Clinical Research |

Phase II Safety and Clinical Comparison With Single-Photon Emission Computed Tomography Myocardial Perfusion Imaging for Detection of Coronary Artery Disease: Flurpiridaz F 18 Positron Emission Tomography

Daniel S. Berman, MD; Jamshid Maddahi, MD; B.K. Tamarappoo, PhD, MD; Johannes Czernin, MD; Raymond Taillefer, MD; James E. Udelson, MD; C. Michael Gibson, MD; Marybeth Devine, BS; Joel Lazewatsky, PhD; Gajanan Bhat, PhD; Dana Washburn, MD
[+] Author Information

Dr. Tamarappoo is currently at the Department of Cardiovascular Medicine, Cleveland Clinic, Cleveland, Ohio. Dr. Maddahi has served as consultant to Lantheus Medical Imaging, Digirad, and Astellas; served on the Speakers' Bureau for Astellas; and contracted research with Lantheus Medical Imaging. Dr. Taillefer has served as consultant to and as a member of the scientific advisory board of Lantheus Medical Imaging. Dr. Gibson has received research grant funding from Lantheus Medical Imaging. Ms. Devine is an employee of Avid Radiopharmaceuticals. Drs. Bhat, Lazewatsky, and Washburn are employees of Lantheus Medical Imaging. All other authors have reported that they have no relationships relevant to the contents of this paper to disclose.

Reprint requests and correspondence: Dr. Daniel S. Berman, Departments of Imaging and Medicine, Cedars-Sinai Heart Institute, 8700 Beverly Boulevard, Room 1258, Los Angeles, California 90048

Copyright 2013, American College of Cardiology Foundation. All Rights Reserved.

J Am Coll Cardiol. 2013;61(4):469-477. doi:10.1016/j.jacc.2012.11.022
Published online

Objectives  This was a phase II trial to assess flurpiridaz F 18 for safety and compare its diagnostic performance for positron emission tomography (PET) myocardial perfusion imaging (MPI) with Tc-99m single-photon emission computed tomography (SPECT) MPI with regard to image quality, interpretative certainty, defect magnitude, and detection of coronary artery disease (CAD) (≥50% stenosis) on invasive coronary angiography (ICA).

Background  In pre-clinical and phase I studies, flurpiridaz F 18 has shown characteristics of an essentially ideal MPI tracer.

Methods  One hundred forty-three patients from 21 centers underwent rest-stress PET and Tc-99m SPECT MPI. Eighty-six patients underwent ICA, and 39 had low-likelihood of CAD. Images were scored by 3 independent, blinded readers.

Results  A higher percentage of images were rated as excellent/good on PET versus SPECT on stress (99.2% vs. 88.5%, p < 0.01) and rest (96.9% vs. 66.4, p < 0.01) images. Diagnostic certainty of interpretation (percentage of cases with definitely abnormal/normal interpretation) was higher for PET versus SPECT (90.8% vs. 70.9%, p < 0.01). In 86 patients who underwent ICA, sensitivity of PET was higher than SPECT (78.8% vs. 61.5%, respectively, p = 0.02). Specificity was not significantly different (PET: 76.5% vs. SPECT: 73.5%). Receiver-operating characteristic curve area was 0.82 ± 0.05 for PET and 0.70 ± 0.06 for SPECT (p = 0.04). Normalcy rate was 89.7% with PET and 97.4% with SPECT (p = NS). In patients with CAD on ICA, the magnitude of reversible defects was greater with PET than SPECT (p = 0.008). Extensive safety assessment revealed that flurpiridaz F 18 was safe in this cohort.

Conclusions  In this phase 2 trial, PET MPI with flurpiridaz F 18 was safe and superior to SPECT MPI for image quality, interpretative certainty, and overall CAD diagnosis.

Figures in this Article

Over the past 4 decades, radionuclide myocardial perfusion imaging (MPI) has become the mainstay of stress imaging for the evaluation of patients with suspected or known coronary artery disease (CAD). The method offers objective means of detecting stress-induced perfusion defects due to hemodynamically significant coronary stenoses, thereby providing an important tool in risk stratification and guiding patient management ((1),(2),3). Although the radiopharmaceuticals commonly used for MPI are distributed to the myocardium in proportion to myocardial perfusion, a limitation of widely available radiopharmaceuticals used for this purpose has been a “roll-off” of their uptake at higher levels of coronary flow ((4),(5),(6),(7),(8),9). As a consequence, mild reductions in maximal achievable coronary flow might not result in corresponding decreases in myocardial tracer uptake. Furthermore, these agents have lower spatial resolution than the positron emission tomography (PET) agents fluorine-18 (F-18) and nitrogen-13. These limitations have long been the impetus toward the development of radiopharmaceuticals that would have superior uptake versus flow relationships at peak flow rates as well as higher image resolution. Short-lived cyclotron-produced PET MPI agents, including oxygen-15 water and nitrogen-13 ammonia, have excellent flow versus uptake characteristics at high flow rates ((10),11); however, these tracers are not practical for widespread clinical use due to the requirement that they be used in very close proximity to medical cyclotrons.

Flurpiridaz F 18, labeled with positron-emitting isotope F-18, has recently been developed as an MPI radiopharmaceutical for use with PET ((12),(13),(14),(15),(16),17). In pre-clinical as well as phase I studies, flurpiridaz F 18 has been shown to have essentially linear myocardial uptake throughout the range of flow ((13),17). It also has high myocardial retention, low background in adjacent organs, and other characteristics that suggest it might be close to an ideal myocardial perfusion tracer ((7),(15),16). The purpose of this phase II trial was to assess the clinical safety of flurpiridaz F 18 and to compare the diagnostic performance of this agent with Tc-99m MPI tracers with respect to image quality, certainty of interpretation, magnitude of reversible perfusion defects, and diagnostic efficacy for detection of CAD.

