrow Articles by Maurer, G
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Baumgartner, H
Right arrow Articles by Maurer, G

Effect of prosthetic aortic valve design on the Doppler-catheter gradient correlation: an in vitro study of normal St. Jude, Medtronic-Hall, Starr-Edwards and Hancock valves

H Baumgartner, S Khan, M DeRobertis, L Czer, and G Maurer

Division of Cardiology and Cardiovascular, Surgery Cedars-Sinai Medical Center, Los Angeles, California.

To evaluate the normal range of Doppler-derived velocities and gradients, their relation to direct flow measurements and the importance of prosthetic valve design on the relation between Doppler and catheter-derived gradients, five sizes of normal St. Jude bileaflet, Medtronic-Hall tilting disc, Starr-Edwards caged ball and Hancock bioprosthetic aortic valves were studied with use of a pulsatile flow model. A strong linear correlation between peak velocity and peak flow, and mean velocity and mean flow, was found in all four valve types (r = 0.96 to 0.99). In small St. Jude and Hancock valves, Doppler velocities and corresponding gradients increased dramatically with increasing flow, resulting in velocities and gradients as high as 4.7 m/s and 89 mm Hg, respectively. The ratio of velocity across the valve to velocity in front of the valve (velocity ratio) was independent of flow in all St. Jude, Medtronic-Hall, Starr-Edwards and Hancock valves when the two lowest flow rates were excluded for Hancock valves. Although Doppler peak and mean gradients correlated well with catheter peak and mean gradients in all four valve types, the actual agreement between the two techniques was acceptable only in Hancock and Medtronic-Hall valves. For St. Jude and Starr-Edwards valves, Doppler gradients significantly and consistently exceeded catheter gradients with differences as great as 44 mm Hg. Thus, Doppler velocities and gradients across normal prosthetic heart valves are highly flow dependent. However, the velocity ratio is independent of flow.(ABSTRACT TRUNCATED AT 250 WORDS)


This article has been cited by other articles:


Home page
Eur Heart J Cardiovasc ImagingHome page
S. Spethmann, H. Dreger, S. Schattke, G. Baldenhofer, D. Saghabalyan, V. Stangl, M. Laule, G. Baumann, K. Stangl, and F. Knebel
Doppler haemodynamics and effective orifice areas of Edwards SAPIEN and CoreValve transcatheter aortic valves
Eur Heart J Cardiovasc Imaging, February 3, 2012; (2012) jes021v1.
[Abstract] [Full Text] [PDF]


Home page
J Am Coll Cardiol ImgHome page
W. J. Stewart and L. L. Rodriguez
Avoiding Patient-Prosthesis Mismatch: Is TAVI a New Solution?
J. Am. Coll. Cardiol. Img., October 1, 2011; 4(10): 1063 - 1066.
[Full Text] [PDF]


Home page
J. Thorac. Cardiovasc. Surg.Home page
J. Price, H. Lapierre, L. Ressler, B.-K. Lam, T. G. Mesana, and M. Ruel
Prosthesis-patient mismatch is less frequent and more clinically indolent in patients operated for aortic insufficiency
J. Thorac. Cardiovasc. Surg., September 1, 2009; 138(3): 639 - 645.
[Abstract] [Full Text] [PDF]


Home page
Ann. Thorac. Surg.Home page
A. Kulik, N. T. Kouchoukos, and M. Ruel
Avoiding Prosthesis-Patient Mismatch in the Elderly: Options Other Than Mechanical Prostheses
Ann. Thorac. Surg., September 1, 2009; 88(3): 1049 - 1050.
[Full Text] [PDF]


Home page
CirculationHome page
A. Kulik, V. Kapila, T. G. Mesana, M. Ruel, and I. G. Burwash
Response to Letter Regarding Article, "Long-Term Outcomes After Valve Replacement for Low-Gradient Aortic Stenosis: Impact of Prosthesis-Patient Mismatch"
Circulation, December 5, 2006; 114(23): e628 - e628.
[Full Text] [PDF]


Home page
CirculationHome page
A. Kulik, I. G. Burwash, V. Kapila, T. G. Mesana, and M. Ruel
Long-Term Outcomes After Valve Replacement for Low-Gradient Aortic Stenosis: Impact of Prosthesis-Patient Mismatch
Circulation, July 4, 2006; 114(1_suppl): I-553 - I-558.
[Abstract] [Full Text] [PDF]


