J Am Coll Cardiol, 1991; 18:997-1003
© 1991 by the American College of Cardiology Foundation
This Article
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Parro, A
Right arrow Articles by Fujii, H
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Parro, A
Right arrow Articles by Fujii, H

Amplitude information from Doppler color flow mapping systems: a preliminary study of the power mode

A Parro, EF Philpot, PJ Czuwala, PH Fan, NC Nanda, EG Cape, AP Yoganathan, F Helmcke, and H Fujii

Division of Cardiovascular Disease, University of Alabama, Birmingham 35294.

The flow of a saline-glycerin solution with sand particles through a continuous in vitro flow system was imaged by using two commercially available Doppler color flow mapping systems in a power mode (Toshiba SSH-160A and Advanced Technology Laboratories [ATL] Ultramark 9). The images generated from seven solutions with particle concentrations ranging from 0.0001 x 10(12) to 6 x 10(12) particles/liter and a mean velocity of 30 cm/s measured with use of pulsed Doppler ultrasound were used to examine the dependence of the power mode on particle concentration. To examine the velocity dependence, 20 mean velocities ranging from 0.1 to 0.53 m/s (3 to 30 liters/min) and three particle concentrations (1, 3 and 6 x 10(12) particles/liter) in the solution were used. The recorded images were digitized and analyzed off-line. The SUM values, or the adjusted color intensity levels in delineated areas of interest in the displayed flow, were compared. In general, the power mode was sensitive in displaying slower velocity flows; in the selected particle concentration and velocity ranges, it was both velocity and concentration dependent. The specific dependence differed for the two color flow mapping systems.


This article has been cited by other articles:


Home page
J Ultrasound MedHome page
C. Macedonia, J. L. Miller, and B. C. Sonies
Power Doppler Imaging of the Fetal Upper Aerodigestive Tract Using a 4-Point Standardized Evaluation: Preliminary Report
J. Ultrasound Med., August 1, 2002; 21(8): 869 - 878.
[Abstract] [Full Text] [PDF]




HOME SUBSCRIPTIONS CURRENT ISSUE PAST ISSUES CARDIOSOURCE SEARCH HELP FEEDBACK
Copyright © 1991 by the American College of Cardiology Foundation.