Updated Standardized Endpoint Definitions for Transcatheter Aortic Valve Implantation: The Valve Academic Research Consortium-2 Consensus Document†  FREE

A. Pieter Kappetein; Stuart J. Head; Philippe Généreux; Nicolo Piazza; Nicolas M. van Mieghem; Eugene H. Blackstone; Thomas G. Brott; David J. Cohen; Donald E. Cutlip; Gerrit-Anne van Es; Rebecca T. Hahn; Ajay J. Kirtane; Mitchell W. Krucoff; Susheel Kodali; Michael J. Mack; Roxana Mehran; Josep Rodés-Cabau; Pascal Vranckx; John G. Webb; Stephan Windecker; Patrick W. Serruys; Martin B. Leon
[+] Author Information

All rights reserved. © The Author 2012. For permissions please email: journals.permissions@oup.com. Grants have been provided to the ARC Board including representatives of Cardialysis, the Cardiovascular Research Foundation, Duke Clinical Research Institute, and Harvard Clinical Research Institute to cover the costs of travel, meeting rooms, and lodging for academic attendees at the Washington and Rotterdam meetings by Abbott Vascular, Boston Scientific, Direct Flow Medical, Edwards Lifesciences, Heart Leaflet Technologies, Medtronic Corporation, and St. Jude Medical. VARC Participants will provide Conflict of Interest Disclosures individually prior to publication. The VARC meetings involved members of the Interventional Cardiology Devices Branch, of the Office of Device Evaluation, Center for Devices and Radiological Health, USFDA. The opinions or assertions herein are the private views of the authors and are not to be construed as reflecting the views of the FDA.

Corresponding author: Tel: +31 0 10 70 35784, Fax: +31 0 10 70 32358

Copyright 2012, American College of Cardiology Foundation. All Rights Reserved.

J Am Coll Cardiol. 2012;60(15):1438-1454. doi:10.1016/j.jacc.2012.09.001
Published online

Objectives  The aim of the current Valve Academic Research Consortium (VARC)-2 initiative was to revisit the selection and definitions of transcatheter aortic valve implantation (TAVI) clinical endpoints to make them more suitable to the present and future needs of clinical trials. In addition, this document is intended to expand the understanding of patient risk stratification and case selection.

Background  A recent study confirmed that VARC definitions have already been incorporated into clinical and research practice and represent a new standard for consistency in reporting clinical outcomes of patients with symptomatic severe aortic stenosis (AS) undergoing TAVI. However, as the clinical experience with this technology has matured and expanded, certain definitions have become unsuitable or ambiguous.

Methods and Results  Two in-person meetings (held in September 2011 in Washington, DC, USA, and in February 2012 in Rotterdam, the Netherlands) involving VARC study group members, independent experts (including surgeons, interventional and non-interventional cardiologists, imaging specialists, neurologists, geriatric specialists, and clinical trialists), the US Food and Drug Administration (FDA), and industry representatives, provided much of the substantive discussion from which this VARC-2 consensus manuscript was derived. This document provides an overview of risk assessment and patient stratification that need to be considered for accurate patient inclusion in studies. Working groups were assigned to define the following clinical endpoints: mortality, stroke, myocardial infarction, bleeding complications, acute kidney injury, vascular complications, conduction disturbances and arrhythmias, and a miscellaneous category including relevant complications not previously categorized. Furthermore, comprehensive echocardiography recommendations are provided for the evaluation of prosthetic valve (dys)function. Definitions for the quality of life assessments are also reported. These endpoints formed the basis for several recommended composite endpoints.

Conclusions  This VARC-2 document has provided further standardization of endpoint definitions for studies evaluating the use of TAVI, which will lead to improved comparability and interpretability of the study results, supplying an increasingly growing body of evidence with respect to TAVI and/or surgical aortic valve replacement. This initiative and document can furthermore be used as a model during current endeavors of applying definitions to other transcatheter valve therapies (for example, mitral valve repair).

Figures in this Article

The first Valve Academic Research Consortium (VARC) consensus manuscript was published in January 2011 with the goal of achieving consensus for (i) selecting appropriate clinical endpoints reflecting device, procedure and patient-related effectiveness and safety, and (ii) standardizing definitions for single and composite clinical endpoints, for transcatheter aortic valve implantation (TAVI) clinical trials ((1),2). A recent pooled analysis, which included 3,519 patients from 16 unique studies, confirms that VARC definitions have already been incorporated into clinical and research practice and represent a new standard for consistency in reporting clinical outcomes of patients with symptomatic severe aortic stenosis (AS) undergoing TAVI (3). However, as the clinical experience with this technology has matured and expanded, certain definitions have become unsuitable or ambiguous ((3),(4),(5),(6),7). The aim of the current VARC was therefore to revisit the selection and definitions of TAVI-related clinical endpoints to make them more suitable to the present and future needs of clinical trials. In addition, this document is intended to expand the understanding of patient risk stratification and case selection.

Similar to the VARC-1 process, two in-person meetings (held in September 2011 in Washington, DC, USA, and in February 2012 in Rotterdam, the Netherlands) involving VARC study group members, independent experts (including surgeons, interventional and non- interventional cardiologists, imaging specialists, neurologists, geriatric specialists, and clinical trialists), the US Food and Drug Administration (FDA), and industry representatives, provided much of the substantive discussion from which this VARC-2 consensus manuscript was derived (see Appendixes).

Risk stratification of patients is crucial to identifying appropriate candidates for specific cardiac procedures. The EuroSCORE and Society of Thoracic Surgeons (STS) score are the most widely used risk scores to predict operative mortality in cardiac surgery. These models were developed and validated in a standard surgical risk population. The predictive power of both models is therefore suboptimal in high-risk patients with valvular disease, although the STS score has shown to outperform the Logistic EuroSCORE (8). These models are even more limited in application to patients who are considered at prohibitive risk for cardiac surgery, a cohort that could particularly benefit from TAVI. Current models could be improved by the addition of specific clinical and anatomical variables that affect mortality (9). As an example, the presence of a porcelain aorta and frailty are important factors not included in either risk model but are routinely considered during patient evaluation ((Figure 01_gr1), Table 1).

Grahic Jump Location
Figure 1

Porcelain Aorta or Severely Atherosclerotic Aorta

Table Grahic Jump Location
Table 1Risk factors Not Captured by Traditional Risk Scores

Perhaps the most important patient characteristic not included in current risk models is frailty (10). Frailty is frequently assessed subjectively based upon an informal ‘eyeball test’. However, physical performance assessments such as gait speed and grip strength are more objective performance measures that may capture an individual's overall functional status (11). These continuous measures are reproducible and can be re-assessed at various time points. In addition, they require no language translation. Assessments of cognition, weight (loss), activity level, and independence in the activities of daily living provide additional information on the overall health state of the individual (11). These limitations are more often found in patients with a high comorbidity burden and may co-exist with certain laboratory findings (e.g. low serum albumin, elevated inflammatory markers, anemia) that further reflect the health state and physiological reserve of the frail patient.

Baseline evaluation of the presence of cognitive dysfunction (mild cognitive impairment or dementia) has also emerged as an essential part of the initial risk stratification, especially in older populations, where the risk, benefit, and cost-effectiveness of invasive procedures must be weighed judiciously. Pre-procedural cognitive assessment may also help avoid attributing post-procedural mental status changes to stroke categories. Among the several clinically established rating scales [e.g. Mini-Mental State Examination, modified Telephone Interview of Cognitive Status (TICS-M), Clinical Dementia Rating Scale] (12), there is no particular standard for TAVI. Nevertheless, some systematic cognitive assessment by neuropsychological experts should be a part of the initial heart team evaluation.

(Table 1) provides an overview of these and other risk factors (Figs. (Figure 01_gr1) to Figure 01_)and VARC-2 recommendations on how each should be assessed. In clinical trials, it will be important to capture variables that predict extreme operative risk and to standardize the evaluation criteria and process. This will help to determine which subsets of patients are likely to benefit from TAVI treatment.

Grahic Jump Location
Figure 3

Patent IMA Graft Crossing Midline and/or Adherent to the Posterior Table of Sternum

Valve Academic Research Consortium-2 recommends the use of a heart team for patient evaluation. The heart team should consist of at least (interventional) cardiologists, cardiovascular surgeons, and imaging specialists, but its composition is dynamic and can also include anesthesiologists, geriatricians, neurologists, etc. This multi-disciplinary team should convene as a group on a regular basis to review and interpret clinical data to arrive at a consensus on the optimal treatment strategy for each patient. The heart team approach also allows for the adjustment of the decision-making process according to local experience and circumstances.

The heart team should agree on an estimated 30-day mortality risk for each patient based upon integrating a careful clinical assessment and utilizing appropriate risk prediction scoring systems, preferably the STS score. Surgical mortality risk strata are difficult to precisely assign, but an estimated 30-day mortality of <4% is considered low risk, 4–10% is intermediate risk, >10% is high risk, and >15% is very high risk. A patient is considered at extreme risk if at least two cardiovascular surgeons from a tertiary centre of excellence deny surgery because of prohibitive operative risks, estimated to be a combined >50% risk of irreversible morbidity or mortality (13). In addition to the specific risk factors that can prohibit patients from undergoing TAVI or surgical aortic valve replacement (SAVR) (Table 1), the operative risk assessment is also important to identify patients who are likely not to benefit from either TAVI or SAVR (the so-called ‘futility’ category of high-risk patients). An expected improvement in the quality of life (QOL) may further be necessary to identify treatment responders vs. non- responders. Individualized life expectancy assumptions should be incorporated by the heart team in the clinical decision-making process as a central factor in weighing the risk-benefit ratio. Prognostic indices of life expectancy may play a central role in moving beyond arbitrary age-based cut-offs (14).