Home page
J. Thorac. Cardiovasc. Surg.Home page
M. Ruel, H. Al-Faleh, A. Kulik, K. L. Chan, T. G. Mesana, and I. G. Burwash
Prosthesis-patient mismatch after aortic valve replacement predominantly affects patients with preexisting left ventricular dysfunction: Effect on survival, freedom from heart failure, and left ventricular mass regression
J. Thorac. Cardiovasc. Surg., May 1, 2006; 131(5): 1036 - 1044.
[Abstract] [Full Text] [PDF]


Home page
HeartHome page
R J Edwards, M Al-Bustami, and C D J Ilsley
A novel method of assessing Starr-Edwards aortic valve stenosis
Heart, December 1, 2004; 90(12): 1479 - 1480.
[Full Text] [PDF]


Home page
Ann. Thorac. Surg.Home page
A. J. Murday, A. Hochstitzky, J. Mansfield, J. Miles, B. Taylor, E. Whitley, and T. Treasure
A prospective controlled trial of st. jude versus starr edwards aortic and mitral valve prostheses
Ann. Thorac. Surg., July 1, 2003; 76(1): 66 - 73.
[Abstract] [Full Text] [PDF]


Home page
J Am Coll CardiolHome page
R. A. Levine and E. Schwammenthal
Stenosis is in the eye of the observer: impact of pressure recovery on assessing aortic valve area
J. Am. Coll. Cardiol., February 5, 2003; 41(3): 443 - 445.
[Full Text] [PDF]


Home page
J. Thorac. Cardiovasc. Surg.Home page
N. Vitale, I. Caldarera, C. Muneretto, R. Sinatra, A. Scafuri, E. Di Rosa, A. Contini, N. Tedesco, A. Pierangeli, M. Abbate, et al.
Clinical evaluation of St Jude Medical Hemodynamic Plus versus standard aortic valve prostheses: The Italian multicenter, prospective, randomized study
J. Thorac. Cardiovasc. Surg., October 1, 2001; 122(4): 691 - 698.
[Abstract] [Full Text] [PDF]


Home page
J. Thorac. Cardiovasc. Surg.Home page
O. Bech-Hanssen, K. Caidahl, I. Wallentin, P. Ask, and B. Wranne
Assessment of effective orifice area of prosthetic aortic valves with Doppler echocardiography: An in vivo and in vitro study
J. Thorac. Cardiovasc. Surg., August 1, 2001; 122(2): 287 - 295.
[Abstract] [Full Text] [PDF]


Home page
J. Thorac. Cardiovasc. Surg.Home page
G. Ismeno, A. Renzulli, M. De Feo, A. Della Corte, F. E. Covino, and M. Cotrufo
Standard versus hemodynamic plus 19-mm St Jude Medical aortic valves
J. Thorac. Cardiovasc. Surg., April 1, 2001; 121(4): 723 - 728.
[Abstract] [Full Text] [PDF]


Home page
J Am Coll CardiolHome page
P. Pibarot and J. G. Dumesnil
Hemodynamic and clinical impact of prosthesis-patient mismatch in the aortic valve position and its prevention
J. Am. Coll. Cardiol., October 1, 2000; 36(4): 1131 - 1141.
[Abstract] [Full Text] [PDF]


Home page
Ann. Thorac. Surg.Home page
S. S. Khan, R. J. Siegel, M. A. DeRobertis, C. E. Blanche, R. M. Kass, W. Cheng, G. P. Fontana, and A. Trento
Regression of hypertrophy after Carpentier-Edwards pericardial aortic valve replacement
Ann. Thorac. Surg., February 1, 2000; 69(2): 531 - 535.
[Abstract] [Full Text] [PDF]


Home page
J. Thorac. Cardiovasc. Surg.Home page
O. Bech-Hanssen, K. Caidahl, B. Wall, P. Myken, S. Larsson, and I. Wallentin
INFLUENCE OF AORTIC VALVE REPLACEMENT, PROSTHESIS TYPE, AND SIZEON FUNCTIONAL OUTCOME AND VENTRICULAR MASS IN PATIENTS WITH AORTIC STENOSIS
J. Thorac. Cardiovasc. Surg., July 1, 1999; 118(1): 57 - 65.
[Abstract] [Full Text] [PDF]