Study population

Patients were eligible for the study if they were referred for clinically indicated single-photon emission computed tomography (SPECT) MPI or if they had undergone invasive coronary angiography (ICA) without percutaneous coronary intervention (PCI) within 60 days. Patients were recruited from 21 centers in the United States. All potential child-bearing female patients underwent pregnancy testing within 24 h before receiving the PET tracer. Patients were not eligible for the study if they were <18 years of age, had PCI or coronary artery bypass grafting within 6 months before either PET or SPECT, had nonischemic cardiomyopathy or an ejection fraction of <35%, or were unable to undergo exercise or pharmacological stress testing. The study was approved by the institutional review boards of all participating centers, and all patients provided written informed consent.

Study design

All patients underwent SPECT MPI within 90 days before or 60 days after flurpiridaz F 18 PET MPI. If SPECT MPI had been performed before enrollment on an accepted SPECT camera, repeat SPECT examination was not performed. In those without such a SPECT study, both SPECT and PET scans were performed for purposes of the research trial during the study. In patients who had ICA, the ICA was performed within 60 days of PET MPI.

All patients were evaluated for flurpiridaz F 18 safety. Patients who underwent any imaging were evaluated for image quality and diagnostic certainty. Patients who underwent ICA were evaluated for test performance characteristics, whereas those who did not undergo ICA but who could be classified as low CAD likelihood were evaluated for normalcy rate.

Stress testing protocols

Stress tests were performed with either exercise or pharmacological stress according to American Society of Nuclear Cardiology Guidelines (18), with the intention to have the same stress test for both procedures. Exercise testing was performed with a standard Bruce exercise treadmill protocol, and the radiopharmaceutical was injected at the same time-point during the SPECT and PET procedures. Pharmacological stress tests were conducted with adenosine, regadenoson, or dipyridamole. Patients were instructed to have no caffeine intake for 24 h before pharmacological stress testing and were taking the same cardiac active medications at the time of both SPECT and PET studies.

SPECT imaging protocol

Tc-99m agents were used for all SPECT imaging (73 Tc-99m sestamibi and 35 Tc-99m tetrofosmin in 1-day protocol; 10 Tc-99m sestamibi and 6 Tc-99m tetrofosmin in 2-day protocol). There were 19 cases performed with the rest Tl-201 and stress Tc-99m dual isotope MPI protocol. All PET and SPECT cameras used in the study were subjected to and passed quality control procedures reviewed by the core laboratory for purposes of the study. The SPECT scanning was performed with standard American Society of Nuclear Cardiology Guidelines with either rest/stress or a stress/rest protocol (18).

PET imaging protocol

All PET imaging was performed at rest and stress in accordance with the pre-defined study protocol with either PET/computed tomography (CT) systems (n = 140) or dedicated PET scanners (n = 3). All PET cameras used in the study were subjected to quality control procedures for the purpose of this study. For PET/CT, attenuation correction CT scans were obtained before rest and stress examinations. For dedicated PET, transmission scans for attenuation correction scans were performed with a radioisotope source. Imaging was performed in list-mode. Rest imaging began at the time of the flurpiridaz F 18 injection and was continuously acquired over 20 min. Stress dosing began 73.4 ± 12.1 min after rest studies in patients undergoing exercise stress and 52.9 ± 11.2 min after rest studies in patients undergoing pharmacological stress testing. One patient undergoing exercise stress had rest and stress procedures completed on different dates and was not included in the aforementioned summary of the rest/stress dose interval. In the pharmacological stress group, 2.91 ± 0.70 mCi of flurpiridaz F 18 was injected at rest, and 6.76 ± 1.78 mCi was injected at stress. In the exercise group, these doses were 1.93 ± 0.37 mCi and 6.48 ± 1.23 mCi, respectively. The mean effective radiation dose to the patients from these injections is estimated to be approximately 6.4 mSv and 4.8 mSv for the pharmacological and exercise rest/stress protocols, respectively (14).

MPI image interpretation

All images were interpreted in an imaging core laboratory (Les Services Nucmed, Montreal, Canada). Both overall qualitative diagnosis and semi-quantitative 17-segment with 5-point (0 = normal, 4 = absent tracer uptake) scoring were employed in the independent, blinded read by 3 expert readers who were experienced in both SPECT and PET MPI interpretation. The same 3 readers interpreted all of the SPECT and PET studies in independent cross-over reading sessions. The readers were blinded to which study was being performed, the type of stress, patient sex, and all clinical data as well as to the reading of other readers. Image quality was classified as excellent, good, fair, poor, or uninterpretable. Certainty of overall interpretation was classified definitely normal, probably normal, probably abnormal, and definitely abnormal.

For the purpose of assessing diagnostic efficacy, each blinded reader rated the rest/stress perfusion and gated images as normal, ischemic, ischemic and scar, or scar. A patient was considered MPI negative if the rating was normal. Summed stress scores (SSS), summed rest scores (SRS), and summed difference scores (SDS) were calculated by the core laboratory (18). A patient was considered to have a reversible defect when the SDS was ≥2. For patients who underwent attenuation-corrected SPECT or both supine and prone SPECT acquisitions, the readers assessed only the non–attenuation-corrected, supine SPECT images.

Coronary angiography interpretation

All coronary angiograms were interpreted with quantitative coronary angiography (QCA) in a coronary angiography core laboratory (Perfuse, Boston, Massachusetts) blinded to the clinical or imaging results. A coronary stenosis was considered present when there was ≥50% diameter stenosis in any epicardial coronary artery. The presence of 1 or more coronary stenoses defined the presence of significant CAD. The core laboratory had no knowledge of history of myocardial infarction (MI). Presence or absence of wall motion abnormality was not considered in interpreting the presence or absence of CAD. The determination of the presence or absence of CAD was strictly related to coronary stenosis in the native coronary arteries. Arteries with patent stents were classified as no significant CAD, regardless of evidence of prior MI. Arteries with patent bypass grafts but native coronary stenosis were classified as significant CAD.