The most important role of the heart team is to provide customized management decisions for common and unusual clinical scenarios in terms of patient selection, procedural performance, and complication management. An example is the frequent situation of severe AS and concomitant coronary artery disease (CAD). The complexity of CAD and appropriate revascularization strategies in the setting of AS should be determined by consensus from interventional cardiologists and cardiovascular surgeons ((15),16) In new TAVI clinical trials, angiographic risk scores (e.g. SYNTAX score) may be utilized to help determine the complexity of CAD, as a basis for the inclusion in the trial. Thresholds for coronary revascularization and the choice for a staged or concomitant PCI with TAVI should be guided by the complexity of the CAD and other factors as determined by the heart team ((17),18). In general, the plan to deal with other co-existing conditions [such as atrial fibrillation (AF), other valvular lesions, and other congenital lesions] should be pre-specified and all complications encountered in the treatment of associated conditions (including treatment after the TAVI procedure) should be captured. Such thorough pre-procedural assessment is also valuable in discriminating new post-procedural complications from simple exacerbations of pre-existing conditions.

Mortality

In addition to the original VARC definitions, VARC-2 recommends the collection of immediate procedural mortality to capture intra-procedural events that result in immediate or consequent death <72 h post-procedure. Taking into account the surgical literature, procedural mortality consists of all-cause mortality within 30 days or during index procedure hospitalization—if the postoperative length of stay is longer than 30 days.

The cause of death should be captured, based on a careful review of narrative summaries and source material. All-cause, cardiovascular, and non-cardiovascular mortality should be reported after 30 days during the follow-up (Table 2). In determining the cause of death, the adjudication committee should consider the clinical context at the time of the index procedure and during the time interval leading up to death. All efforts (including the use of national death registries) should be made to identify, precisely characterize, and appropriately classify any death.

Table Grahic Jump Location
Table 2Mortality
Myocardial Infarction

Myocardial injury as determined by a significant rise in cardiac biomarkers occurs frequently following TAVI, and a significant magnitude of myocardial injury has been associated with worse outcomes (19). Valve Academic Research Consortium-2 recommends the systematic collection of biomarkers of myocardial injury prior to the procedure, within 12–24 h after the procedure, at 24 h thereafter, at 72 h or at discharge, and, if still elevated, daily until values show a decline. Similar to the previous VARC recommendations, the definition of peri-procedural (<72 h following TAVI) MI will be based on a combination of clinical criteria and cardiac biomarkers. However, the threshold values have been adjusted (Table 3). Acute ischemic events occurring after 72 h should be considered spontaneous myocardial infarctions and defined in accordance with the universal MI guidelines (20).

Table Grahic Jump Location
Table 3Myocardial Infarction
Stroke

With increasing attention to stroke as an important peri-procedural complication of TAVI (21), the FDA has emphasized the need for an accurate assessment of stroke and has participated actively in recommending specific details of the VARC-2 definitions. In an attempt to further align with the fundamental definitions now endorsed by the FDA (22), consensus was reached at VARC-2 to further refine the definition of stroke and recommend the use of these definitions in future TAVI clinical trials (Table 4). The definitions endorsed by the FDA are intended to apply to a wide range of clinical trials and to enable those trials to assess the clinically relevant consequences of vascular brain injury for determining the safety or effectiveness of an intervention.

Table Grahic Jump Location
Table 4Stroke and TIA

Stroke is defined as an acute episode of focal or global neurological dysfunction caused by the brain, spinal cord, or retinal vascular injury as a result of hemorrhage or infarction. Stroke may be classified as ischemic or hemorrhagic with appropriate subdefinitions. Ischemic stroke is defined as an acute episode of focal cerebral, spinal, or retinal dysfunction caused by infarction of central nervous system tissue. Hemorrhagic stroke is defined as an acute episode of focal or global cerebral or spinal dysfunction caused by intraparenchymal, intraventricular, or subarachnoid hemorrhage. A stroke may be classified as ‘undetermined’ if there is insufficient information to allow the categorization as ischemic or hemorrhagic.

An entity closely related to an ischemic stroke that should be assessed is a transient ischemic attack (TIA). Transient ischemic attack is defined as a transient episode of focal neurological dysfunction caused by the brain, spinal cord, or retinal ischemia, without acute infarction. The difference between TIA and ischemic stroke is the presence of tissue damage on neuro-imaging studies or new sensory-motor deficit persisting >24 h. By definition, a TIA does not produce a lasting disability.

Valve Academic Research Consortium-2 recognizes that an assessment of stroke is incomplete without an appropriate measurement of the disability resulting from the stroke. Valve Academic Research Consortium-2 recommends the use of the modified Rankin Scale (mRS) to assess this clinical disability ((23),(24),25). The assessment of the mRS should occur at all scheduled visits in a trial and at 90 days after the onset of any stroke. This approach will maximize the detection of new or recurrent strokes, assist in the ongoing evaluation of events previously determined as TIAs, and provide an accepted and reliable indicator of the long-term impact of a given stroke.

Previously, VARC recommended categorizing strokes as ‘major’ and ‘minor’ based upon mRS scores. To enhance the accuracy in the description of a given stroke and to provide accurate categorization of strokes within a given trial, VARC-2 now recommends the use of the terms ‘disabling’ and ‘non-disabling.' A disabling stroke is one that results (at 90 days after stroke onset) in an mRS score of >2 and an increase in >1 mRS category from an individual's pre-stroke baseline. A non-disabling stroke is one that results (at 90 days after stroke onset) in an mRS score of <2 or that does not result in an increase in >1 mRS category from an individual's pre-stroke baseline. In addition to this categorization of disabling and non-disabling strokes, the endpoint of all strokes should be reported.

Although brain imaging (typically, MRI for acute and chronic ischemia and hemorrhage, and CT for acute and chronic hemorrhage and chronic ischemia) is often used to supplement the clinical diagnosis of stroke (26), a diagnosis of stroke may be made on clinical grounds alone. Valve Academic Research Consortium-2 recognizes that stroke symptoms are protean and not well suited to a pre-specified itemized listing. Accordingly, VARC-2 recommends that a vascular neurologist experienced in clinical trials involving stroke be included in all phases of trial planning, execution, and monitoring, including involvement in the Clinical Events Committee and the Data and Safety Monitoring Board.

New insights into the timing of events show delayed or late occurrence of strokes, beyond the early post-implantation phase (27). This may suggest that the cause of stroke is additionally related to other factors or patient susceptibilities and should necessitate active investigation of devices and adjunctive pharmacotherapy to reduce the frequency and severity of strokes after TAVI, including precise documentation of the use and dosage of antithrombotic and antiplatelet medication. Patient baseline characteristics (e.g. carotid stenosis) and postoperative complications (e.g. AF) need to be carefully documented to be able to identify the contributing causes of stroke.

Invasive stroke management (catheter-based intracranial intervention) is gaining an increasingly important role and may impact morbidity and mortality. Valve Academic Research Consortium-2 therefore recommends the ascertainment of any acute stroke management strategy (e.g. aspiration, thrombolysis, or conservative management).

Bleeding Complications

Valve Academic Research Consortium-2 acknowledges the fact that the Bleeding Academic Research Consortium (BARC) recently convened and established standardized bleeding definitions for patients receiving antithrombotic therapy and undergoing coronary revascularization (PCI or CABG) ((28),29). However, because the current definitions have been well adopted and shown to be accurate in predicting adverse events (30), VARC-2 has chosen to maintain the original VARC definitions with BARC classifications (Table 5), recognizing that future validation of BARC criteria in this population may warrant revision of the current recommendations.

Table Grahic Jump Location
Table 5Bleeding

With respect to blood transfusions, it is critical to acknowledge that a bleeding complication has to be the result of overt bleeding and cannot be adjudicated based on blood transfusions alone.

Acute Kidney Injury

The original VARC definitions recommended the use of a modified version of the RIFLE classification. However, we now recommend using the AKIN system (Table 6), which is a modified version of RIFLE that has been adopted by many in the nephrology community, including the KDIGO initiative ((31),21). As a result, acute kidney injury (AKI) can also be diagnosed according to urine output measures (Table 6).

Table Grahic Jump Location
Table 6Acute Kidney Injury (AKIN ClassificationTable fn7)

In comparison with the original VARC, the timing for the diagnosis of AKI is extended from 72 h to 7 days. Patients who experience AKI should have follow-up renal function assessments after 7 days until stabilization.

