A prospective, randomized evaluation of effect of ventricular fibrillation duration on defibrillation thresholds in humans
GH Bardy,
TD Ivey,
M Allen,
and
G Johnson
Department of Medicine, University of Washington, Seattle.
The effect of ventricular fibrillation duration in humans on defibrillation efficacy as it pertains to the time of intervention of an automatic implantable defibrillator is unknown. If a difference in defibrillation efficacy exists in the early period after ventricular fibrillation onset, it may affect algorithms used by antiarrhythmic devices for arrhythmia detection and therapy. Therefore, a prospective, randomized evaluation was performed of the effect of ventricular fibrillation durations of 10 s and 20 s on defibrillation thresholds in 10 survivors of sudden cardiac arrest undergoing implantation of an automatic cardioverter defibrillator. The initial duration of ventricular fibrillation was chosen randomly. Subsequently, each patient served as his or her own control for the alternate duration of ventricular fibrillation to that chosen initially. The mean leading edge defibrillation threshold voltage was 411 +/- 114 V when ventricular fibrillation persisted for 10 s and 419 +/- 125 V when it persisted for 20 s (p = 0.73). The mean defibrillation threshold current was 11.4 +/- 2.8 A when ventricular fibrillation persisted for 10 s and 11.4 +/- 3.2 A when it persisted for 20 s (p = 0.97). The delivered energy defibrillation threshold was 11.5 +/- 5.9 J when ventricular fibrillation persisted for 10 s and 12.0 +/- 6.9 J when it persisted for 20 s (p = 0.67). These findings show that the defibrillation threshold does not change between 10 and 20 s of ventricular fibrillation in out-of-hospital survivors of cardiac arrest at the time of surgical implantation of an automatic defibrillator. The data may have influence on the programming of defibrillator detection algorithms.
This article has been cited by other articles:

|
 |

|
 |
 
C. T. Leng, N. A. Paradis, H. Calkins, R. D. Berger, A. C. Lardo, K. C. Rent, and H. R. Halperin
Resuscitation After Prolonged Ventricular Fibrillation With Use of Monophasic and Biphasic Waveform Pulses for External Defibrillation
Circulation,
June 27, 2000;
101(25):
2968 - 2974.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
G. P. Walcott, S. B. Melnick, F. W. Chapman, J. L. Jones, W. M. Smith, and R. E. Ideker
Relative Efficacy of Monophasic and Biphasic Waveforms for Transthoracic Defibrillation After Short and Long Durations of Ventricular Fibrillation
Circulation,
November 17, 1998;
98(20):
2210 - 2215.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
S. A. Strickberger, E. G. Daoud, T. Davidson, R. Weiss, F. Bogun, B. P. Knight, M. Bahu, R. Goyal, K. C. Man, and F. Morady
Probability of Successful Defibrillation at Multiples of the Defibrillation Energy Requirement in Patients With an Implantable Defibrillator
Circulation,
August 19, 1997;
96(4):
1217 - 1223.
[Abstract]
[Full Text]
|
 |
|

|
 |

|
 |
 
E. U. Alt, P. C. Fotuhi, R. L. Callihan, E. Mestre, W. M. Smith, and R. E. Ideker
Endocardial Carbon-Braid Electrodes : A New Concept for Lower Defibrillation Thresholds
Circulation,
September 15, 1995;
92(6):
1627 - 1633.
[Abstract]
[Full Text]
|
 |
|

|
 |

|
 |
 
A. S. Manolis, H. Rastegar, and N. A. M. Estes III
Automatic Implantable Cardioverter Defibrillator: Current Status
JAMA,
September 8, 1989;
262(10):
1362 - 1368.
[Abstract]
[PDF]
|
 |
|
|