Safety analysis

Safety assessments included blood and urine testing as well as clinical monitoring for possible adverse events (AEs). Patients were monitored for 30 min after the rest injection and for 60 min after the stress injections for clinical symptoms, vital signs, and electrocardiogram changes. They were followed for 2 weeks after the drug administration for AEs. All AEs were adjudicated by the site principal investigator, including whether they were considered to be related to the drug or to the stress test. A serious AE was defined as any AE considered life threatening, resulting in death, requiring in-patient hospital stay, or persistent or significant disability/incapacity.

Statistical analysis

Continuous variables were expressed as mean (or median) and range or SDs. Paired continuous variables that were not normally distributed were compared with the Wilcoxon signed rank test for matched pairs or the paired t test if normally distributed. Categorical variables were expressed as frequencies. Primary diagnostic efficacy measures were sensitivity and specificity and were assessed after blinded and independent reads of PET MPI and SPECT MPI data with ICA as the truth standard (n = 86). Efficacy estimates for PET MPI and SPECT MPI were compared individually with a 2-sided paired test of proportions (McNemar test) at 5% level of significance. Normalcy rate was determined in the patients with a low likelihood of CAD and no ICA and was calculated as percentage of patients with normal/probably normal studies. Analysis of diagnostic efficacy was performed by reader as well as with majority rule, and the results of majority rule analysis were used in the overall efficacy analysis. Where majority rule was used for the comparison between PET and SPECT MPI, the estimate was obtained with the rating obtained by 2 of 3 blinded readers for each patient. The analysis of diagnostic certainty used a modified majority rule in which the median rating was used when 3 readers have 3 different ratings. Summaries and comparisons of SRS, SSS, and SDS measures similarly used the median of the results of 3 blinded readers. Receiver-operating characteristic (ROC) curve analysis was performed in the 86 patients undergoing ICA with the comparison of area under ROC between PET MPI and SPECT MPI diagnostic efficacy. The area under ROC and confidence intervals used the empirical approach developed by DeLong et al. (19).

The interreader agreement between 3 blinded readers was estimated by comparison of dichotomized ratings (normal/abnormal) and was presented with kappa and percentage agreement. Intrareader agreement for each of the readers was performed by randomizing 10% of the repeat images in the blinded read and was estimated with kappa and percentage agreement.

Patient population characteristics

There were 143 patients enrolled into the overall study population, 107 men, and the mean age was 62.5 years (Table 1). The mean body mass index was 28.3. There were 108 white, 3 Asian, 16 African American individuals, and 16 who classified themselves as “other” race/ethnicity. Patients were stratified into 3 pre-test CAD likelihood categories: low (n = 45); intermediate (n = 68); and high (n = 30) ((18),20), the latter including patients with known CAD. In the angiographic cohort (n = 86), 52 patients had significant CAD by QCA, and 34 patients had no significant CAD or normal coronary arteriograms. In this cohort, 70 patients had SPECT studies performed before PET and ICA; 15 patients had ICA before PET and SPECT studies, and 1 patient had PET study completed before ICA and SPECT study. In this cohort, 22 had prior MI, 42 had prior revascularizations, and 43 had either prior MI or revascularization. (Figure 22_gr1) depicts the manner in which the 143 patients were enrolled and evaluated. Eleven patients were excluded for the reasons listed, leaving 132 patients who had both SPECT MPI and PET MPI studies evaluable for image quality and diagnostic certainty comparison. Seven patients of the 132 with an intermediate or high likelihood of CAD did not undergo ICA. The remaining 125 patients were evaluable for diagnostic efficacy and included 86 with ICA evaluated for sensitivity and specificity for angiographic CAD and 39 patients with a low likelihood of CAD assessed for normalcy rate.

Table Grahic Jump Location
Table 1Patient Characteristics
Grahic Jump Location
Figure 1

Summary of Patient Enrollment and Disposition

CAD = coronary artery disease; ICA = invasive coronary angiography; MPI = myocardial perfusion imaging; PET = positron emission tomography; SPECT = single-photon emission computed tomography.

Stress flurpiridaz F 18 PET MPI was performed with either exercise (n = 76) or adenosine (n = 67) stress. For SPECT MPI, pharmacological stress was conducted with adenosine (n = 44), regadenoson (n = 7), or dipyridamole (n = 10). A total of 5 of 143 patients had different stress test types between 2 MPI procedures.

Clinical safety

Among the 132 patients in the safety population, 2 patients reported a total of 3 serious AEs (dizziness, hypertension, and knee effusion) that were judged to be unrelated to study drug. Sixty-one patients (42.7% of the 143) reported 108 AEs. Thirty-six of these 61 patients received pharmacological stress with a total of 75 AEs, and the remaining 25 patients received exercise stress with a total of 33 AEs. One hundred of the 108 AEs were considered not related to study drug. Of the remaining 8 AEs judged as possibly or probably related to the study drug, all were of either mild or moderate severity and resolved without any complications. Five of these 8 were also judged as possibly or probably related to the stress test itself. The remaining 3 of the 8 patients had the following: transient hypertension, mild cough, and a metallic taste. The AEs associated with the Tc-99m and Tl-201 injections were not obtained. There was no evidence of clinically relevant flurpiridaz F 18 effect on QT interval in this study. An independent data-monitoring committee that reviewed safety data at regular intervals did not raise any significant safety concerns.

Image quality

Among the 132 patients assessed for image quality, image quality was judged to be superior with flurpiridaz F 18 for both stress and rest (Figure 22_gr2). On the stress studies, 99.2% were considered to be good or excellent with flurpiridaz F 18 versus 88.5% with Tc-99m (p < 0.01). On the rest studies, image quality was considered to be good or excellent in 96.9% with flurpiridaz F 18 versus 66.4% with Tc-99m (p < 0.01). Excellent image quality alone was observed on stress images in 80.9% with flurpiridaz F 18 and in 24.4% with Tc-99m (p < 0.01) and on rest studies in 69.5% with flurpiridaz F 18 versus 7.6% with Tc-99m (p < 0.01).

