JACC
HOME SUBSCRIPTIONS CURRENT ISSUE PAST ISSUES CARDIOSOURCE SEARCH HELP FEEDBACK
 QUICK SEARCH:   [advanced]


     


J Am Coll Cardiol, 1991; 18:1259-1262
© 1991 by the American College of Cardiology Foundation
This Article
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Demer, L.
Right arrow Articles by Siegel, R.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Demer, L.
Right arrow Articles by Siegel, R.

High intensity ultrasound increases distensibility of calcific atherosclerotic arteries

LL Demer, M Ariani, and RJ Siegel

Division of Cardiology, University of California, Los Angeles School of Medicine 90024-1679.

To assess the effect of high intensity ultrasound energy on atherosclerotic arteries, the pressure-volume relation of stenoses in 14 atherosclerotic human cadaver arteries was measured before and after intraarterial application of ultrasound energy. The pressure-volume relation was measured by inflating a 3-mm angioplasty balloon within each artery with use of a syringe equipped with pressure and volume transducers. To minimize potential effects of balloon inflations on compliance, balloon inflation pressure was set at less than or equal to 3 atm before ultrasound application. Ultrasound energy was applied by using a titanium wire probe at a frequency of 20 kHz for 2 min. After exposure to ultrasound, arterial distensibility (measured as volume at 1 atm of balloon pressure) increased by a mean value of 82 +/- 60 microliters. These data suggest that ultrasound energy probably increases the pliability of atherosclerotic lesions by interrupting calcified plaque. This effect may enhance in vivo distensibility and render calcified atherosclerotic lesions more amenable to balloon angioplasty.


This article has been cited by other articles:


Home page
CirculationHome page
P. G. Yock and P. J. Fitzgerald
Catheter-Based Ultrasound Thrombolysis: Shake, Rattle, and Reperfuse
Circulation, March 18, 1997; 95(6): 1360 - 1362.
[Full Text]




HOME SUBSCRIPTIONS CURRENT ISSUE PAST ISSUES CARDIOSOURCE SEARCH HELP FEEDBACK
Copyright © 1991 by the American College of Cardiology Foundation.