Vascular Complications

(Table 7) lists VARC-2 definitions for major and minor vascular complications. Further clarifications of these definitions to supplement the original VARC document are as follows. Pre-planned surgical access or a planned endovascular approach to vascular closure (e.g. ‘pre- closure') ((33),34) should be considered as part of the TAVI procedure and not as a complication, unless untoward clinical consequences are documented (e.g. bleeding complications, limb ischemia, distal embolization, or neurological impairment). Unplanned endovascular stenting or surgical repair for any vascular complications during the index procedure without other clinical sequelae should be considered a minor vascular complication, except if associated with qualifying consequences (Table 7). Complications related to alternative access sites, including the left-ventricular apex, subclavian artery, or aorta should be systematically recorded. To ensure accurate capture of these elements, VARC-2 strongly recommends that detailed information regarding the access site and pre-planned vascular closure technique be recorded as well as the use of any additional unplanned access or closure techniques (surgical repair, endovascular stenting, or endovascular balloon therapy). Since many vascular complications will also result in a bleeding complication, events that meet VARC-2 definitions for both categories should be reported in both categories. Finally, VARC-2 recommends that all vascular complications be recorded as either access (e.g. iliac rupture) or non-access site-related (e.g. ascending aorta dissection or rupture unless aortic access is used and the event originates from the cannulation site).

Table Grahic Jump Location
Table 7Vascular Access Site and Access-Related Complications
Conduction Disturbances and Arrhythmias

Valve Academic Research Consortium-2 proposes the systematic collection of data on the frequency of implant-related new and/or worsened conduction disturbances and the incidence and indication for permanent pacemaker implantation (Table 8). In addition, the frequency of specific arrhythmias following TAVI should be recorded as they may result in prolonged hospitalization and impaired clinical outcomes. New-onset AF (or flutter) is diagnosed as any arrhythmia within hospitalization that has the ECG characteristics of AF and lasts sufficiently long to be recorded on a 12-lead ECG, or for at least 30 s on a rhythm strip (35). The therapeutic approach to new-onset AF (spontaneous conversion, electrical or medical cardioversion, initiation of oral anticoagulation, and rate or rhythm control medications) and any clinical consequences should be thoroughly documented in the case report form.

Table Grahic Jump Location
Table 8Conduction Disturbances and Arrhythmias
Other TAVI-Related Complications

The original VARC document recommended the collection of a number of TAVI-related complications, but did not provide specific endpoint definitions for several endpoints. Valve Academic Research Consortium-2 recommends reporting any other complications related to the TAVI procedure, even those occurring less frequently, and provides formal VARC-2 definitions (Table 9) ((36),(37),38).

Table Grahic Jump Location
Table 9Other TAVI-Related Complications
Additional Considerations

For studies or trials where the occurrence, prevention, or treatment of cerebral infarction is a fundamental feature (e.g. embolic protection devices) additional appropriate imaging in all or a subset of patients may be necessary to allow determination of effectiveness.

Valve Academic Research Consortium-2 maintains the original recommendations to use echocardiography as the primary imaging modality for the assessment of prosthetic valve function (39). This should include the valve position, morphology, function, and evaluation of the left ventricle (LV) and right ventricle (RV) size and function. The suggested time points for routine follow-up transthoracic echocardiography (TTE) following valve implantation are: immediately (before discharge) following the implantation for transarterial approaches or within 30 days for transapical or transaortic approaches, 6 months following implantation, 1 year following implantation, and yearly thereafter. At these endpoints, prosthetic aortic valve stenosis and regurgitation should be reported.

Transcatheter Valve Stenosis

The assessment of prosthetic valve stenosis should be an integrative process utilizing multiple parameters of valve function. (Table 10) outlines the primary parameters used for assessing prosthetic valve function based on published guidelines (40). Divergence from the guidelines is based on a number of studies ((41),42), as well as methods used in large randomized control trials of TAVI ((43),44). In addition, VARC-2 does not recommend using acceleration time, which is dependent on ventricular function and heart rate (42). The limitation of flow-dependent parameters such as peak jet velocity or mean transprosthetic gradient is obvious, however, even flow-independent parameters such as the effective orifice area (EOA) and the Doppler velocity index (DVI) have limitations: (i) the absolute EOA does not account for the cardiac output requirements in relation to the patient's body size; thus lower criteria should be used to define prosthetic valve stenosis in patients with BSA <1.6 m2 (Table 10), (ii) the indexed EOA may overestimate the valve-related hemodynamic burden in obesity; hence, lower criteria may be more appropriate in patients with a body mass index >30 kg/m2, (iii) DVI severity criteria are dependent on the left ventricular outflow tract (LVOT) size; thus a lower threshold may be more appropriate in patients with LVOT diameters of >25 mm. The EOA should generally be calculated with the use of the LVOT diameter and the velocity measured just underneath the apical margin of the valve stent ((45),46). In cases where the landing zone of the stent is low in the LVOT, the diameter and velocity may both be measured in the proximal portion of the stent. Unlike the surgically implanted valve, the transcatheter prosthetic valve EOA is defined not only by the size of the valve but also by the patient's aortic valve/ annular anatomy and procedural variables. Thus, well-established normal transcatheter valve gradients and EOAs based on pre-implant aortic annular dimensions do not currently exist. Clinicians should be aware of this variability when assessing a patient for transcatheter valve function and VARC-2 strongly recommends that the patient's own initial post-implant study be used as a reference for serial comparisons.

Table Grahic Jump Location
Table 10Prosthetic Valve Dysfunction

The assessment of transcatheter valve dysfunction includes the immediate post-TAVI hemodynamics and the follow-up evaluation. The immediate post-TAVI evaluation documents initial valve appearance (position and circularity of the stent, and leaflet morphology and motion) and a comprehensive hemodynamic evaluation. Valve Academic Research Consortium-2 advocates using the integrative approach outlined in the algorithm shown in (Figure 01_gr4) as part of a comprehensive hemodynamic evaluation by initially using one flow dependent (e.g. mean gradient) and one flow independent criterion (e.g. EOA) for the initial hemodynamic evaluation. If there is discordance between these measurements, then the DVI should be calculated. An abnormal DVI indicates possible prosthetic valve dysfunction. A normal DVI indicates intrinsically normal prosthetic valve function, and the indexed EOA can then be used to determine the reason for the initial measurement discordance. When the indexed EOA is low in the setting of a normal DVI, the patient probably has a prosthesis-patient mismatch (PPM), an indicator of the intrinsic relationship of the implanted valve to the cardiac output requirements of the patient (47). Prosthesis-patient mismatch occurs in the setting of a morphologically normal valve and is considered to be hemodynamically insignificant if the indexed EOA is >0.85 cm2/m2, moderate if between 0.65 and 0.85 cm2/m2, and severe if <0.65 cm2/m2. However, for obese patients (body mass index >30 kg/m2) lower criteria may be more appropriate (Table 10).

Grahic Jump Location
Figure 4

Transcatheter Heart Valve Haemodynamic Evaluation Algorithm

Transcatheter Valve Regurgitation

There is growing evidence suggesting a significant association of post-procedural paravalvular aortic regurgitation (AR) with short- and long- term mortality ((48),49). As the duration of implanted transcatheter heart valves increases, valve durability and dysfunction become more crucial issues. Evaluating the presence and severity of regurgitation should include an assessment of both central and paravalvular components, with a combined measurement of 'total' aortic regurgitation (AR) reflecting the total volume load imposed on the LV (Table 10). The quantitative and semi-quantitative hemodynamic assessment of AR severity should be performed with Doppler echocardiography according to the guidelines ((39),(50),51). Color Doppler evaluation should be performed just below the valve stent for paravalvular jets, and at the coaptation point of the leaflets for central regurgitation. Although all imaging windows should be used, the parasternal short-axis view is critical in assessing the number and severity of paravalvular jets. Whenever possible, the quantification of the prosthetic regurgitant volume, effective regurgitant orifice area, and regurgitant fraction (Table 10) should be performed ((40),(51),52). The regurgitant volume may be calculated as the difference between the stroke volume across any non-regurgitant orifice (RVOT or mitral valve) and the stroke volume across the LVOT.

It is important to realize that at this time the body of evidence supporting the numerical criteria used in (Table 10) as well as (Figure 01_gr4) may be limited. These criteria should be used as guidelines for clinical decision-making and require further validation as our experience continues to expand.

Follow-Up Assessments

The follow-up assessment should also begin with valve imaging and documentation of changes in morphology. When determining whether a patient has developed hemodynamically significant structural valve failure, the patient's own baseline echocardiographic parameters should be used as a reference. An increase in the mean gradient >10 mm Hg, a decrease in the EOA >0.3–0.4 cm2, or a reduction in the DVI >0.1–0.13 probably indicates a change in valve function and should trigger a comprehensive hemodynamic evaluation. Whenever valve dysfunction is suspected, the careful evaluation of valve morphology should confirm a structurally abnormal valve. In addition, measurement error must be excluded; the use of a consistent LVOT diameter for more accurate follow-up study comparisons is recommended. Finally, changes in ventricular morphology would be expected in the setting of long-standing significant valvular dysfunction and this parameter may support the clinical assessment of severity.

Although the rate of moderate or severe regurgitation may appear to be less at the follow-up, this may be the result of attrition of the sickest patients. To assess such time trends, it is recommended to report an individual patient's progression of regurgitation, in a table that provides changes between short-term and long-term regurgitation, including mortality (48).