Grahic Jump Location
Figure 2

Image Quality Comparing Flurpiridaz F 18 PET MPI and SPECT MPI (N = 132)

N = 132. Shown are the proportion of patients with excellent or good image quality (left) and excellent only (right). Abbreviations as in (Figure 1).

Diagnostic certainty

In the overall efficacy population of 132 patients, diagnostic certainty was greater with flurpiridaz F 18 than with Tc-99m (Figure 22_gr3). Certainty of interpretation was classified as definitely normal or definitely abnormal in 90.8% of patients by flurpiridaz F 18 and in 70.9% of patients by Tc-99m (p < 0.01), with approximately a 3-fold reduction in the patients in the probably normal and probably abnormal groups.

Grahic Jump Location
Figure 3

Diagnostic Certainty Comparing Flurpiridaz F 18 PET MPI and SPECT MPI

N = 132. Def. = definitely; Prob. = probably; other abbreviations as in (Figure 1).

Detection of CAD

In the 86 patients who underwent ICA and thus had a truth standard for coronary stenosis, 52 had significant CAD on ICA. The sensitivity of PET was higher than that of SPECT (78.8% vs. 61.5%, respectively, p = 0.02). The specificity was not significantly different (76.5% for PET vs. 73.5% for SPECT) (Figure 22_gr4). The area under the ROC curve was 0.82 ± 0.05 for PET and 0.70 ± 0.06 for SPECT (p = 0.04) (Figure 22_gr5). In the 39 patients with a low likelihood of CAD, the normalcy rate was 89.7% with PET and 97.4% with SPECT (p = NS).

Grahic Jump Location
Figure 4

Sensitivity and Specificity Comparing Flurpiridaz F 18 PET MPI and SPECT MPI for Detection of CAD

Abbreviations as in (Figure 1).

Grahic Jump Location
Figure 5

Receiver-Operating Characteristics Analysis

PET = positron emission tomography; SPECT = single-photon emission computed tomography.

Magnitude of stress and rest perfusion defect

The SSS and SRS scores in the patients who underwent ICA are shown in (Table 2) according to the number of vessels abnormal on ICA. In patients with 1- and 2-vessel stenosis, the median SSS was greater with PET than with SPECT.

Table Grahic Jump Location
Table 2Summary of Median SSS and SRS Among Readers by Number of Vessels With Coronary Stenosis
Magnitude and frequency of reversible perfusion defect

The magnitude of reversible perfusion defect on PET and SPECT in the patients who underwent ICA is shown in Tables (Table 3) and (Table 4). In the patients with coronary stenosis (Table 3), the mean SDS and the proportion of patients with SDS ≥2 was higher with PET than SPECT as interpreted by each reader. In the patients without coronary stenosis (Table 4), the mean SDS and the proportion of patients with SDS ≥2 was not significantly different between the 2 MPI modalities for any of the readers.

Table Grahic Jump Location
Table 3Summary of SDS in Patients With Angiography-Positive Results
Table Grahic Jump Location
Table 4Summary of SDS in Patients With Angiography-Negative Results
Reader agreement

Interreader agreement rates for presence of abnormality were ≥90% in PET MPI between 3 readers with kappa values of ≥0.80 between readers. Intrareader agreement rates were slightly lower for SPECT MPI (≥80%) and kappa values of at least 0.60. Intrareader agreement rates were 80% or greater for both imaging modalities.

Case examples

Examples of PET and SPECT images from a patient with no coronary stenosis on ICA are shown in (Figure 22_gr6) and from a patient with left anterior descending coronary artery stenosis in (Figure 22_gr7).

Grahic Jump Location
Figure 6

FLUR PET and MIBI SPECT Images From a 75-Year-Old Man

Flurpiridaz F 18 (FLUR) PET (top) and Tc-99m sestamibi (MIBI) SPECT (bottom) images from a 75-year-old man, with a body mass index of 32.6 and no CAD on ICA. The FLUR images were interpreted as normal, whereas the MIBI images were interpreted as probably abnormal with a possible reversible defect in the basal inferior wall. Abbreviations as in (Figure 1).

Grahic Jump Location
Figure 7

FLUR PET and MIBI SPECT Images From an 82-Year-Old Man

The FLUR PET (top) and MIBI SPECT (bottom) images from an 82-year-old man with shortness of breath and an occluded native proximal left anterior descending (LAD) coronary artery and an occluded left internal mammary graft to the LAD and no other significant native CAD. The FLUR images show a severe reversible perfusion defect throughout the territory of the occluded proximal LAD, whereas the MIBI images show only a moderate perfusion defect in the distal LAD territory (apical slices). Abbreviations as in (Figure 1) and (Figure 6).

This is the first multicenter clinical trial comparing the diagnostic performance of flurpiridaz F 18, a new myocardial perfusion PET tracer with conventional Tc-99m SPECT. Interpreted with rigorous core laboratory image analysis, the findings show that flurpiridaz F 18 can detect hemodynamically significant obstructive CAD with greater sensitivity, superior image quality, and with greater diagnostic certainty compared with SPECT MPI. The radiation doses associated with the flurpiridaz F 18 injections were approximately 2× lower than those associated with standard rest/stress Tc-99m SPECT MPI protocols. Importantly, in patients with significant coronary stenosis, the frequency and magnitude of reversible perfusion defect was greater with PET than SPECT. The extensive safety assessment conducted in this phase II study revealed that flurpiridaz F 18 was safe and well-tolerated in this study.