Home page
J Am Coll CardiolHome page
H. Baumgartner, T. Stefenelli, J. Niederberger, H. Schima, and G. Maurer
"overestimation" of catheter gradients by doppler ultrasound in patients with aortic stenosis: a predictable manifestation of pressure recovery
J. Am. Coll. Cardiol., May 1, 1999; 33(6): 1655 - 1661.
[Abstract] [Full Text] [PDF]


Home page
CirculationHome page
D. Y. Leung, J. Wong, L. Rodriguez, M. Pu, P. M. Vandervoort, and J. D. Thomas
Application of Color Doppler Flow Mapping to Calculate Orifice Area of St Jude Mitral Valve
Circulation, September 22, 1998; 98(12): 1205 - 1211.
[Abstract] [Full Text] [PDF]


Home page
J. Thorac. Cardiovasc. Surg.Home page
R. De Paulis, L. Sommariva, G. M. De Matteis, E. Caprara, F. Tomai, A. P. de Peppo, P. Polisca, C. Bassano, and L. Chiariello
EXTENT AND PATTERN OF REGRESSION OF LEFT VENTRICULAR HYPERTROPHY IN PATIENTS WITH SMALL SIZE CARBOMEDICS AORTIC VALVES
J. Thorac. Cardiovasc. Surg., May 1, 1997; 113(5): 901 - 909.
[Abstract] [Full Text]


Home page
CirculationHome page
J. Niederberger, H. Schima, G. Maurer, and H. Baumgartner
Importance of Pressure Recovery for the Assessment of Aortic Stenosis by Doppler Ultrasound: Role of Aortic Size, Aortic Valve Area, and Direction of the Stenotic Jet In Vitro
Circulation, October 15, 1996; 94(8): 1934 - 1940.
[Abstract] [Full Text]


Home page
Ann. Thorac. Surg.Home page
T. Carrel, U. Zingg, R. Jenni, B. Aeschbacher, and M. I. Turina
Early In Vivo Experience With the Hemodynamic Plus St. Jude Medical Heart Valves in Patients With Narrowed Aortic Annulus
Ann. Thorac. Surg., May 1, 1996; 61(5): 1418 - 1422.
[Abstract] [Full Text]


Home page
Ann. Thorac. Surg.Home page
A. Laske, R. Jenni, M. Maloigne, G. Vassalli, O. Bertel, and M. I. Turina
Pressure Gradients Across Bileaflet Aortic Valves by Direct Measurement and Echocardiography
Ann. Thorac. Surg., January 1, 1996; 61(1): 48 - 57.
[Abstract] [Full Text]


Home page
CirculationHome page
P. M. Vandervoort, N. L. Greenberg, M. Pu, K. A. Powell, D. M. Cosgrove, and J. D. Thomas
Pressure Recovery in Bileaflet Heart Valve Prostheses : Localized High Velocities and Gradients in Central and Side Orifices With Implications for Doppler-Catheter Gradient Relation in Aortic and Mitral Position
Circulation, December 15, 1995; 92(12): 3464 - 3472.
[Abstract] [Full Text]


Home page
J. Thorac. Cardiovasc. Surg.Home page
R. De Paulis, L. Sommariva, F. Russo, F. Tomai, A. Tondo, C. Pagliaricci, C. Bassano, and L. Chiariello
Doppler echocardiography evaluation of the CarboMedics valve in patients with small aortic anulus and valve prosthesis-body surface area mismatch
J. Thorac. Cardiovasc. Surg., July 1, 1994; 108(1): 57 - 62.
[Abstract] [Full Text]


Home page
Ann. Thorac. Surg.Home page
R. M. Bojar, H. Rastegar, D. D. Payne, C. A. Mack, and S. L. Schwartz
Clinical and hemodynamic performance of the 19-mm Carpentier-Edwards porcine bioprosthesis
Ann. Thorac. Surg., November 1, 1993; 56(5): 1141 - 1147.
[Abstract] [PDF]



 
  CME Topic Collections Past Issues Search Current Issue Home

Advertisement