Quality of Life Evaluation in Aortic Stenosis

New York Heart Association (NYHA) classification is limited by the discrete nature of the scale, which provides only modest resolution to detect clinically relevant changes. Moreover, since the NYHA class is assessed by an external body rather than the patient, it does not reflect the patient's perspective. Thus, the NYHA class is more properly considered a measure of the functional status than the QOL.

The Minnesota Living with Heart Failure Questionnaire (MLHF) (53) and the Kansas City Cardiomyopathy Questionnaire (KCCQ) ((54),55) have a number of desirable properties for the evaluation of health-related QOL (HRQOL) in the setting of AS. Both instruments produce outcomes on a continuous scale, which improves responsiveness and sensitivity. Although only the MLHF has been specifically validated in patients with aortic valve disease (56), preliminary experience with the KCCQ in patients undergoing TAVI has also demonstrated a high degree of responsiveness and internal consistency (57).

Recommended Endpoints and Timing of Assessment

Valve Academic Research Consortium-2 recommends that a comprehensive assessment of HRQOL for patients undergoing TAVI incorporate both a heart failure-specific measure (such as the KCCQ or MLHF) as well as one or more generic measures [such as the Medical Outcomes Study Short-Form 36 (SF-36), the Short-Form 12 (SF-12), or the EuroQOL (EQ-5D)] ((58),(59),60). The disease-specific measures offer improved sensitivity/responsiveness as well as clinical interpretability, whereas the inclusion of a generic health status measure is useful because it captures some additional domains. Furthermore, generic measures can enhance the comparability across different diseases and populations and can be used to compare patients with population-level benchmarks.

For the comparison of TAVI vs. SAVR (or for the comparison of alternative access sites for TAVI), we recommend that early QOL assessment be performed at 2 weeks, 1 month, and 3 months using a combination of generic instruments and pain scales (e.g. visual analogue scale) to assess the early recovery process. The evaluation of the QOL at an intermediate time point (e.g. 6 months) could also be considered in order to confirm that QOL recovery is complete by this stage. At later time points (1–5 years), the use of heart failure-specific instruments to identify the consequences of long-term valve performance may be more useful. Finally, the assessment of cognitive function at later time points (1–5 years) may be valuable for the comparison of surgical vs. catheter-based techniques, although these endpoints generally require highly specialized and demanding neuropsychiatric testing (61). In contrast, for the comparison of alternative TAVI systems (as may be expected in the near future), HRQOL assessment should focus mainly on heart failure-specific endpoints at intermediate and later time points (1–5 years), wherein between-device differences in the hemodynamic performance or structural valve deterioration may emerge. The inclusion of disease-specific QOL measures in these studies can also provide insight into the consequences of valve-related complications such as the need for pacemaker insertion.

Additional Considerations

It is essential to ensure complete ascertainment of HRQOL at each time point, as missing data cannot be retrieved retrospectively and statistical adjustment techniques (e.g. multiple imputation) that assume that data are 'missing at random' may not be adequate. Differential mortality between two treatments may complicate the interpretation of QOL results since the QOL may appear to 'improve' over time even with an ineffective therapy simply because of attrition of the sickest patients. The use of categorical endpoints that characterize outcomes as favorable (e.g. survival AND improvement of QOL endpoints) ((44),57) or endpoints that integrate survival and the QOL (e.g. quality-adjusted life expectancy) may provide more interpretable results. In such cases, reporting the outcomes in both ways (i.e. among the entire study cohort and separately among only the surviving patients) will provide the most complete description of the results.

Rationale and Caveats

Comparisons of the success, safety, and effectiveness with achievable study cohort sample sizes may at times require the use of composite endpoints. However, it is important that composites contain components that have roughly similar impacts on the patient. A family of single endpoints tending in the same direction may, as a family of hypotheses, be statistically significant when individual endpoints are not.

Each post-procedural event has a different temporal risk profile (hazard function) modulated by different risk factors. Therefore, traditionally, the evaluation of the safety and efficacy of procedures has focused on in-hospital events (complications and morbidity), events within 30 days of the procedure, and ‘late’ events.

Specific Composite Endpoints

The assessment of TAVI, SAVR, and their alternatives or new devices should include device, procedure, and patient-oriented endpoints. These endpoints have been devised to be applicable to both TAVI and SAVR. Previous clinical trials have used the all-cause mortality at 1 year as the primary clinical endpoint. Owing to the emergence of stroke as an important clinical event, future trials should also require the composite of all-cause mortality and disabling stroke as a primary or secondary endpoint.

The first VARC document proposed three composite endpoints: device success, early safety, and clinical efficacy. Valve Academic Research Consortium-2 goes beyond the early and intermediate experience of TAVI, drawing upon prior surgical AVR guidelines to include time-related safety endpoints (62). Therefore, VARC-2 recommends a new composite endpoint, time-related valve safety, which combines valve dysfunction, endocarditis, and thrombotic complications of the prosthesis (Table 11).

Table Grahic Jump Location
Table 11Composite Endpoints

Although the original VARC standardized endpoint definitions were fundamentally useful and have been widely adopted, growing experience with TAVI studies has identified some definitions as ambiguous, of limited clinical utility, or in need of updating or extension ((5),(6),(63),64). This need provided the rationale for a VARC-2 document with such improvements and additions. As was the case with the original VARC process, it should be emphasized that this consensus manuscript is not intended to be a guidelines document, but rather a practical tool to facilitate and inform clinical research in TAVI.

Current clinical trials are focusing more on intermediate risk patients, and more studies are comparing TAVI with surgical AVR. Therefore, it becomes increasingly important to identify those patients who benefit from either treatment. Specific risk categories have been defined to allow universal clinical study designs and outcome comparisons.

Changes and additions that have been applied to improve the interpretation of clinical endpoint definitions and provide further insights on TAVI-related outcomes are as follows: (i) risk stratification should be done by a dedicated ‘heart team’ and include other factors (e.g. frailty, porcelain aorta) beyond the traditional risk scores, and should take into account co-existing conditions; (ii) immediate procedural mortality has been added to capture intra-procedural events that result in immediate or consequent death; (iii) stroke ascertainment requires the use of precise definitions, standardized assessments, close collaboration with neurology experts including the consideration of acute stroke management, and has been re-categorized as non-disabling or disabling; (iv) detailed documentation of the etiology of strokes and concomitant therapies is needed to provide insights into the multi-factorial nature of acute, early, and late strokes; (v) closure device failure is now a separate category within vascular complications, and if unplanned percutaneous or surgical intervention does not lead to adverse outcomes, these are not considered as a major vascular complication per se; (vi) the time for AKI diagnosis has been extended from 72 h to 7 days; (vii) AKI is diagnosed according to AKIN guidelines, which include classification by the urine output to detect a wider range of etiologies; (viii) peri-procedural myocardial infarction is defined by troponin or CK-MB elevation and the troponin threshold has changed from 10× ULN to 15× ULN based on recent data (19); (ix) assessment of conduction disturbances and arrhythmias has been reinforced ((65),(66),(67),68); (x) new definitions for several TAVI-related complications and valve malpositioning are reported; (xi) echocardiography parameters of prosthetic valve stenosis and regurgitation have been updated and now include the assessment of the prosthesis-patient mismatch; (xii) for the QOL assessment, VARC-2 recommends the use of both heart failure-specific and generic measures during the follow-up between 30 days and 5 years to fully assess the impact of the procedure and the durability of clinical benefit. These definitions can be used in studies comparing TAVI to surgical AVR, as well as in future trials comparing first generation to next generation TAVI devices.

The composite endpoint of device success has specifically been criticized for being too strict with regard to valve performance; for example, an AVA >1.2 cm2 seems unachievable in patients with smaller body habitus (5). The current VARC-2 definition therefore corrects for the body surface area so that valve performance is now assessed through the indexed EOA. It is notable that valve-in-valve procedures for failing bioprostheses will frequently have a low device success, even with this modified definition (69). Considering that stroke has emerged as an important concern, the composite of all-cause mortality and disabling stroke should be considered as a primary or secondary endpoint in future trials. Two ongoing large randomized trials [PARTNER II (NCT01314313) and SURTAVI (NCT01586910)] are already incorporating these composite endpoints.

With longer follow-up duration, it becomes more critical to include time-related valve safety composite endpoints. This will eventually provide linearized rates of complications with transcatheter valves, known as ‘objective performance criteria’, as has been used to evaluate surgical valves (70).

With this VARC-2 document, we have provided further standardization of endpoint definitions and hope that the adoption of these criteria will continue to increase, ultimately leading to improved comparability and interpretability of the study results. (32)

†The Valve Academic Research Consortium (VARC) consists of representatives from several independent Academic Research Organization, several Surgery and Cardiology Societies, members of the U.S. Food and Drug Administration (FDA), and several independents experts. However, it is not a society document. Neither the societies nor the FDA have been asked to endorse the document.