Image quality and diagnostic certainty

Variation of image quality and image interpretation has been considered one of the limitations of conventional SPECT MPI. In this study, image quality and diagnostic certainty were both higher with flurpiridaz F 18 PET than with SPECT. Image quality was good-to-excellent in a greater proportion of PET than SPECT scans and excellent in a far greater proportion of patients, both at stress and at rest. This superior image quality was associated with a higher diagnostic certainty with flurpiridaz F 18 by independent readers. The improved image quality and diagnostic certainty are likely due to the greater conspicuity of perfusion defects, improved resolution associated with F-18 PET, as well as the improved diagnostic certainty associated with the use of PET compared with SPECT and the use of attenuation correction for the PET studies (21). Of note, interreader agreement rates were high with both PET and SPECT and were higher with the PET examinations.

Sensitivity and specificity for detection of CAD and magnitude of perfusion defects

In this study we show that flurpiridaz F 18 PET was associated with improved sensitivity when compared with SPECT, with specificity similar to SPECT. Improved diagnostic accuracy was confirmed by a significantly greater receiver operating characteristics curve area for flurpiridaz F 18 PET compared with SPECT. Several factors might explain this improved sensitivity for CAD detection, including the higher extraction fraction of the tracer, the improved spatial resolution of PET versus SPECT, and the use of attenuation correction in the PET studies.

Magnitude of perfusion defects

The magnitude of stress and reversible perfusion defect with flurpiridaz F 18 PET was nearly double that of the Tc-99m SPECT studies in patients with coronary stenosis. The increased magnitude of perfusion defects in patients with angiographic stenosis was seen with each of the individual readers, as was the proportion of patients with disease-manifesting significant amounts of ischemia (SDS ≥2). The magnitude of perfusion defect was not different between the tracers in the patients without significant stenosis.

Potential explanations for greater stress perfusion defect magnitude and higher sensitivity for CAD detection

Flurpiridaz F 18 exhibits near-linear myocardial extraction and retention across the range of achievable myocardial blood flow (17). In contrast, a “roll off” of extraction at high myocardial perfusion rates has been reported for the commonly used SPECT agents Tc-99m-sestamibi, Tc-99m-tetrofosmin, and Tl-201 ((4),(5),(6),7) as well as for the generator-produced PET agent Rb-82 ((8),9). The higher extraction fraction of flurpiridaz F 18 compared with the SPECT tracers (17) at high flow rates is likely to be an important factor explaining the differences in magnitude of perfusion defect as well as higher sensitivity for perfusion defect detection observed in this study. During maximal exercise or with maximal coronary vasodilation, areas of the myocardium supplied by vessels without stenosis demonstrate myocardial perfusion rates in the range of 3 to 4 ml/min/g ((8),22). These are perfusion rates at which a marked difference in uptake between tracers with high extraction fraction and low extraction fraction has been observed ((4),(5),17). By contrast, areas of the myocardium supplied by vessels with high-grade coronary stenosis have lower levels of maximal myocardial perfusion, which would manifest similar uptake between tracers with high- and low-extraction fractions (4). These factors would result in a greater magnitude of a stress perfusion defect with the more highly extracted agent. In regions supplied by a vessel with mild coronary stenosis, maximal myocardial perfusion would be expected to be only slightly reduced compared with those supplied by vessels without stenosis. This mild reduction in myocardial perfusion might be detectable with the more highly extracted tracer but might be insufficient to be detected by radiopharmaceuticals with lower extraction fraction (7), potentially contributing to the higher sensitivity observed with flurpiridaz F 18 than with the Tc-99m SPECT agents in this study. Improved spatial resolution of PET compared with SPECT and the use of attenuation correction with the PET studies might have contributed to the higher accuracy of flurpiridaz F 18 PET studies. In patients who had attenuation-corrected SPECT or both supine and prone SPECT imaging, only the supine SPECT imaging was assessed. Of note, the greater magnitude of reversible perfusion defect noted with flurpiridaz F 18 PET raises the possibility that reassessment of the threshold for ischemia used clinically to consider a study as indicating potential benefit from revascularization might have to be recalibrated in future studies.

Study limitations

This study has several limitations. Not all patients underwent ICA. The overall sensitivity of SPECT MPI in this study was generally lower than that reported in published clinical reports. These results are consistent with recent well-controlled SPECT MPI studies that implemented independent blinded read with multiple readers and have been conducted for regulatory purposes ((23),24). Several factors might have led to this low sensitivity in both imaging modalities. Patient selection is likely to have been a significant contributor. Because the majority of the patients underwent both SPECT and PET before ICA, it is possible that patients with greater amounts of ischemia by SPECT or with greater symptoms might have been more likely to have been not referred for this trial, due to the desire of the clinician or the patient not to have an additional stress test before ICA. Because patients with recent ICA but no coronary intervention were not excluded from the trial, patients undergoing the PET examinations after ICA are more likely to have had angiographic lesions of only borderline significance at ICA, thus potentially reducing the sensitivity of both SPECT and PET. Additionally, in individuals with prior MI, the use of 50% stenosis QCA criteria might have led to a misclassification of patients, because fixed perfusion defects in these patients would be expected even if the infarct-related artery had been successfully opened at a prior PCI (i.e., if vessels in patients with prior PCI did not show ≥50% stenosis, the patients were categorized as normal, even if the vessels were supplying a region of MI). In this regard, 48 patients had undergone prior revascularization >6 months before the MPI studies, and 26 had prior MI. The determination of the presence or absence of CAD was based on coronary stenosis in the native coronary arteries; arteries with patent bypass grafts but native coronary stenosis were classified as significant CAD. The acceptance of patients into this trial who completed SPECT before ICA and had the PET study after ICA might have biased the population toward lower specificity by SPECT. The use of the >50% stenosis by ICA as the criterion for abnormality might have included patients as abnormal for the truth standard who do not have flow-limiting stenosis, falsely lowering the sensitivity of stenosis detection. The absence of flow-limiting stenosis in a substantial proportion of patients with >50% stenosis has been documented in studies comparing angiographic stenosis with abnormalities of coronary flow reserve as measured invasively by fractional flow reserve (24). Higher spatial resolution associated with PET compared with SPECT imaging methods might have contributed to the results. Attenuation correction was applied in the PET studies and not in the SPECT studies. Although this is the largest group of patients to be studied with this novel perfusion agent, the number of patients is still relatively small, and this might affect the statistical significance of our results.

