J Am Coll Cardiol, 1991; 17:707-711
© 1991 by the American College of Cardiology Foundation
This Article
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by O'Neill, P.
Right arrow Articles by Pacifico, A
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by O'Neill, P.
Right arrow Articles by Pacifico, A

The automatic implantable cardioverter-defibrillator: effect of patch polarity on defibrillation threshold

PG O'Neill, KA Boahene, GM Lawrie, LF Harvill, and A Pacifico

Department of Medicine (Cardiology), Baylor College of Medicine, Houston, Texas.

An automatic implantable cardioverter-defibrillator (AICD) was implanted in 40 patients with sudden cardiac arrest (n = 29), sustained monomorphic ventricular tachycardia (n = 10) or recurrent syncope (n = 1) who were unsuitable for direct ablative surgery or had had unsuccessful medical therapy. The effect of patch electrode polarity on the defibrillation threshold was prospectively evaluated. Two large epicardial patches were used. Initial polarity was selected at random. Ventricular fibrillation was induced by direct current and a preestablished defibrillation protocol employed to assess the minimal energy that would reproducibly defibrillate the heart. Nineteen patients had a lower defibrillation threshold with the inferior left ventricular patch as an anode and nine patients had a lower defibrillation threshold with this patch as a cathode. In general, the defibrillation threshold was lower when this patch was used as an anode than when it was used as a cathode (18 +/- 10 versus 22.6 +/- 12.2 J; p less than 0.01). No preoperative variable predicted optimal polarity. Therefore, the effect of patch polarity on defibrillation threshold should be assessed in each patient at the time of AICD implantation so that the safety margin for satisfactory device function can be maximized.


This article has been cited by other articles:


Home page
CirculationHome page
J. Huang, B. H. KenKnight, G. P. Walcott, D. L. Rollins, W. M. Smith, and R. E. Ideker
Effects of Transvenous Electrode Polarity and Waveform Duration on the Relationship Between Defibrillation Threshold and Upper Limit of Vulnerability
Circulation, August 19, 1997; 96(4): 1351 - 1359.
[Abstract] [Full Text]




HOME SUBSCRIPTIONS CURRENT ISSUE PAST ISSUES CARDIOSOURCE SEARCH HELP FEEDBACK
Copyright © 1991 by the American College of Cardiology Foundation.