Leon  M.B., Piazza  N., Nikolsky  E.; Standardized endpoint definitions for transcatheter aortic valve implantation clinical trials: a consensus report from the Valve Academic Research Consortium. J Am Coll Cardiol. 2011;57:253-269.
PubMed | CrossRef
Leon  M.B., Piazza  N., Nikolsky  E.; Standardized endpoint definitions for transcatheter aortic valve implantation clinical trials: a consensus report from the Valve Academic Research Consortium. Eur Heart J. 2011;32:205-217.
Généreux  P., Head  S.J., Van Mieghem  N.M.; Clinical outcomes after transcatheter aortic valve replacement using Valve Academic Research Consortium definitions: a weighted meta-analysis of 3,519 patients from 16 studies. J Am Coll Cardiol. 2012;59:2317-2326.
Clavel  M.A., Rodes-Cabau  J., Dumont  E.; Validation and characterization of transcatheter aortic valve effective orifice area measured by Doppler echocardiography. J Am Coll Cardiol Img. 2011;4:1053-1062.
Ikeda  K., Ho  M., Kawahara  M.; Valve academic research consortium consensus report the pharmaceutical and medical devices agency perspective. J Am Coll Cardiol. 2011;58:777
Gurvitch  R., Toggweiler  S., Willson  A.B.; Outcomes and complications of transcatheter aortic valve replacement using a balloon expandable valve according to the Valve Academic Research Consortium (VARC) guidelines. Eurolntervention. 2011;7:41-48.
Delgado  V., Schuijf  J.D., Bax  J.J.; Pre-operative aortic valve implantation evaluation: multimodality imaging. Eurolntervention. 2010;6:G38-G47.
Piazza  N., Wenaweser  P., van Gameren  M.; Relationship between the logistic EuroSCORE and the Society of Thoracic Surgeons Predicted Risk of Mortality score in patients implanted with the CoreValve ReValving system—a Bern-Rotterdam Study. Am Heart J. 2010;159:323-329.
Rosenhek  R., Iung  B., Tornos  P.; ESC Working Group on Valvular Heart Disease Position Paper: assessing the risk of interventions in patients with valvular heart disease. Eur Heart J. 2012;33:822-828.
Afilalo  J., Mottillo  S., Eisenberg  M.J.; Addition of frailty and disability to cardiac surgery risk scores identifies elderly patients at high risk of mortality or major morbidity. Circ Cardiovasc Qual Outcomes. 2012;5:222-228.
Pal  S.K., Katheria  V., Hurria  A.; Evaluating the older patient with cancer: understanding frailty and the geriatric assessment. CA Cancer J Clin. 2010;60:120-132.
Morris  J.C.; The clinical dementia rating (CDR): current version and scoring rules. Neurology. 1993;43:2412-2414.
Holmes  D.R., Mack  M.J., Kaul  S.; 2012 ACCF/AATS/SCAI/STS expert consensus document on transcatheter aortic valve replacement: developed in collaboration with the American Heart Association, American Society of Echocardiography, European Association for Cardio-Thoracic Surgery, Heart Failure Society of America, Mended Hearts, Society of Cardiovascular Anesthesiologists, Society of Cardiovascular Computed Tomography, and Society for Cardiovascular Magnetic Resonance. Ann Thorac Surg. 2012;93:1340-1345.
Yourman  L.C., Lee  S.J., Schonberg  M.A., Widera  E.W., Smith  A.K.; Prognostic indices for older adults: a systematic review. JAMA. 2012;307:182-192.
Sianos  G., Morel  M.A., Kappetein  A.P.; The SYNTAX Score: an angio- graphic tool grading the complexity of coronary artery disease. EuroIntervention. 2005;1:219-227.
Serruys  P.W., Morice  M.C., Kappetein  A.P.; Percutaneous coronary intervention versus coronary-artery bypass grafting for severe coronary artery disease. N Engl J Med. 2009;360:961-972.
Dewey  T.M., Brown  D.L., Herbert  M.A.; Effect of concomitant coronary artery disease on procedural and late outcomes of transcatheter aortic valve implantation. Ann Thorac Surg. 2010;89:758-767.
Wenaweser  P., Pilgrim  T., Guerios  E.; Impact of coronary artery disease and percutaneous coronary intervention on outcomes in patients with severe aortic stenosis undergoing transcatheter aortic valve implantation. EuroIntervention. 2011;7:541-548.
Rodes-Cabau  J., Gutierrez  M., Bagur  R.; Incidence, predictive factors, and prognostic value of myocardial injury following uncomplicated transcatheter aortic valve implantation. J Am Coll Cardiol. 2011;57:1988-1999.
Thygesen  K., Alpert  J.S., White  H.D.; Joint ESC/ACCF/AHA/WHF Task Force for the Redefinition of Myocardial Infarction. Eur Heart J. 2007;28:2525-2538.
Daneault  B., Kirtane  A.J., Kodali  S.K.; Stroke associated with surgical and transcatheter treatment of aortic stenosis: a comprehensive review. J Am Coll Cardiol. 2011;58:2143-2150.
Brott  T., Adams  H.P., Olinger  C.P.; Measurements of acute cerebral infarction: a clinical examination scale. Stroke. 1989;20:864-870.
Bonita  R., Beaglehole  R.; Recovery of motor function after stroke. Stroke. 1988;19:1497-1500.
Lyden  P.D., Lau  G.T.; A critical-appraisal of stroke evaluation and rating-scales. Stroke. 1991;22:1345-1352.
Wintermark  M., Albers  G.W., Alexandrov  A.V.; Acute stroke imaging research roadmap. Stroke. 2008;39:1621-1628.
Miller  D.C., Blackstone  E.H., Mack  M.J.; The PARTNER Trial Investigators and Patients, The PARTNER Stroke Substudy Writing Group Executive Committee. J Thorac Cardiovasc Surg. 2012;143:832-843 e13.
Ndrepepa  G., Schuster  T., Hadamitzky  M.; Validation of the bleeding academic research consortium definition of bleeding in patients with coronary artery disease undergoing percutaneous coronary intervention. Circulation. 2012;125:1424-1431.
Mehran  R., Rao  S.V., Bhatt  D.L.; Standardized bleeding definitions for cardiovascular clinical trials: a consensus report from the Bleeding Academic Research Consortium. Circulation. 2011;123:2736-2747.
Ussia  G.P., Barbanti  M., Petronio  A.S.; CoreValve Italian Registry Investigators Transcatheter aortic valve implantation: 3-year outcomes of self-expanding CoreValve prosthesis. Eur Heart J. 2012;33:969-976.
Mehta  R.L., Kellum  J.A., Shah  S.V.; Acute Kidney Injury N. Crit Care. 2007;11:R31
Kidney Disease,  Improving Global Outcomes (KDIGO) Acute Kidney Injury Work Group. Kidney int Suppl. 2012;2:1-138.
Généreux  P., Kodali  S., Leon  M.B.; Clinical outcomes using a new crossover balloon occlusion technique for percutaneous closure after transfemoral aortic valve implantation. JACC Cardiovasc Inters. 2011;4:861-867.
Sharp  A.S., Michev  I., Maisano  F.; A new technique for vascular access management in transcatheter aortic valve implantation. Catheter Cardiovasc Interv. 2010;75:784-793.
Camm  A.J., Kirchhof  P., Lip  G.Y.;European Heart Rhythm Association, European Association for Cardio-Thoracic Surgery,  Guidelines for the management of atrial fibrillation: the Task Force for the Management of Atrial Fibrillation of the European Society of Cardiology (ESC). Eur Heart J. 2010;31:2369-2429.
Al Ali  A.M., Altwegg  L., Horlick  E.M.; Prevention and management of transcatheter balloon-expandable aortic valve malposition. Catheter Cardiovasc Interv. 2008;72:573-578.
Geisbusch  S., Bleiziffer  S., Mazzitelli  D., Ruge  H., Bauernschmitt  R., Lange  R.; Incidence and management of CoreValve dislocation during transcatheter aortic valve implantation. Circ Cardiovasc Interv. 2010;3:531-536.
Gerckens  U., Latsios  G., Mueller  R.; Procedural and mid-term results in patients with aortic stenosis treated with implantation of 2 (in-series) CoreValve prostheses in 1 procedure. J Am Coll Cardiol Intv. 2010;3:244-250.
Bonow  R.O., Carabello  B.A., Chatterjee  K.; American College of Cardiology/American Heart Association Task Force on Practice Guidelines. 2008 focused update incorporated into the ACC/AHA 2006 guidelines for the management of patients with valvular heart disease: a report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines (Writing Committee to revise the 1998 guidelines for the management of patients with valvular heart disease). J Am Coll Cardiol. 2008;52:e1-e142.
Zoghbi  W.A., Chambers  J.B., Dumesni  J.G.; Recommendations for evaluation of prosthetic valves with echocardiography and Doppler ultrasound: a report From the American Society of Echocardiography's Guidelines and Standards Committee and the Task Force on Prosthetic Valves, developed in conjunction with the American College of Cardiology Cardiovascular Imaging Committee, Cardiac Imaging Committee of the American Heart Association, the European Association of Echocardiography, a registered branch of the European Society of Cardiology, the Japanese Society of Echocardiography and the Canadian Society of Echocardiography, endorsed by the American College of Cardiology Foundation, American Heart Association, European Association of Echocardiography, a registered branch of the European Society of Cardiology, the Japanese Society of Echocardiography, and Canadian Society of Echocardiography. J Am Soc Echocardiogr. 2009;22:975-1014.
Rothbart  R.M., Castriz  J.L., Harding  L.V., Russo  C.D., Teague  S.M.; Determination of aortic valve area by two-dimensional and Doppler echocardiography in patients with normal and stenotic bioprosthetic valves. J Am Coll Cardiol. 1990;15:817-824.
Ben Zekry  S., Saad  R.M., Ozkan  M.; Flow acceleration time and ratio of acceleration time to ejection time for prosthetic aortic valve function. J Am Coll Cardiol Img. 2011;4:1161-1170.
Leon  M.B., Smith  C.R., Mack  M.; PARTNER Trial Investigators. N Engl J Med. 2010;363:1597-1607.
Smith  C.R., Leon  M.B., Mack  M.J.; PARTNER Trial Investigators. N Engl J Med. 2011;364:2187-2198.
Clavel  M.A., Webb  J.G., Pibarot  P.; Comparison of the hemodynamic performance of percutaneous and surgical bioprostheses for the treatment of severe aortic stenosis. J Am Coll Cardiol. 2009;53:1883-1891.
Shames  S., Koczo  A., Hahn  R., Jin  Z., Picard  M.H., Gillam  L.D.; Flow characteristics of the SAPIEN aortic valve: The importance of recognizing in-stent flow acceleration for the echocardiographic assessment of valve function. J Am Soc Echocardiogr. 2012;25:603-609.
Head  S.J., Mokhles  M.M., Osnabrugge  R.L.; The impact of prosthesis-patient mismatch on long-term survival after aortic valve replacement: a systematic review and meta-analysis of 34 observational studies comprising 27186 patients with 133 141 patient-years. Eur Heart J. 2012;33:1518-1529.
Kodali  S.K., Williams  M.R., Smith  C.R.; Two-year outcomes after transcatheter or surgical aortic-valve replacement. N Engl J Med. 2012;366:1686-1695.
Tamburino  C., Capodanno  D., Ramondo  A.; Incidence and predictors of early and late mortality after transcatheter aortic valve implantation in 663 patients with severe aortic stenosis. Circulation. 2011;123:299-308.
Vahanian  A., Baumgartner  H., Bax  J.; Task Force on the Management of Valvular Hearth Disease of the European Society of Cardiology, E. Eur Heart J. 2007;28:230-268.
Zamorano  J.L., Badano  L.P., Bruce  C.; EAE/ASE recommendations for the use of echocardiography in new transcatheter interventions for valvular heart disease. Eur Heart J. 2011;32:2189-2214.
Zoghbi  W.A., Enriquez-Sarano  M., Foster  E.; Recommendations for evaluation of the severity of native valvular regurgitation with two-dimensional and Doppler echocardiography. J Am Soc Echocardiogr. 2003;16:777-802.
Rector  T.S., Cohn  J.N.;Pimobendan Multicenter Research Group,  Assessment of patient outcome with the Minnesota Living with Heart Failure questionnaire: reliability and validity during a randomized, double-blind, placebo-controlled trial of pimobendan. Am Heart J. 1992;124:1017-1025.
Green  C.P., Porter  C.B., Bresnahan  D.R., Spertus  J.A.; Development and evaluation of the Kansas City Cardiomyopathy Questionnaire: a new health status measure for heart failure. J Am Coll Cardiol. 2000;35:1245-1255.
Spertus  J., Peterson  E., Conard  M.W.; Monitoring clinical changes in patients with heart failure: a comparison of methods. Am Heart J. 2005;150:707-715.
Supino  P.G., Borer  J.S., Franciosa  J.A.; Acceptability and psychometric properties of the Minnesota Living With Heart Failure Questionnaire among patients undergoing heart valve surgery: validation and comparison with SF-36. J Card Fail. 2009;15:267-277.
Reynolds  M.R., Magnuson  E.A., Lei  Y.; Placement of Aortic Transcatheter Valves (PARTNER) Investigators. Circulation. 2011;124:1964-1972.
McHorney  C.A., Ware  J.E., Raczek  A.E., Ware  J.E.; The MOS 36-Item Short-Form Health Survey (SF-36): II. Med Care. 1993;31:247-263.
Shaw  J.W., Johnson  J.A., Coons  S.J.; US valuation of the EQ-5D health states: development and testing of the D1 valuation model. Med Care. 2005;43:203-220.
Ware  J., Kosinski  M., Keller  S.D., Ware  J.; A 12-Item Short-Form Health Survey: construction of scales and preliminary tests of reliability and validity. Med Care. 1996;34:220-233.
Van Dijk  D., Jansen  E.W., Hijman  R.; Cognitive outcome after off-pump and on-pump coronary artery bypass graft surgery: a randomized trial. JAMA. 2002;287:1405-1412.
Akins  C.W., Miller  D.C., Turina  M.I.; Guidelines for reporting mortality and morbidity after cardiac valve interventions. Eur J Cardiothorac Surg. 2008;33:523-528.
Gotzmann  M., Pljakic  A., Bojara  W., Lindstaedt  M., Ewers  A., Germing  A., Mugge  A.; Transcatheter aortic valve implantation in patients with severe symptomatic aortic valve stenosis-predictors of mortality and poor treatment response. Am Heart J. 2011;162:238-245 e1.
Giugliano  G.R., Lotfi  A.S.; Transcatheter aortic valve implantation: comparing apples to apples. J Invasive Cardiol. 2011;23:313-315.
Guetta  V., Goldenberg  G., Segev  A.; Predictors and course of high-degree atrioventricular block after transcatheter aortic valve implantation using the CoreValve Revalving System. Am J Cardiol. 2011;108:1600-1605.
Amat-Santos  I.J., Rodes-Cabau  J., Urena  M.; Incidence, predictive factors, and prognostic value of new-onset atrial fibrillation following transcatheter aortic valve implantation. J Am Coll Cardiol. 2012;59:178-188.
Fraccaro  C., Buja  G., Tarantini  G.; Incidence, predictors, and outcome of conduction disorders after transcatheter self-expandable aortic valve implantation. Am J Cardiol. 2011;107:747-754.
Nuis  R.J., Van Mieghem  N.M., Schultz  C.J.; Timing and potential mechanisms of new conduction abnormalities during the implantation of the Medtronic CoreValve System in patients with aortic stenosis. Eur Heart J. 2011;32:2067-2074.
Seiffert  M., Conradi  L., Baldus  S.; Impact of patient-prosthesis mismatch after transcatheter aortic valve-in-valve implantation in degenerated bioprostheses. J Thorac Cardiovasc Surg. 2012;143:617-624.
Stassano  P., Di Tommaso  L., Monaco  M.; Aortic valve replacement: a prospective randomized evaluation of mechanical versus biological valves in patients ages 55 to 70 years. J Am Coll Cardiol. 2009;54:1862-1868.
Rudski  L.G., Lai  W.W., Afilalo  J., Hua  L.; Guidelines for the echocardiographic assessment of the right heart in adults: a report from the American Society of Echocardiography endorsed by the European Association of Echocardiography, a registered branch of the European Society of Cardiology, and the Canadian Society of Echocardiography. J Am Soc Echocardiogr. 2010;23:685-713.
Durack  D.T., Lukes  A.S., Bright  D.K.; New criteria for diagnosis of infective endocarditis: utilization of specific echocardiographic findings. Am J Med. 1994;96:200-209.