Flurpiridaz F 18 PET, when compared with Tc-99m SPECT, has better image quality and shows higher diagnostic certainty. In patients with coronary stenosis by ICA, flurpiridaz F 18 PET manifests higher sensitivity than SPECT for detection of patients with coronary stenosis as well as greater magnitude of ischemia. These findings suggest that this new tracer might significantly improve the assessment of patients with radionuclide MPI compared with the standard SPECT MPI methods.

Berman  D.S., Hachamovitch  R., Shaw  L.J.; Nuclear cardiology.:544-576.
Hachamovitch  R., Rozanski  A., Shaw  L.J.; Impact of ischaemia and scar on the therapeutic benefit derived from myocardial revascularization vs. medical therapy among patients undergoing stress-rest myocardial perfusion scintigraphy. Eur Heart J. 2011;32:1012-1024.
Hendel  R.C., Berman  D.S., Di Carli  M.F.; ACCF/ASNC/ACR/AHA/ASE/SCCT/SCMR/SNM 2009 appropriate use criteria for cardiac radionuclide imaging: a report of the American College of Cardiology Foundation Appropriate Use Criteria Task Force, the American Society of Nuclear Cardiology, the American College of Radiology, the American Heart Association, the American Society of Echocardiography, the Society of Cardiovascular Computed Tomography, the Society for Cardiovascular Magnetic Resonance, and the Society of Nuclear Medicine. J Am Coll Cardiol. 2009;53:2201-2229.
Glover  D.K., Ruiz  M., Edwards  N.C.; Comparison between 201Tl and 99mTc sestamibi uptake during adenosine-induced vasodilation as a function of coronary stenosis severity. Circulation. 1995;91:813-820.
Glover  D.K., Ruiz  M., Yang  J.Y., Smith  W.H., Watson  D.D., Beller  G.A.; Myocardial 99mTc-tetrofosmin uptake during adenosine-induced vasodilatation with either a critical or mild coronary stenosis: comparison with 201Tl and regional myocardial blood flow. Circulation. 1997;96:2332-2338.
Glover  D.K., Gropler  R.J.; Journey to find the ideal PET flow tracer for clinical use: are we there yet?. J Nucl Cardiol. 2007;14:765-768.
Maddahi  J.; Properties of an ideal PET perfusion tracer: new PET tracer cases and data. J Nucl Cardiol. 2012;19:S30-S37.
Lortie  M., Beanlands  R.S., Yoshinaga  K., Klein  R., Dasilva  J.N., DeKemp  R.A.; Quantification of myocardial blood flow with 82Rb dynamic PET imaging. Eur J Nucl Med Mol Imaging. 2007;34:1765-1774.
Mullani  N.A., Goldstein  R.A., Gould  K.L.; Myocardial perfusion with rubidium-82. J Nucl Med. 1983;24:898-906.
Schelbert  H.R., Phelps  M.E., Huang  S.C.; N-13 ammonia as an indicator of myocardial blood flow. Circulation. 1981;63:1259-1272.
Bergmann  S.R., Herrero  P., Markham  J., Weinheimer  C.J., Walsh  M.N.; Noninvasive quantitation of myocardial blood flow in human subjects with oxygen-15-labeled water and positron emission tomography. J Am Coll Cardiol. 1989;14:639-652.
Huisman  M.C., Higuchi  T., Reder  S.; Initial characterization of an 18F-labeled myocardial perfusion tracer. J Nucl Med. 2008;49:630-636.
Nekolla  S.G., Reder  S., Saraste  A.; Evaluation of the novel myocardial perfusion positron-emission tomography tracer 18F-BMS-747158-02: comparison to 13N-ammonia and validation with microspheres in a pig model. Circulation. 2009;119:2333-2342.
Maddahi  J.B.F., Huang  S.-C., Czernin  J., Schelbert  H., Zhu  Q.; Phase 1 rest-stress study of F-18 labeled BMS747158 myocardial perfusion PET tracer: human safety, dosimetry, biodistribution, and myocardial imaging characteristics. J Nucl Med. 2009;50:1
Maddahi  J., Czernin  J., Lazewatsky  J.; Phase I, first-in-human study of BMS747158, a novel 18F-labeled tracer for myocardial perfusion PET: dosimetry, biodistribution, safety, and imaging characteristics after a single injection at rest. J Nucl Med. 2011;52:1490-1498.
Yalamanchili  P., Wexler  E., Hayes  M.; Mechanism of uptake and retention of F-18 BMS-747158-02 in cardiomyocytes: a novel PET myocardial imaging agent. J Nucl Cardiol. 2007;14:782-788.
Yu  M., Guaraldi  M.T., Mistry  M.; BMS-747158-02: a novel PET myocardial perfusion imaging agent. J Nucl Cardiol. 2007;14:789-798.
Holly  T.A., Abbott  B.G., Al-Mallah  M.;American Society of Nuclear Cardiology,  Single photon-emission computed tomography. J Nucl Cardiol. 2010;17:941-973.
DeLong  E.R., DeLong  D.M., Clarke-Pearson  D.L.; Comparing the areas under two or more correlated receiver operating characteristic curves: a nonparametric approach. Biometrics. 1988;44:837-845.
Gibbons  R.J., Abrams  J., Chatterjee  K.; ACC/AHA 2002 guideline update for the management of patients with chronic stable angina—summary article: a report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines (Committee on the Management of Patients With Chronic Stable Angina). J Am Coll Cardiol. 2003;41:159-168.
Bateman  T.M., Heller  G.V., McGhie  A.I.; Diagnostic accuracy of rest/stress ECG-gated Rb-82 myocardial perfusion PET: comparison with ECG-gated Tc-99m sestamibi SPECT. J Nucl Cardiol. 2006;13:24-33.
Slomka  P.J., Alexanderson  E., Jacome  R.; Comparison of clinical tools for measurements of regional stress and rest myocardial blood flow assessed with 13N-ammonia PET/CT. J Nucl Med. 2012;53:171-181.
Senior  R., Monaghan  M., Main  M.L.; Detection of coronary artery disease with perfusion stress echocardiography using a novel ultrasound imaging agent: two Phase 3 international trials in comparison with radionuclide perfusion imaging. Eur J Echocardiogr. 2009;10:26-35.
Uren  N.G., Melin  J.A., De Bruyne  B., Wijns  W., Baudhuin  T., Camici  P.G.; Relation between myocardial blood flow and the severity of coronary-artery stenosis. N Engl J Med. 1994;330:1782-1788.

