Clinical interactions between pacemakers and automatic implantable cardioverter-defibrillators
H Calkins,
J Brinker,
EP Veltri,
T Guarnieri,
and
JH Levine
Department of Medicine, Johns Hopkins Hospital, Baltimore, Maryland.
Concomitant use of a pacemaker and an automatic implantable cardioverter-defibrillator (AICD) is common. Seventeen percent of patients receiving an AICD at The Johns Hopkins Hospital also had a permanent pacemaker implanted before (16 patients), at the same time as (2 patients) or after (12 patients) AICD implantation. Four types of interactions were noted: 1) transient failure to sense or capture immediately after AICD discharge (seven patients); 2) oversensing of the pacemaker stimulus by the AICD, leading to double counting (one patient); 3) AICD failure to sense ventricular fibrillation resulting from pacemaker stimulus oversensing (three patients, one only at high asynchronous output); and 4) pacemaker reprogramming caused by AICD discharge (three patients). No clinical sequelae of these interactions were noted during follow-up study. Thus, potentially adverse clinical interactions are common and routine screening is recommended. With proper attention to lead placements and programming of the devices, clinical consequences of these interactions can be avoided.
This article has been cited by other articles:

|
 |

|
 |
 
T. Rauwolf, M. Guenther, N. Hass, A. Schnabel, M. Bock, M.U. Braun, and R.H. Strasser
Ventricular oversensing in 518 patients with implanted cardiac defibrillators: incidence, complications, and solutions
Europace,
November 1, 2007;
9(11):
1041 - 1047.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
M. H. Schoenfeld
Contemporary Pacemaker and Defibrillator Device Therapy: Challenges Confronting the General Cardiologist
Circulation,
February 6, 2007;
115(5):
638 - 653.
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
M. I. Cohen, R. Jedeikin, S. Pfeiffer, and S. Pedersen
Unipolar ventricular pacing reduces inappropriate shocks from a separate cardioverter defibrillator in a post-Fontan patient.
Europace,
April 1, 2006;
8(4):
293 - 296.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
Z. Goldberger, B. Elbel, C. A. McPherson, A. D. Paltiel, and R. Lampert
Cost Advantage of Dual-Chamber Versus Single-Chamber Cardioverter-Defibrillator Implantation
J. Am. Coll. Cardiol.,
September 6, 2005;
46(5):
850 - 857.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
T. V. Salukhe, D. Dob, and R. Sutton
Pacemakers and defibrillators: anaesthetic implications
Br. J. Anaesth.,
July 1, 2004;
93(1):
95 - 104.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
A. Quesada, J.ús Almendral, F. Arribas, R. Ricci, C. Wolpert, P. Adragao, E. Cobo, X. Navarro, and on behalf DATAS investigators
The DATAS rationale and design: a controlled, randomized trial to assess the clinical benefit of dual chamber (DDED) defibrillator
Europace,
January 1, 2004;
6(2):
142 - 150.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
E. Himmrich, O. Przibille, C. Zellerhoff, A. Liebrich, S. Rosocha, K. Andreas, D. Nebeling, B. Omogbehin, and J. Meyer
Proarrhythmic Effect of Pacemaker Stimulation in Patients With Implanted Cardioverter-Defibrillators
Circulation,
July 15, 2003;
108(2):
192 - 197.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
V. Kuhlkamp and InSync 7272 ICD World Wide Investigators
Initial experience with an implantable cardioverter-defibrillator incorporating cardiac resynchronization therapy
J. Am. Coll. Cardiol.,
March 6, 2002;
39(5):
790 - 797.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
F. Duru, U. Bauersfeld, and R. Candinas
Avoiding inappropriate ventricular tachycardia detection due to T-wave oversensing in an implantable cardioverter defibrillator: a novel application of the electrogram width criterion
Europace,
January 1, 2001;
3(1):
80 - 84.
[Abstract]
[PDF]
|
 |
|

|
 |

|
 |
 
A. Proclemer, P. Della Bella, D. Facchin, L. Fattore, C. Carbucicchio, C. Tondo, M. Lunati, M. R. Vecchi, E. Petz, and M. Zecchin
Indications for dual-chamber cardioverter defibrillators at implant and at 1 year follow-up: a retrospective analysis in the single-chamber defibrillator era
Europace,
January 1, 2001;
3(2):
132 - 135.
[Abstract]
[PDF]
|
 |
|

|
 |

|
 |
 
G. R. Haddow and M. Neville
Anesthetic Implications for Patients With Implantable Cardioverter Defibrillators
Seminars in Cardiothoracic and Vascular Anesthesia,
September 1, 2000;
4(3):
187 - 195.
[Abstract]
[PDF]
|
 |
|

|
 |

|
 |
 
S. Walker, T.M. Levy, A.J.S. Coats, N.S. Peters, and V.E. Paul
Bi-ventricular pacing in congestive cardiac failure. Current experience and future directions
Eur. Heart J.,
June 1, 2000;
21(11):
884 - 889.
[PDF]
|
 |
|

|
 |

|
 |
 
C. Sticherling, A. Schaumann, T. Klingenheben, S. H. Hohnloser, and for the Ventak AV II DR investigators
First worldwide clinical experience with a new dual chamber implantable cardioverter defibrillator: Advantages and complications
Europace,
January 1, 1999;
1(2):
96 - 102.
[Abstract]
[PDF]
|
 |
|

|
 |

|
 |
 
S. Walker, T. Levy, S. Rex, and V.E. Paul
Biventricular implantable cardioverter defibrillator use in a patient with heart failure and ventricular tachycardia secondary to Emery-Dreifuss syndrome
Europace,
January 1, 1999;
1(3):
206 - 209.
[Abstract]
[PDF]
|
 |
|

|
 |

|
 |
 
S. L. Pinski and G. J. Fahy
The Proarrhythmic Potential of Implantable Cardioverter-Defibrillators
Circulation,
September 15, 1995;
92(6):
1651 - 1664.
[Abstract]
[Full Text]
|
 |
|

|
 |

|
 |
 
T. J. Cohen, P. R. Reid, M. M. Mower, M. Mirowski, D. Aarons, J. Juanteguy, and E. P. Veltri
The Automatic Implantable Cardioverter-Defibrillator: Long-term Clinical Experience and Outcome at a Hospital Without an Open-Heart Surgery Program
Arch Intern Med,
January 1, 1992;
152(1):
65 - 69.
[Abstract]
[PDF]
|
 |
|

|
 |

|
 |
 
INTERACTIONS BETWEEN PACEMAKERS AND IMPLANTABLE DEFIBRILLATORS
Journal Watch (General),
September 14, 1990;
1990(914):
4 - 4.
[Full Text]
|
 |
|
|