Figures

Grahic Jump Location
Figure 1

Porcelain Aorta or Severely Atherosclerotic Aorta

Grahic Jump Location
Figure 3

Patent IMA Graft Crossing Midline and/or Adherent to the Posterior Table of Sternum

Grahic Jump Location
Figure 4

Transcatheter Heart Valve Haemodynamic Evaluation Algorithm

Tables

Table Grahic Jump Location
Table 1Risk factors Not Captured by Traditional Risk Scores
Table Grahic Jump Location
Table 2Mortality
Table Grahic Jump Location
Table 3Myocardial Infarction
Table Grahic Jump Location
Table 4Stroke and TIA
Table Grahic Jump Location
Table 5Bleeding
Table Grahic Jump Location
Table 6Acute Kidney Injury (AKIN ClassificationTable fn7)
Table Grahic Jump Location
Table 7Vascular Access Site and Access-Related Complications
Table Grahic Jump Location
Table 8Conduction Disturbances and Arrhythmias
Table Grahic Jump Location
Table 9Other TAVI-Related Complications
Table Grahic Jump Location
Table 10Prosthetic Valve Dysfunction
Table Grahic Jump Location
Table 11Composite Endpoints

Interactive Graphics

Video

References

Leon  M.B., Piazza  N., Nikolsky  E.; Standardized endpoint definitions for transcatheter aortic valve implantation clinical trials: a consensus report from the Valve Academic Research Consortium. J Am Coll Cardiol. 2011;57:253-269.
PubMed | CrossRef
Leon  M.B., Piazza  N., Nikolsky  E.; Standardized endpoint definitions for transcatheter aortic valve implantation clinical trials: a consensus report from the Valve Academic Research Consortium. Eur Heart J. 2011;32:205-217.
Généreux  P., Head  S.J., Van Mieghem  N.M.; Clinical outcomes after transcatheter aortic valve replacement using Valve Academic Research Consortium definitions: a weighted meta-analysis of 3,519 patients from 16 studies. J Am Coll Cardiol. 2012;59:2317-2326.
Clavel  M.A., Rodes-Cabau  J., Dumont  E.; Validation and characterization of transcatheter aortic valve effective orifice area measured by Doppler echocardiography. J Am Coll Cardiol Img. 2011;4:1053-1062.
Ikeda  K., Ho  M., Kawahara  M.; Valve academic research consortium consensus report the pharmaceutical and medical devices agency perspective. J Am Coll Cardiol. 2011;58:777
Gurvitch  R., Toggweiler  S., Willson  A.B.; Outcomes and complications of transcatheter aortic valve replacement using a balloon expandable valve according to the Valve Academic Research Consortium (VARC) guidelines. Eurolntervention. 2011;7:41-48.
Delgado  V., Schuijf  J.D., Bax  J.J.; Pre-operative aortic valve implantation evaluation: multimodality imaging. Eurolntervention. 2010;6:G38-G47.
Piazza  N., Wenaweser  P., van Gameren  M.; Relationship between the logistic EuroSCORE and the Society of Thoracic Surgeons Predicted Risk of Mortality score in patients implanted with the CoreValve ReValving system—a Bern-Rotterdam Study. Am Heart J. 2010;159:323-329.
Rosenhek  R., Iung  B., Tornos  P.; ESC Working Group on Valvular Heart Disease Position Paper: assessing the risk of interventions in patients with valvular heart disease. Eur Heart J. 2012;33:822-828.
Afilalo  J., Mottillo  S., Eisenberg  M.J.; Addition of frailty and disability to cardiac surgery risk scores identifies elderly patients at high risk of mortality or major morbidity. Circ Cardiovasc Qual Outcomes. 2012;5:222-228.
Pal  S.K., Katheria  V., Hurria  A.; Evaluating the older patient with cancer: understanding frailty and the geriatric assessment. CA Cancer J Clin. 2010;60:120-132.
Morris  J.C.; The clinical dementia rating (CDR): current version and scoring rules. Neurology. 1993;43:2412-2414.
Holmes  D.R., Mack  M.J., Kaul  S.; 2012 ACCF/AATS/SCAI/STS expert consensus document on transcatheter aortic valve replacement: developed in collaboration with the American Heart Association, American Society of Echocardiography, European Association for Cardio-Thoracic Surgery, Heart Failure Society of America, Mended Hearts, Society of Cardiovascular Anesthesiologists, Society of Cardiovascular Computed Tomography, and Society for Cardiovascular Magnetic Resonance. Ann Thorac Surg. 2012;93:1340-1345.
Yourman  L.C., Lee  S.J., Schonberg  M.A., Widera  E.W., Smith  A.K.; Prognostic indices for older adults: a systematic review. JAMA. 2012;307:182-192.
Sianos  G., Morel  M.A., Kappetein  A.P.; The SYNTAX Score: an angio- graphic tool grading the complexity of coronary artery disease. EuroIntervention. 2005;1:219-227.
Serruys  P.W., Morice  M.C., Kappetein  A.P.; Percutaneous coronary intervention versus coronary-artery bypass grafting for severe coronary artery disease. N Engl J Med. 2009;360:961-972.
Dewey  T.M., Brown  D.L., Herbert  M.A.; Effect of concomitant coronary artery disease on procedural and late outcomes of transcatheter aortic valve implantation. Ann Thorac Surg. 2010;89:758-767.
Wenaweser  P., Pilgrim  T., Guerios  E.; Impact of coronary artery disease and percutaneous coronary intervention on outcomes in patients with severe aortic stenosis undergoing transcatheter aortic valve implantation. EuroIntervention. 2011;7:541-548.
Rodes-Cabau  J., Gutierrez  M., Bagur  R.; Incidence, predictive factors, and prognostic value of myocardial injury following uncomplicated transcatheter aortic valve implantation. J Am Coll Cardiol. 2011;57:1988-1999.
Thygesen  K., Alpert  J.S., White  H.D.; Joint ESC/ACCF/AHA/WHF Task Force for the Redefinition of Myocardial Infarction. Eur Heart J. 2007;28:2525-2538.
Daneault  B., Kirtane  A.J., Kodali  S.K.; Stroke associated with surgical and transcatheter treatment of aortic stenosis: a comprehensive review. J Am Coll Cardiol. 2011;58:2143-2150.
Brott  T., Adams  H.P., Olinger  C.P.; Measurements of acute cerebral infarction: a clinical examination scale. Stroke. 1989;20:864-870.
Bonita  R., Beaglehole  R.; Recovery of motor function after stroke. Stroke. 1988;19:1497-1500.
Lyden  P.D., Lau  G.T.; A critical-appraisal of stroke evaluation and rating-scales. Stroke. 1991;22:1345-1352.
Wintermark  M., Albers  G.W., Alexandrov  A.V.; Acute stroke imaging research roadmap. Stroke. 2008;39:1621-1628.
Miller  D.C., Blackstone  E.H., Mack  M.J.; The PARTNER Trial Investigators and Patients, The PARTNER Stroke Substudy Writing Group Executive Committee. J Thorac Cardiovasc Surg. 2012;143:832-843 e13.
Ndrepepa  G., Schuster  T., Hadamitzky  M.; Validation of the bleeding academic research consortium definition of bleeding in patients with coronary artery disease undergoing percutaneous coronary intervention. Circulation. 2012;125:1424-1431.
Mehran  R., Rao  S.V., Bhatt  D.L.; Standardized bleeding definitions for cardiovascular clinical trials: a consensus report from the Bleeding Academic Research Consortium. Circulation. 2011;123:2736-2747.
Ussia  G.P., Barbanti  M., Petronio  A.S.; CoreValve Italian Registry Investigators Transcatheter aortic valve implantation: 3-year outcomes of self-expanding CoreValve prosthesis. Eur Heart J. 2012;33:969-976.
Mehta  R.L., Kellum  J.A., Shah  S.V.; Acute Kidney Injury N. Crit Care. 2007;11:R31
Kidney Disease,  Improving Global Outcomes (KDIGO) Acute Kidney Injury Work Group. Kidney int Suppl. 2012;2:1-138.
Généreux  P., Kodali  S., Leon  M.B.; Clinical outcomes using a new crossover balloon occlusion technique for percutaneous closure after transfemoral aortic valve implantation. JACC Cardiovasc Inters. 2011;4:861-867.
Sharp  A.S., Michev  I., Maisano  F.; A new technique for vascular access management in transcatheter aortic valve implantation. Catheter Cardiovasc Interv. 2010;75:784-793.
Camm  A.J., Kirchhof  P., Lip  G.Y.;European Heart Rhythm Association, European Association for Cardio-Thoracic Surgery,  Guidelines for the management of atrial fibrillation: the Task Force for the Management of Atrial Fibrillation of the European Society of Cardiology (ESC). Eur Heart J. 2010;31:2369-2429.
Al Ali  A.M., Altwegg  L., Horlick  E.M.; Prevention and management of transcatheter balloon-expandable aortic valve malposition. Catheter Cardiovasc Interv. 2008;72:573-578.
Geisbusch  S., Bleiziffer  S., Mazzitelli  D., Ruge  H., Bauernschmitt  R., Lange  R.; Incidence and management of CoreValve dislocation during transcatheter aortic valve implantation. Circ Cardiovasc Interv. 2010;3:531-536.