Figures

Grahic Jump Location
Figure 1

Summary of Patient Enrollment and Disposition

CAD = coronary artery disease; ICA = invasive coronary angiography; MPI = myocardial perfusion imaging; PET = positron emission tomography; SPECT = single-photon emission computed tomography.

Grahic Jump Location
Figure 2

Image Quality Comparing Flurpiridaz F 18 PET MPI and SPECT MPI (N = 132)

N = 132. Shown are the proportion of patients with excellent or good image quality (left) and excellent only (right). Abbreviations as in (Figure 1).

Grahic Jump Location
Figure 3

Diagnostic Certainty Comparing Flurpiridaz F 18 PET MPI and SPECT MPI

N = 132. Def. = definitely; Prob. = probably; other abbreviations as in (Figure 1).

Grahic Jump Location
Figure 4

Sensitivity and Specificity Comparing Flurpiridaz F 18 PET MPI and SPECT MPI for Detection of CAD

Abbreviations as in (Figure 1).

Grahic Jump Location
Figure 5

Receiver-Operating Characteristics Analysis

PET = positron emission tomography; SPECT = single-photon emission computed tomography.

Grahic Jump Location
Figure 6

FLUR PET and MIBI SPECT Images From a 75-Year-Old Man

Flurpiridaz F 18 (FLUR) PET (top) and Tc-99m sestamibi (MIBI) SPECT (bottom) images from a 75-year-old man, with a body mass index of 32.6 and no CAD on ICA. The FLUR images were interpreted as normal, whereas the MIBI images were interpreted as probably abnormal with a possible reversible defect in the basal inferior wall. Abbreviations as in (Figure 1).

Grahic Jump Location
Figure 7

FLUR PET and MIBI SPECT Images From an 82-Year-Old Man

The FLUR PET (top) and MIBI SPECT (bottom) images from an 82-year-old man with shortness of breath and an occluded native proximal left anterior descending (LAD) coronary artery and an occluded left internal mammary graft to the LAD and no other significant native CAD. The FLUR images show a severe reversible perfusion defect throughout the territory of the occluded proximal LAD, whereas the MIBI images show only a moderate perfusion defect in the distal LAD territory (apical slices). Abbreviations as in (Figure 1) and (Figure 6).

Tables

Table Grahic Jump Location
Table 1Patient Characteristics
Table Grahic Jump Location
Table 2Summary of Median SSS and SRS Among Readers by Number of Vessels With Coronary Stenosis
Table Grahic Jump Location
Table 3Summary of SDS in Patients With Angiography-Positive Results
Table Grahic Jump Location
Table 4Summary of SDS in Patients With Angiography-Negative Results

Interactive Graphics

Video

References

Berman  D.S., Hachamovitch  R., Shaw  L.J.; Nuclear cardiology.:544-576.
Hachamovitch  R., Rozanski  A., Shaw  L.J.; Impact of ischaemia and scar on the therapeutic benefit derived from myocardial revascularization vs. medical therapy among patients undergoing stress-rest myocardial perfusion scintigraphy. Eur Heart J. 2011;32:1012-1024.
Hendel  R.C., Berman  D.S., Di Carli  M.F.; ACCF/ASNC/ACR/AHA/ASE/SCCT/SCMR/SNM 2009 appropriate use criteria for cardiac radionuclide imaging: a report of the American College of Cardiology Foundation Appropriate Use Criteria Task Force, the American Society of Nuclear Cardiology, the American College of Radiology, the American Heart Association, the American Society of Echocardiography, the Society of Cardiovascular Computed Tomography, the Society for Cardiovascular Magnetic Resonance, and the Society of Nuclear Medicine. J Am Coll Cardiol. 2009;53:2201-2229.
Glover  D.K., Ruiz  M., Edwards  N.C.; Comparison between 201Tl and 99mTc sestamibi uptake during adenosine-induced vasodilation as a function of coronary stenosis severity. Circulation. 1995;91:813-820.
Glover  D.K., Ruiz  M., Yang  J.Y., Smith  W.H., Watson  D.D., Beller  G.A.; Myocardial 99mTc-tetrofosmin uptake during adenosine-induced vasodilatation with either a critical or mild coronary stenosis: comparison with 201Tl and regional myocardial blood flow. Circulation. 1997;96:2332-2338.
Glover  D.K., Gropler  R.J.; Journey to find the ideal PET flow tracer for clinical use: are we there yet?. J Nucl Cardiol. 2007;14:765-768.
Maddahi  J.; Properties of an ideal PET perfusion tracer: new PET tracer cases and data. J Nucl Cardiol. 2012;19:S30-S37.
Lortie  M., Beanlands  R.S., Yoshinaga  K., Klein  R., Dasilva  J.N., DeKemp  R.A.; Quantification of myocardial blood flow with 82Rb dynamic PET imaging. Eur J Nucl Med Mol Imaging. 2007;34:1765-1774.
Mullani  N.A., Goldstein  R.A., Gould  K.L.; Myocardial perfusion with rubidium-82. J Nucl Med. 1983;24:898-906.
Schelbert  H.R., Phelps  M.E., Huang  S.C.; N-13 ammonia as an indicator of myocardial blood flow. Circulation. 1981;63:1259-1272.
Bergmann  S.R., Herrero  P., Markham  J., Weinheimer  C.J., Walsh  M.N.; Noninvasive quantitation of myocardial blood flow in human subjects with oxygen-15-labeled water and positron emission tomography. J Am Coll Cardiol. 1989;14:639-652.
Huisman  M.C., Higuchi  T., Reder  S.; Initial characterization of an 18F-labeled myocardial perfusion tracer. J Nucl Med. 2008;49:630-636.
Nekolla  S.G., Reder  S., Saraste  A.; Evaluation of the novel myocardial perfusion positron-emission tomography tracer 18F-BMS-747158-02: comparison to 13N-ammonia and validation with microspheres in a pig model. Circulation. 2009;119:2333-2342.
Maddahi  J.B.F., Huang  S.-C., Czernin  J., Schelbert  H., Zhu  Q.; Phase 1 rest-stress study of F-18 labeled BMS747158 myocardial perfusion PET tracer: human safety, dosimetry, biodistribution, and myocardial imaging characteristics. J Nucl Med. 2009;50:1
Maddahi  J., Czernin  J., Lazewatsky  J.; Phase I, first-in-human study of BMS747158, a novel 18F-labeled tracer for myocardial perfusion PET: dosimetry, biodistribution, safety, and imaging characteristics after a single injection at rest. J Nucl Med. 2011;52:1490-1498.
Yalamanchili  P., Wexler  E., Hayes  M.; Mechanism of uptake and retention of F-18 BMS-747158-02 in cardiomyocytes: a novel PET myocardial imaging agent. J Nucl Cardiol. 2007;14:782-788.
Yu  M., Guaraldi  M.T., Mistry  M.; BMS-747158-02: a novel PET myocardial perfusion imaging agent. J Nucl Cardiol. 2007;14:789-798.
Holly  T.A., Abbott  B.G., Al-Mallah  M.;American Society of Nuclear Cardiology,  Single photon-emission computed tomography. J Nucl Cardiol. 2010;17:941-973.
DeLong  E.R., DeLong  D.M., Clarke-Pearson  D.L.; Comparing the areas under two or more correlated receiver operating characteristic curves: a nonparametric approach. Biometrics. 1988;44:837-845.
Gibbons  R.J., Abrams  J., Chatterjee  K.; ACC/AHA 2002 guideline update for the management of patients with chronic stable angina—summary article: a report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines (Committee on the Management of Patients With Chronic Stable Angina). J Am Coll Cardiol. 2003;41:159-168.
Bateman  T.M., Heller  G.V., McGhie  A.I.; Diagnostic accuracy of rest/stress ECG-gated Rb-82 myocardial perfusion PET: comparison with ECG-gated Tc-99m sestamibi SPECT. J Nucl Cardiol. 2006;13:24-33.
Slomka  P.J., Alexanderson  E., Jacome  R.; Comparison of clinical tools for measurements of regional stress and rest myocardial blood flow assessed with 13N-ammonia PET/CT. J Nucl Med. 2012;53:171-181.
Senior  R., Monaghan  M., Main  M.L.; Detection of coronary artery disease with perfusion stress echocardiography using a novel ultrasound imaging agent: two Phase 3 international trials in comparison with radionuclide perfusion imaging. Eur J Echocardiogr. 2009;10:26-35.
Uren  N.G., Melin  J.A., De Bruyne  B., Wijns  W., Baudhuin  T., Camici  P.G.; Relation between myocardial blood flow and the severity of coronary-artery stenosis. N Engl J Med. 1994;330:1782-1788.

Correspondence

Latest JACC CME

Continuing Medical Education through JACC is a convenient way to fulfill your CME requirements while learning important information about the latest advances in cardiovascular medicine.

April 2013- JACC CME Activity
Repeat Revascularization and Outcome

March 2013- JACC CME Activity
Extreme Lipoprotein(a) Levels and Improved Cardiovascular Risk Prediction

Feb 2013- JACC CME Activity
Results from the BARI 2D Trial

Jan 2013- JACC CME Activity
Prognosis Among Healthy Individuals Discharged With a Primary Diagnosis of Syncope

Dec 2012- JACC CME Activity
Incidence of Heart Failure or Cardiomyopathy After Adjuvant Trastuzumab Therapy for Breast Cancer

Nov 2012- JACC CME Activity
A Collaborative Analysis of Individual Patient Data From 10 Randomized Trials

Oct 2012- JACC CME Activity
Radiofrequency Ablation of Premature Ventricular Ectopy Improves the Efficacy of Cardiac Resynchronization Therapy in Nonresponders

Sept 2012- JACC CME Activity
Exercise and Pharmacological Treatment of Depressive Symptoms in Patients With Coronary Heart Disease

Aug 2012- JACC CME Activity
Reduction in Life-Threatening Ventricular Tachyarrhythmias in Statin-Treated Patients With Nonischemic Cardiomyopathy Enrolled in the MADIT-CRT (Multicenter Automatic Defibrillator Implantation Trial with Cardiac Resynchronization Therapy)

July 2012- JACC CME Activity
Relationship of Beta-Blocker Dose With Outcomes in Ambulatory Heart Failure Patients With Systolic Dysfunction

For previous CME quizzes, please follow this link to CardioSource Lifelong Learning and MOC.

 

NOTE:
Citing articles are presented as examples only. In non-demo SCM6 implementation, integration with CrossRef’s “Cited By” API will populate this tab (http://www.crossref.org/citedby.html).
Submit a Comment
Submit a Comment

Some tools below are only available to our subscribers or users with an online account.

Related Content

Customize your page view by dragging & repositioning the boxes below.

Articles Related By Topic
Related Topics
PubMed Articles