Gerckens  U., Latsios  G., Mueller  R.; Procedural and mid-term results in patients with aortic stenosis treated with implantation of 2 (in-series) CoreValve prostheses in 1 procedure. J Am Coll Cardiol Intv. 2010;3:244-250.
Bonow  R.O., Carabello  B.A., Chatterjee  K.; American College of Cardiology/American Heart Association Task Force on Practice Guidelines. 2008 focused update incorporated into the ACC/AHA 2006 guidelines for the management of patients with valvular heart disease: a report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines (Writing Committee to revise the 1998 guidelines for the management of patients with valvular heart disease). J Am Coll Cardiol. 2008;52:e1-e142.
Zoghbi  W.A., Chambers  J.B., Dumesni  J.G.; Recommendations for evaluation of prosthetic valves with echocardiography and Doppler ultrasound: a report From the American Society of Echocardiography's Guidelines and Standards Committee and the Task Force on Prosthetic Valves, developed in conjunction with the American College of Cardiology Cardiovascular Imaging Committee, Cardiac Imaging Committee of the American Heart Association, the European Association of Echocardiography, a registered branch of the European Society of Cardiology, the Japanese Society of Echocardiography and the Canadian Society of Echocardiography, endorsed by the American College of Cardiology Foundation, American Heart Association, European Association of Echocardiography, a registered branch of the European Society of Cardiology, the Japanese Society of Echocardiography, and Canadian Society of Echocardiography. J Am Soc Echocardiogr. 2009;22:975-1014.
Rothbart  R.M., Castriz  J.L., Harding  L.V., Russo  C.D., Teague  S.M.; Determination of aortic valve area by two-dimensional and Doppler echocardiography in patients with normal and stenotic bioprosthetic valves. J Am Coll Cardiol. 1990;15:817-824.
Ben Zekry  S., Saad  R.M., Ozkan  M.; Flow acceleration time and ratio of acceleration time to ejection time for prosthetic aortic valve function. J Am Coll Cardiol Img. 2011;4:1161-1170.
Leon  M.B., Smith  C.R., Mack  M.; PARTNER Trial Investigators. N Engl J Med. 2010;363:1597-1607.
Smith  C.R., Leon  M.B., Mack  M.J.; PARTNER Trial Investigators. N Engl J Med. 2011;364:2187-2198.
Clavel  M.A., Webb  J.G., Pibarot  P.; Comparison of the hemodynamic performance of percutaneous and surgical bioprostheses for the treatment of severe aortic stenosis. J Am Coll Cardiol. 2009;53:1883-1891.
Shames  S., Koczo  A., Hahn  R., Jin  Z., Picard  M.H., Gillam  L.D.; Flow characteristics of the SAPIEN aortic valve: The importance of recognizing in-stent flow acceleration for the echocardiographic assessment of valve function. J Am Soc Echocardiogr. 2012;25:603-609.
Head  S.J., Mokhles  M.M., Osnabrugge  R.L.; The impact of prosthesis-patient mismatch on long-term survival after aortic valve replacement: a systematic review and meta-analysis of 34 observational studies comprising 27186 patients with 133 141 patient-years. Eur Heart J. 2012;33:1518-1529.
Kodali  S.K., Williams  M.R., Smith  C.R.; Two-year outcomes after transcatheter or surgical aortic-valve replacement. N Engl J Med. 2012;366:1686-1695.
Tamburino  C., Capodanno  D., Ramondo  A.; Incidence and predictors of early and late mortality after transcatheter aortic valve implantation in 663 patients with severe aortic stenosis. Circulation. 2011;123:299-308.
Vahanian  A., Baumgartner  H., Bax  J.; Task Force on the Management of Valvular Hearth Disease of the European Society of Cardiology, E. Eur Heart J. 2007;28:230-268.
Zamorano  J.L., Badano  L.P., Bruce  C.; EAE/ASE recommendations for the use of echocardiography in new transcatheter interventions for valvular heart disease. Eur Heart J. 2011;32:2189-2214.
Zoghbi  W.A., Enriquez-Sarano  M., Foster  E.; Recommendations for evaluation of the severity of native valvular regurgitation with two-dimensional and Doppler echocardiography. J Am Soc Echocardiogr. 2003;16:777-802.
Rector  T.S., Cohn  J.N.;Pimobendan Multicenter Research Group,  Assessment of patient outcome with the Minnesota Living with Heart Failure questionnaire: reliability and validity during a randomized, double-blind, placebo-controlled trial of pimobendan. Am Heart J. 1992;124:1017-1025.
Green  C.P., Porter  C.B., Bresnahan  D.R., Spertus  J.A.; Development and evaluation of the Kansas City Cardiomyopathy Questionnaire: a new health status measure for heart failure. J Am Coll Cardiol. 2000;35:1245-1255.
Spertus  J., Peterson  E., Conard  M.W.; Monitoring clinical changes in patients with heart failure: a comparison of methods. Am Heart J. 2005;150:707-715.
Supino  P.G., Borer  J.S., Franciosa  J.A.; Acceptability and psychometric properties of the Minnesota Living With Heart Failure Questionnaire among patients undergoing heart valve surgery: validation and comparison with SF-36. J Card Fail. 2009;15:267-277.
Reynolds  M.R., Magnuson  E.A., Lei  Y.; Placement of Aortic Transcatheter Valves (PARTNER) Investigators. Circulation. 2011;124:1964-1972.
McHorney  C.A., Ware  J.E., Raczek  A.E., Ware  J.E.; The MOS 36-Item Short-Form Health Survey (SF-36): II. Med Care. 1993;31:247-263.
Shaw  J.W., Johnson  J.A., Coons  S.J.; US valuation of the EQ-5D health states: development and testing of the D1 valuation model. Med Care. 2005;43:203-220.
Ware  J., Kosinski  M., Keller  S.D., Ware  J.; A 12-Item Short-Form Health Survey: construction of scales and preliminary tests of reliability and validity. Med Care. 1996;34:220-233.
Van Dijk  D., Jansen  E.W., Hijman  R.; Cognitive outcome after off-pump and on-pump coronary artery bypass graft surgery: a randomized trial. JAMA. 2002;287:1405-1412.
Akins  C.W., Miller  D.C., Turina  M.I.; Guidelines for reporting mortality and morbidity after cardiac valve interventions. Eur J Cardiothorac Surg. 2008;33:523-528.
Gotzmann  M., Pljakic  A., Bojara  W., Lindstaedt  M., Ewers  A., Germing  A., Mugge  A.; Transcatheter aortic valve implantation in patients with severe symptomatic aortic valve stenosis-predictors of mortality and poor treatment response. Am Heart J. 2011;162:238-245 e1.
Giugliano  G.R., Lotfi  A.S.; Transcatheter aortic valve implantation: comparing apples to apples. J Invasive Cardiol. 2011;23:313-315.
Guetta  V., Goldenberg  G., Segev  A.; Predictors and course of high-degree atrioventricular block after transcatheter aortic valve implantation using the CoreValve Revalving System. Am J Cardiol. 2011;108:1600-1605.
Amat-Santos  I.J., Rodes-Cabau  J., Urena  M.; Incidence, predictive factors, and prognostic value of new-onset atrial fibrillation following transcatheter aortic valve implantation. J Am Coll Cardiol. 2012;59:178-188.
Fraccaro  C., Buja  G., Tarantini  G.; Incidence, predictors, and outcome of conduction disorders after transcatheter self-expandable aortic valve implantation. Am J Cardiol. 2011;107:747-754.
Nuis  R.J., Van Mieghem  N.M., Schultz  C.J.; Timing and potential mechanisms of new conduction abnormalities during the implantation of the Medtronic CoreValve System in patients with aortic stenosis. Eur Heart J. 2011;32:2067-2074.
Seiffert  M., Conradi  L., Baldus  S.; Impact of patient-prosthesis mismatch after transcatheter aortic valve-in-valve implantation in degenerated bioprostheses. J Thorac Cardiovasc Surg. 2012;143:617-624.
Stassano  P., Di Tommaso  L., Monaco  M.; Aortic valve replacement: a prospective randomized evaluation of mechanical versus biological valves in patients ages 55 to 70 years. J Am Coll Cardiol. 2009;54:1862-1868.
Rudski  L.G., Lai  W.W., Afilalo  J., Hua  L.; Guidelines for the echocardiographic assessment of the right heart in adults: a report from the American Society of Echocardiography endorsed by the European Association of Echocardiography, a registered branch of the European Society of Cardiology, and the Canadian Society of Echocardiography. J Am Soc Echocardiogr. 2010;23:685-713.
Durack  D.T., Lukes  A.S., Bright  D.K.; New criteria for diagnosis of infective endocarditis: utilization of specific echocardiographic findings. Am J Med. 1994;96:200-209.

Correspondence

Latest JACC CME

Continuing Medical Education through JACC is a convenient way to fulfill your CME requirements while learning important information about the latest advances in cardiovascular medicine.

April 2013- JACC CME Activity
Repeat Revascularization and Outcome

March 2013- JACC CME Activity
Extreme Lipoprotein(a) Levels and Improved Cardiovascular Risk Prediction

Feb 2013- JACC CME Activity
Results from the BARI 2D Trial

Jan 2013- JACC CME Activity
Prognosis Among Healthy Individuals Discharged With a Primary Diagnosis of Syncope

Dec 2012- JACC CME Activity
Incidence of Heart Failure or Cardiomyopathy After Adjuvant Trastuzumab Therapy for Breast Cancer

Nov 2012- JACC CME Activity
A Collaborative Analysis of Individual Patient Data From 10 Randomized Trials

Oct 2012- JACC CME Activity
Radiofrequency Ablation of Premature Ventricular Ectopy Improves the Efficacy of Cardiac Resynchronization Therapy in Nonresponders

Sept 2012- JACC CME Activity
Exercise and Pharmacological Treatment of Depressive Symptoms in Patients With Coronary Heart Disease

Aug 2012- JACC CME Activity
Reduction in Life-Threatening Ventricular Tachyarrhythmias in Statin-Treated Patients With Nonischemic Cardiomyopathy Enrolled in the MADIT-CRT (Multicenter Automatic Defibrillator Implantation Trial with Cardiac Resynchronization Therapy)

July 2012- JACC CME Activity
Relationship of Beta-Blocker Dose With Outcomes in Ambulatory Heart Failure Patients With Systolic Dysfunction

For previous CME quizzes, please follow this link to CardioSource Lifelong Learning and MOC.

 

NOTE:
Citing articles are presented as examples only. In non-demo SCM6 implementation, integration with CrossRef’s “Cited By” API will populate this tab (http://www.crossref.org/citedby.html).
Submit a Comment
Submit a Comment
Data Supplement
Supporting Data
Supplementary Content

Some tools below are only available to our subscribers or users with an online account.

Related Content

Customize your page view by dragging & repositioning the boxes below.

Related Editorial Content
Articles Related By Topic
Related Topics